Roberto Barreto Osaki, Raimundo Sales de Oliveira-Neto, Gabriel Nakamura Ramos de Luca, Ricardo Gariba da Silva, Guilherme Ferreira da Silva, Bruno Carvalho de Vasconcelos, Rodrigo Ricci Vivan, Murilo Priori Alcalde, Marco Antonio Hungaro Duarte
{"title":"In vitro assessment of integrated apex locator systems: working length accuracy, operational parameters, and mechanical effects on NiTi instruments.","authors":"Roberto Barreto Osaki, Raimundo Sales de Oliveira-Neto, Gabriel Nakamura Ramos de Luca, Ricardo Gariba da Silva, Guilherme Ferreira da Silva, Bruno Carvalho de Vasconcelos, Rodrigo Ricci Vivan, Murilo Priori Alcalde, Marco Antonio Hungaro Duarte","doi":"10.4317/jced.64103","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study aimed to evaluate the accuracy of four hybrid endodontic motors-VDW Gold (VDW), iRoot Pro (IRP), E-Connect S (ECS), and TriAuto ZX2 (TZX)-in maintaining length, speed, and angulation during alternating and continuous rotation kinematics and the influence of these parameters on instrument fatigue.</p><p><strong>Materials and methods: </strong>Forty-eight mandibular incisors (n=12 per group) were used to assess apex locator accuracy during glide path procedures with alternating rotation and auto apical stop (AAS) set to the apical foramen (AF: 0.0). Positioning precision was evaluated using micro-CT, while torsional resistance of instruments was tested per ISO 3630-1. Actual speed and angulation were measured using a tachometer and Arduino system. Cyclic fatigue testing was conducted with Only One File 25.08 instruments in an artificial curved canal (60°, 5 mm radius). Data were statistically analyzed with a significance level of 5%.</p><p><strong>Results: </strong>The evaluated devices exhibited similar precision (p>0.05). The TZX showed the lowest number of determinations beyond the AF (p<0.05). Regarding torsional testing, no significant differences were observed between groups (p>0.05). The equipment showed substantial differences compared to the values displayed on the screen (p<0.05). The Optimum Glide Path (OGP) motion provided significantly higher resistance to cyclic fatigue (p<0.05).</p><p><strong>Conclusions: </strong>All hybrid motors were accurate in length control, with TZX being the most conservative. Kinematics did not influence torsional resistance, but significant differences were found between programmed and real parameters.</p>","PeriodicalId":15376,"journal":{"name":"Journal of Clinical and Experimental Dentistry","volume":"18 7","pages":"e921-e928"},"PeriodicalIF":0.0000,"publicationDate":"2026-06-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13386448/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical and Experimental Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4317/jced.64103","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"Dentistry","Score":null,"Total":0}
引用次数: 0
Abstract
Background: This study aimed to evaluate the accuracy of four hybrid endodontic motors-VDW Gold (VDW), iRoot Pro (IRP), E-Connect S (ECS), and TriAuto ZX2 (TZX)-in maintaining length, speed, and angulation during alternating and continuous rotation kinematics and the influence of these parameters on instrument fatigue.
Materials and methods: Forty-eight mandibular incisors (n=12 per group) were used to assess apex locator accuracy during glide path procedures with alternating rotation and auto apical stop (AAS) set to the apical foramen (AF: 0.0). Positioning precision was evaluated using micro-CT, while torsional resistance of instruments was tested per ISO 3630-1. Actual speed and angulation were measured using a tachometer and Arduino system. Cyclic fatigue testing was conducted with Only One File 25.08 instruments in an artificial curved canal (60°, 5 mm radius). Data were statistically analyzed with a significance level of 5%.
Results: The evaluated devices exhibited similar precision (p>0.05). The TZX showed the lowest number of determinations beyond the AF (p<0.05). Regarding torsional testing, no significant differences were observed between groups (p>0.05). The equipment showed substantial differences compared to the values displayed on the screen (p<0.05). The Optimum Glide Path (OGP) motion provided significantly higher resistance to cyclic fatigue (p<0.05).
Conclusions: All hybrid motors were accurate in length control, with TZX being the most conservative. Kinematics did not influence torsional resistance, but significant differences were found between programmed and real parameters.
期刊介绍:
Indexed in PUBMED, PubMed Central® (PMC) since 2012 and SCOPUSJournal of Clinical and Experimental Dentistry is an Open Access (free access on-line) - http://www.medicinaoral.com/odo/indice.htm. The aim of the Journal of Clinical and Experimental Dentistry is: - Periodontology - Community and Preventive Dentistry - Esthetic Dentistry - Biomaterials and Bioengineering in Dentistry - Operative Dentistry and Endodontics - Prosthetic Dentistry - Orthodontics - Oral Medicine and Pathology - Odontostomatology for the disabled or special patients - Oral Surgery