CBCT dose reduction with HyperSight on an Ethos linear accelerator: A phantom-based TLD measurement study.

IF 4.2
Svenja Ganz, Sven Knobe, Marc Ries, Jan Palm, Markus Hecht, Hendrik Auerbach
{"title":"CBCT dose reduction with HyperSight on an Ethos linear accelerator: A phantom-based TLD measurement study.","authors":"Svenja Ganz, Sven Knobe, Marc Ries, Jan Palm, Markus Hecht, Hendrik Auerbach","doi":"10.1016/j.zemedi.2026.07.003","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The novel HyperSight cone-beam CT (CBCT) system (Varian Medical Systems, Siemens Healthineers, Palo Alto, USA) provides image quality comparable to native CT and offers advantages for contouring in online-adaptive radiotherapy. However, daily imaging increases organ dose. This study evaluates and compares organ doses across HyperSight protocols on an O-ring Linac and calculates the effective dose and secondary cancer risk associated with daily imaging.</p><p><strong>Methods: </strong>Thermoluminescent dosimeter (TLD) measurements were performed on an anthropomorphic phantom for all pre-configured CBCT protocols in HyperSight, with TLDs placed according to each entity (head, head and neck, thorax, chest, and pelvis). The resulting imaging doses were compared across the various protocols and TLD positions, followed by the calculation of effective doses and the associated secondary cancer risk.</p><p><strong>Results: </strong>HyperSight CBCT demonstrated lower organ doses compared with previous systems. Mean bladder and rectum dose was 10mGy versus 27mGy for OBI Version 1.4.13, while mean head dose was 2mGy compared with 1.2mGy for Halcyon 3.0. A left-right dose asymmetry was observed due to partial gantry rotation, resulting in higher exposure in beam-facing organs. Imaging dose varied considerably between protocols. Low Dose protocols reduced exposure by up to 55%, whereas slow and large protocols increased dose by 40-60%.</p><p><strong>Conclusion: </strong>HyperSight provides high-quality imaging with lower dose levels than previous CBCT generations. The observed beam geometry-related asymmetry should be considered in clinical use, and individualized protocol selection is essential to minimize unnecessary exposure in image guided and adaptive radiotherapy under ALARA principles.</p>","PeriodicalId":101315,"journal":{"name":"Zeitschrift fur medizinische Physik","volume":" ","pages":""},"PeriodicalIF":4.2000,"publicationDate":"2026-07-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zeitschrift fur medizinische Physik","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.zemedi.2026.07.003","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The novel HyperSight cone-beam CT (CBCT) system (Varian Medical Systems, Siemens Healthineers, Palo Alto, USA) provides image quality comparable to native CT and offers advantages for contouring in online-adaptive radiotherapy. However, daily imaging increases organ dose. This study evaluates and compares organ doses across HyperSight protocols on an O-ring Linac and calculates the effective dose and secondary cancer risk associated with daily imaging.

Methods: Thermoluminescent dosimeter (TLD) measurements were performed on an anthropomorphic phantom for all pre-configured CBCT protocols in HyperSight, with TLDs placed according to each entity (head, head and neck, thorax, chest, and pelvis). The resulting imaging doses were compared across the various protocols and TLD positions, followed by the calculation of effective doses and the associated secondary cancer risk.

Results: HyperSight CBCT demonstrated lower organ doses compared with previous systems. Mean bladder and rectum dose was 10mGy versus 27mGy for OBI Version 1.4.13, while mean head dose was 2mGy compared with 1.2mGy for Halcyon 3.0. A left-right dose asymmetry was observed due to partial gantry rotation, resulting in higher exposure in beam-facing organs. Imaging dose varied considerably between protocols. Low Dose protocols reduced exposure by up to 55%, whereas slow and large protocols increased dose by 40-60%.

Conclusion: HyperSight provides high-quality imaging with lower dose levels than previous CBCT generations. The observed beam geometry-related asymmetry should be considered in clinical use, and individualized protocol selection is essential to minimize unnecessary exposure in image guided and adaptive radiotherapy under ALARA principles.

HyperSight在Ethos直线加速器上降低CBCT剂量:基于幻像的TLD测量研究。
目的:新型HyperSight锥形束CT (CBCT)系统(Varian Medical Systems, Siemens Healthineers, Palo Alto, USA)提供与原生CT相当的图像质量,并为在线自适应放疗的轮廓提供优势。然而,每日影像学检查增加了器官剂量。本研究评估和比较了在o型环Linac上HyperSight方案的器官剂量,并计算了与每日成像相关的有效剂量和继发性癌症风险。方法:对HyperSight中所有预配置CBCT方案的拟人体进行热发光剂量计(TLD)测量,并根据每个实体(头部、头颈、胸部、胸部和骨盆)放置TLD。比较不同方案和TLD位置的成像剂量,然后计算有效剂量和相关的继发性癌症风险。结果:与以前的系统相比,HyperSight CBCT显示出更低的器官剂量。膀胱和直肠的平均剂量为10mGy,而OBI版本1.4.13为27mGy,头部的平均剂量为2mGy,而Halcyon 3.0为1.2mGy。由于部分龙门旋转,观察到左右剂量不对称,导致在面向光束的器官中较高的暴露。不同方案的成像剂量差异很大。低剂量方案减少了高达55%的暴露,而缓慢和大剂量方案则使剂量增加了40-60%。结论:与前几代CBCT相比,HyperSight以更低的剂量水平提供了高质量的成像。在临床应用中应考虑到观察到的光束几何相关的不对称性,并根据ALARA原则选择个性化的方案以减少图像引导和适应性放疗中不必要的暴露。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书