Evaluation of an Immunoturbidometric Assay for Measurement of Total Adiponectin Concentration in Equine Serum Samples.

IF 1 4区 农林科学 Q3 VETERINARY SCIENCES
Veterinary clinical pathology Pub Date : 2026-06-01 Epub Date: 2026-07-22 DOI:10.1111/vcp.70118
Rose D Nolen-Walston, Caitrin R Lowndes, Jeaneen C Kulp, Daniela Luethy, Darko Stefanovski, Andrew W Van Eps
{"title":"Evaluation of an Immunoturbidometric Assay for Measurement of Total Adiponectin Concentration in Equine Serum Samples.","authors":"Rose D Nolen-Walston, Caitrin R Lowndes, Jeaneen C Kulp, Daniela Luethy, Darko Stefanovski, Andrew W Van Eps","doi":"10.1111/vcp.70118","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Total adiponectin (TA) is a marker of metabolic health in equids, with values below a critical decision point of < 7.9 μg/mL conferring increased risk of laminitis.</p><p><strong>Objectives: </strong>To evaluate an IT assay to measure serum [TA] in equids and to assess its correlation with [high molecular weight adiponectin] (HMWA) measured by ELISA.</p><p><strong>Methods: </strong>Total serum adiponectin concentrations from 207 adult horses were measured using both an IT assay for TA and HMWA using a previously validated ELISA assay. Data were analyzed using Spearman correlation and Passing-Bablok regression. Intra- and inter-assay precision, linearity, lower limit of quantification (LLoQ), and freeze-thaw stability were assessed for the [TA] assay.</p><p><strong>Results: </strong>Mean (range) intra-assay coefficient of variation (CV%) was 2.3% (1.8%-2.8%); inter-assay, across-day CV% was 3.3% (3.25%-3.4%), and across reagent lot CV was 11.7% (9.0%-17.9%). Mean recovery upon dilution was 133.7% (122.6%-140.6%), and LLoQ 1.3 μg/mL. The median CV for 5 freeze-thaw cycles was 3.0% (range 1.0%-3.0%) with a median change in concentration of -11% (range -[9.9-17.1]%). A strong correlation between [TA] and [HMWA] is observed (ρ = 0.90 [95% confidence interval 0.86-0.92], p < 0.001), although in 18% of samples, [HMWA] > [TA].</p><p><strong>Conclusions: </strong>The turbidimetric immunoassay is sufficiently precise and reliable for measurement of [TA] in equids, especially around the critical decision point, and LLoQ was below the lowest clinical cut-off value. [TA] overall correlated well with [HMWA] in this population, although instances of higher [HMWA] than [TA] were noted at high [HMWA].</p>","PeriodicalId":23593,"journal":{"name":"Veterinary clinical pathology","volume":" ","pages":"239-245"},"PeriodicalIF":1.0000,"publicationDate":"2026-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13686539/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary clinical pathology","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.1111/vcp.70118","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/22 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Total adiponectin (TA) is a marker of metabolic health in equids, with values below a critical decision point of < 7.9 μg/mL conferring increased risk of laminitis.

Objectives: To evaluate an IT assay to measure serum [TA] in equids and to assess its correlation with [high molecular weight adiponectin] (HMWA) measured by ELISA.

Methods: Total serum adiponectin concentrations from 207 adult horses were measured using both an IT assay for TA and HMWA using a previously validated ELISA assay. Data were analyzed using Spearman correlation and Passing-Bablok regression. Intra- and inter-assay precision, linearity, lower limit of quantification (LLoQ), and freeze-thaw stability were assessed for the [TA] assay.

Results: Mean (range) intra-assay coefficient of variation (CV%) was 2.3% (1.8%-2.8%); inter-assay, across-day CV% was 3.3% (3.25%-3.4%), and across reagent lot CV was 11.7% (9.0%-17.9%). Mean recovery upon dilution was 133.7% (122.6%-140.6%), and LLoQ 1.3 μg/mL. The median CV for 5 freeze-thaw cycles was 3.0% (range 1.0%-3.0%) with a median change in concentration of -11% (range -[9.9-17.1]%). A strong correlation between [TA] and [HMWA] is observed (ρ = 0.90 [95% confidence interval 0.86-0.92], p < 0.001), although in 18% of samples, [HMWA] > [TA].

Conclusions: The turbidimetric immunoassay is sufficiently precise and reliable for measurement of [TA] in equids, especially around the critical decision point, and LLoQ was below the lowest clinical cut-off value. [TA] overall correlated well with [HMWA] in this population, although instances of higher [HMWA] than [TA] were noted at high [HMWA].

免疫浊度测定法测定马血清样品中总脂联素浓度的评价。
背景:总脂联素(TA)是马科动物代谢健康的标志,其值低于关键决策点。目的:评价一种用于测量马科动物血清[TA]的IT测定方法,并评估其与ELISA测定的[高分子量脂联素](HMWA)的相关性。方法:对207匹成年马的血清总脂联素浓度进行了测定,采用了TA的IT测定法和HMWA的ELISA测定法。数据分析采用Spearman相关和Passing-Bablok回归。评估了[TA]测定法的测定内和测定间的精密度、线性、定量下限(LLoQ)和冻融稳定性。结果:平均(范围)内变异系数(CV%)为2.3% (1.8% ~ 2.8%);测定间、跨日CV%为3.3%(3.25%-3.4%),跨试剂批次CV为11.7%(9.0%-17.9%)。稀释后平均回收率为133.7%(122.6% ~ 140.6%),定量限为1.3 μg/mL。5个冻融循环的中位数CV为3.0%(范围1.0%-3.0%),中位数浓度变化为-11%(范围-[9.9-17.1]%)。[TA]与[HMWA]之间存在很强的相关性(ρ = 0.90[95%置信区间0.86-0.92],p [TA])。结论:浊度免疫分析法对马科动物的[TA]测量足够精确和可靠,特别是在关键决策点附近,LLoQ低于最低临床临界值。在该人群中,[TA]总体上与[HMWA]相关良好,尽管[TA]高于[HMWA]的情况在高[HMWA]中被注意到。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Veterinary clinical pathology
Veterinary clinical pathology 农林科学-兽医学
CiteScore
1.70
自引率
16.70%
发文量
133
审稿时长
18-36 weeks
期刊介绍: Veterinary Clinical Pathology is the official journal of the American Society for Veterinary Clinical Pathology (ASVCP) and the European Society of Veterinary Clinical Pathology (ESVCP). The journal''s mission is to provide an international forum for communication and discussion of scientific investigations and new developments that advance the art and science of laboratory diagnosis in animals. Veterinary Clinical Pathology welcomes original experimental research and clinical contributions involving domestic, laboratory, avian, and wildlife species in the areas of hematology, hemostasis, immunopathology, clinical chemistry, cytopathology, surgical pathology, toxicology, endocrinology, laboratory and analytical techniques, instrumentation, quality assurance, and clinical pathology education.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书