Luca Conci, Anne-Kristin Schaefer, Thomas Schlöglhofer, Emilio Osorio-Jaramillo, Amila Kahrovic, Alvaro Perazzo, Marcus Granegger, Daniel Zimpfer, Dominik Wiedemann, Julia Riebandt
{"title":"Post-Market Analysis of the Xenios Console With the Xlung Tubing Kit 230 for Extracorporeal Membrane Oxygenation After Cardiac Surgery","authors":"Luca Conci, Anne-Kristin Schaefer, Thomas Schlöglhofer, Emilio Osorio-Jaramillo, Amila Kahrovic, Alvaro Perazzo, Marcus Granegger, Daniel Zimpfer, Dominik Wiedemann, Julia Riebandt","doi":"10.1111/aor.70194","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objectives</h3>\n \n <p>The use of veno-arterial extracorporeal membrane oxygenation (V-A ECMO) has led to an increasing number of devices available on the market. Most studies pay less attention to which devices have been used; therefore, the information on performance for each device is limited. The current study is a comprehensive post-market analysis of all the major functions and adverse events rates of the Xenios console and XLung kit 230 preconfigured ECMO tubing set in post-cardiotomy applications.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>This retrospective, single-center analysis, conducted between 2017 and 2021, included 47 adult patients who received V-A ECMO support using the Xenios system following cardiac surgery at our institution.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The Xenios device was found to effectively improve and maintain microcirculation, organ perfusion, gas exchanges, organ function support, and stabilization of the cardiocirculatory system, while maintaining pump performance and stable body core temperature. Device-related adverse events occurred in 28% of patients but resolved in 81% of cases without leaving clinical sequelae. Fewer hemolysis events than reported in the literature were recorded. Every patient experienced at least two patient-related adverse events, but 43% had no relation to the device and 45% resolved without permanent damage.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>The performance of the evaluated device was demonstrated in all key clinical areas, with an adequate safety profile.</p>\n </section>\n </div>","PeriodicalId":8450,"journal":{"name":"Artificial organs","volume":"50 10","pages":"1464-1475"},"PeriodicalIF":2.2000,"publicationDate":"2026-09-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/aor.70194","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Artificial organs","FirstCategoryId":"5","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/aor.70194","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/14 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives
The use of veno-arterial extracorporeal membrane oxygenation (V-A ECMO) has led to an increasing number of devices available on the market. Most studies pay less attention to which devices have been used; therefore, the information on performance for each device is limited. The current study is a comprehensive post-market analysis of all the major functions and adverse events rates of the Xenios console and XLung kit 230 preconfigured ECMO tubing set in post-cardiotomy applications.
Methods
This retrospective, single-center analysis, conducted between 2017 and 2021, included 47 adult patients who received V-A ECMO support using the Xenios system following cardiac surgery at our institution.
Results
The Xenios device was found to effectively improve and maintain microcirculation, organ perfusion, gas exchanges, organ function support, and stabilization of the cardiocirculatory system, while maintaining pump performance and stable body core temperature. Device-related adverse events occurred in 28% of patients but resolved in 81% of cases without leaving clinical sequelae. Fewer hemolysis events than reported in the literature were recorded. Every patient experienced at least two patient-related adverse events, but 43% had no relation to the device and 45% resolved without permanent damage.
Conclusions
The performance of the evaluated device was demonstrated in all key clinical areas, with an adequate safety profile.
期刊介绍:
Artificial Organs is the official peer reviewed journal of The International Federation for Artificial Organs (Members of the Federation are: The American Society for Artificial Internal Organs, The European Society for Artificial Organs, and The Japanese Society for Artificial Organs), The International Faculty for Artificial Organs, the International Society for Rotary Blood Pumps, The International Society for Pediatric Mechanical Cardiopulmonary Support, and the Vienna International Workshop on Functional Electrical Stimulation. Artificial Organs publishes original research articles dealing with developments in artificial organs applications and treatment modalities and their clinical applications worldwide. Membership in the Societies listed above is not a prerequisite for publication. Articles are published without charge to the author except for color figures and excess page charges as noted.