Potential of Image2Image translation in reducing AI bias attributed to differences in CT reconstruction methods: proof-of-concept study on a paired dataset.

IF 4
Frontiers in radiology Pub Date : 2026-07-03 eCollection Date: 2026-01-01 DOI:10.3389/fradi.2026.1875867
Feifei Li, Liliana Lourenco Caldeira, Astha Jaiswal, Nils Große Hokamp, Oya Beyan, Ekaterina Kutafina
{"title":"Potential of Image2Image translation in reducing AI bias attributed to differences in CT reconstruction methods: proof-of-concept study on a paired dataset.","authors":"Feifei Li, Liliana Lourenco Caldeira, Astha Jaiswal, Nils Große Hokamp, Oya Beyan, Ekaterina Kutafina","doi":"10.3389/fradi.2026.1875867","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The choice of image reconstruction approach, such as Iterative Model Reconstruction (IMR) and iDose <math><msup><mi></mi> <mn>4</mn></msup> </math> , is critical in computed tomography (CT) as distribution shifts between them can introduce systematic bias in clinical AI models, posing a barrier to efficient, scalable AI deployment in radiology. While Image-to-Image (I2I) translation offers a path toward harmonization, efficient and translation-feasible AI in radiology demands that such approaches be validated not only on image quality metrics but also on downstream clinical task performance.</p><p><strong>Methods: </strong>We compared three harmonization strategies: CycleGAN, Denoising Diffusion Models, and a conventional Gaussian filter, using a paired CT dataset. Performance was evaluated using task-agnostic metrics, including Structural Similarity Index Measure (SSIM), Peak Signal-to-Noise Ratio (PSNR), Frechet Inception Distance (FID), and Sliced Wasserstein Distance (SWD). Crucially, we assessed clinical utility by measuring the downstream impact on multi-organ segmentation using Mean Relative Volume Error (RVE), focusing on consistency across reconstructed modes.</p><p><strong>Results: </strong>Our findings reveal a significant discrepancy between visual fidelity and clinical utility. While Diffusion and CycleGAN models achieve competitive feature-level similarity, they introduce catastrophic systematic biases in downstream segmentation, particularly in adipose tissues (Body Composition Analysis, BCA), where volume errors can reach <math><mo>-</mo></math> 100%, indicating a failure to preserve critical anatomical fat structures. In contrast, the computationally efficient Gaussian filter served as the most robust baseline, maintaining minimal RVE and the highest consistency between IMR and iDose <math><msup><mi></mi> <mn>4</mn></msup> </math> modes.</p><p><strong>Conclusions: </strong>The results suggest that in the context of efficient AI for radiology imaging, where efficiency encompasses not only computational cost but also reliability of clinical outputs, task-agnostic image quality metrics are insufficient proxies for downstream clinical performance. High-parameter generative models may introduce structural losses that remain invisible to such metrics yet critically affect clinical utility. The underperformance of generative approaches observed here is likely attributable to 3D slice inconsistencies during volumetric reconstruction and suboptimal model adaptation, rather than to an inherent limitation of I2I methods, which retain significant potential for mitigating reconstruction-related bias in radiology AI. Future work developing harmonization pipelines must therefore prioritize task-specific, clinically grounded validation as the primary benchmark-and carefully consider whether computationally simpler approaches, such as conventional filtering, may already provide comparably stable harmonization for specific use cases, before resorting to high-parameter generative solutions.</p>","PeriodicalId":73101,"journal":{"name":"Frontiers in radiology","volume":"6 ","pages":"1875867"},"PeriodicalIF":4.0000,"publicationDate":"2026-07-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13375877/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in radiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fradi.2026.1875867","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The choice of image reconstruction approach, such as Iterative Model Reconstruction (IMR) and iDose 4 , is critical in computed tomography (CT) as distribution shifts between them can introduce systematic bias in clinical AI models, posing a barrier to efficient, scalable AI deployment in radiology. While Image-to-Image (I2I) translation offers a path toward harmonization, efficient and translation-feasible AI in radiology demands that such approaches be validated not only on image quality metrics but also on downstream clinical task performance.

