Dongyue Wu, Florian Lux, Max Mai Tobon, Holger Muenz, Soeren Schmidt, Eleonora Tagliabue, Philipp Matten, Çağlar Ataman
{"title":"Sensor-integrated dual-clad fiber probe for OCT-guided retinal endolaser photocoagulation.","authors":"Dongyue Wu, Florian Lux, Max Mai Tobon, Holger Muenz, Soeren Schmidt, Eleonora Tagliabue, Philipp Matten, Çağlar Ataman","doi":"10.1117/1.JBO.31.7.077001","DOIUrl":null,"url":null,"abstract":"<p><strong>Significance: </strong>Retinal endolaser photocoagulation is a widely performed vitreoretinal procedure but currently lacks real-time intraoperative feedback on treatment outcomes. Integrating optical coherence tomography (OCT) directly into the endolaser probe enables co-localized, intraoperative sensing of tissue response during laser delivery, offering the potential to significantly enhance surgical safety and efficacy.</p><p><strong>Aim: </strong>The aim of the study is to design, fabricate, and demonstrate a fiber-optic smart instrument with a 3D nano-printed microlens and instrument-integrated OCT sensing that enables co-localized, concurrent surgical laser delivery and tomographic sensing of tissue alterations during REPC.</p><p><strong>Approach: </strong>A dual-clad fiber was used to co-axially deliver surgical and OCT beams through the outer multi-mode and inner single-mode cores. A 3D nano-printed aspherical microlens ( <math><mrow><mn>300</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> diameter) at the fiber tip optimized beam shaping for both lasers in the intravitreal cavity. The optical performance of the 23G instrument-integrated OCT (iiOCT) probe was characterized, and <i>ex vivo</i> validation experiments were conducted on porcine eyes.</p><p><strong>Results: </strong>Optical performance assessments demonstrated close alignment between measured and simulated beam profiles, with an OCT spot size of <math><mrow><mn>37.9</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> at best focus and a surgical laser beam divergence half-angle of 6.0 deg. The probe demonstrated stable power transmission with a mean deviation of 0.35% across repeated pulses. <i>Ex vivo</i> experiments confirmed co-localized, concurrent surgical laser delivery and OCT sensing of retinal tissue alterations.</p><p><strong>Conclusions: </strong>The miniaturized iiOCT endolaser probe combines surgical laser delivery with OCT sensing, achieving adequate performance for the targeted clinical application. This sensor-integrated instrument has the potential to enable intraocular, intraoperative, and quantitative assessment of treatment outcomes. It could support feedback-driven, robotic endolaser photocoagulation, enhancing the safety and efficacy of retinal endolaser surgery.</p>","PeriodicalId":15264,"journal":{"name":"Journal of Biomedical Optics","volume":"31 7","pages":"077001"},"PeriodicalIF":3.3000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13371110/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Biomedical Optics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1117/1.JBO.31.7.077001","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/15 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
Significance: Retinal endolaser photocoagulation is a widely performed vitreoretinal procedure but currently lacks real-time intraoperative feedback on treatment outcomes. Integrating optical coherence tomography (OCT) directly into the endolaser probe enables co-localized, intraoperative sensing of tissue response during laser delivery, offering the potential to significantly enhance surgical safety and efficacy.
Aim: The aim of the study is to design, fabricate, and demonstrate a fiber-optic smart instrument with a 3D nano-printed microlens and instrument-integrated OCT sensing that enables co-localized, concurrent surgical laser delivery and tomographic sensing of tissue alterations during REPC.
Approach: A dual-clad fiber was used to co-axially deliver surgical and OCT beams through the outer multi-mode and inner single-mode cores. A 3D nano-printed aspherical microlens ( diameter) at the fiber tip optimized beam shaping for both lasers in the intravitreal cavity. The optical performance of the 23G instrument-integrated OCT (iiOCT) probe was characterized, and ex vivo validation experiments were conducted on porcine eyes.
Results: Optical performance assessments demonstrated close alignment between measured and simulated beam profiles, with an OCT spot size of at best focus and a surgical laser beam divergence half-angle of 6.0 deg. The probe demonstrated stable power transmission with a mean deviation of 0.35% across repeated pulses. Ex vivo experiments confirmed co-localized, concurrent surgical laser delivery and OCT sensing of retinal tissue alterations.
Conclusions: The miniaturized iiOCT endolaser probe combines surgical laser delivery with OCT sensing, achieving adequate performance for the targeted clinical application. This sensor-integrated instrument has the potential to enable intraocular, intraoperative, and quantitative assessment of treatment outcomes. It could support feedback-driven, robotic endolaser photocoagulation, enhancing the safety and efficacy of retinal endolaser surgery.
期刊介绍:
The Journal of Biomedical Optics publishes peer-reviewed papers on the use of modern optical technology for improved health care and biomedical research.