Methods: We compared three harmonization strategies: CycleGAN, Denoising Diffusion Models, and a conventional Gaussian filter, using a paired CT dataset. Performance was evaluated using task-agnostic metrics, including Structural Similarity Index Measure (SSIM), Peak Signal-to-Noise Ratio (PSNR), Frechet Inception Distance (FID), and Sliced Wasserstein Distance (SWD). Crucially, we assessed clinical utility by measuring the downstream impact on multi-organ segmentation using Mean Relative Volume Error (RVE), focusing on consistency across reconstructed modes.

Results: Our findings reveal a significant discrepancy between visual fidelity and clinical utility. While Diffusion and CycleGAN models achieve competitive feature-level similarity, they introduce catastrophic systematic biases in downstream segmentation, particularly in adipose tissues (Body Composition Analysis, BCA), where volume errors can reach - 100%, indicating a failure to preserve critical anatomical fat structures. In contrast, the computationally efficient Gaussian filter served as the most robust baseline, maintaining minimal RVE and the highest consistency between IMR and iDose 4 modes.

Conclusions: The results suggest that in the context of efficient AI for radiology imaging, where efficiency encompasses not only computational cost but also reliability of clinical outputs, task-agnostic image quality metrics are insufficient proxies for downstream clinical performance. High-parameter generative models may introduce structural losses that remain invisible to such metrics yet critically affect clinical utility. The underperformance of generative approaches observed here is likely attributable to 3D slice inconsistencies during volumetric reconstruction and suboptimal model adaptation, rather than to an inherent limitation of I2I methods, which retain significant potential for mitigating reconstruction-related bias in radiology AI. Future work developing harmonization pipelines must therefore prioritize task-specific, clinically grounded validation as the primary benchmark-and carefully consider whether computationally simpler approaches, such as conventional filtering, may already provide comparably stable harmonization for specific use cases, before resorting to high-parameter generative solutions.

image2图像翻译在减少CT重建方法差异导致的人工智能偏差方面的潜力:配对数据集的概念验证研究。
背景:图像重建方法的选择,如迭代模型重建(IMR)和iDose 4,在计算机断层扫描(CT)中至关重要,因为它们之间的分布变化会在临床人工智能模型中引入系统偏差,对放射学中高效、可扩展的人工智能部署构成障碍。虽然图像到图像(I2I)翻译为协调提供了一条途径,但放射学中高效和翻译可行的人工智能要求这些方法不仅要在图像质量指标上得到验证,还要在下游临床任务表现上得到验证。方法:我们使用配对CT数据集比较了三种协调策略:CycleGAN、去噪扩散模型和传统高斯滤波器。使用与任务无关的指标评估性能,包括结构相似指数测量(SSIM),峰值信噪比(PSNR), Frechet Inception距离(FID)和切片Wasserstein距离(SWD)。至关重要的是,我们通过使用平均相对体积误差(RVE)测量对多器官分割的下游影响来评估临床效用,重点关注重建模式之间的一致性。结果:我们的研究结果揭示了视觉保真度与临床实用性之间的显著差异。虽然Diffusion和CycleGAN模型实现了竞争性的特征级相似性,但它们在下游分割中引入了灾难性的系统偏差,特别是在脂肪组织(Body Composition Analysis, BCA)中,体积误差可以达到- 100%,表明未能保留关键的解剖脂肪结构。相比之下,计算效率高的高斯滤波器作为最稳健的基线,保持最小的RVE和IMR与iDose 4模式之间的最高一致性。结论:研究结果表明,在高效人工智能放射成像的背景下,效率不仅包括计算成本,还包括临床输出的可靠性,与任务无关的图像质量指标不足以代表下游临床表现。高参数生成模型可能会引入结构性损失,这些损失仍然不可见,但严重影响临床效用。这里观察到的生成方法的性能不佳可能归因于体积重建过程中的3D切片不一致和次优模型适应,而不是I2I方法的固有局限性,I2I方法在减轻放射学人工智能中与重建相关的偏差方面仍具有很大的潜力。因此,未来开发协调管道的工作必须优先考虑特定于任务的、基于临床的验证作为主要基准,并在采用高参数生成解决方案之前,仔细考虑计算更简单的方法(如传统过滤)是否已经可以为特定用例提供相对稳定的协调。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.20
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书