Adaptation of a Brain-Dedicated Multipinhole SPECT System for Imaging Higher-Energy Photons of Theranostic Agents—A Simulation Study

IF 3.8 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Sophia Pells;Kesava S. Kalluri;Micaehla May;Benjamin Auer;Lars R. Furenlid;Matthew A. Kupinski;Phillip H. Kuo;Robert Licho;Michael A. King
{"title":"Adaptation of a Brain-Dedicated Multipinhole SPECT System for Imaging Higher-Energy Photons of Theranostic Agents—A Simulation Study","authors":"Sophia Pells;Kesava S. Kalluri;Micaehla May;Benjamin Auer;Lars R. Furenlid;Matthew A. Kupinski;Phillip H. Kuo;Robert Licho;Michael A. King","doi":"10.1109/TRPMS.2026.3651065","DOIUrl":null,"url":null,"abstract":"Multipinhole single-photon emission computed tomography (SPECT) of the brain is capable of producing high-resolution images. However, pinhole imaging of high-energy photons remains challenging due to the tradeoff of greatly sacrificing sensitivity to limit penetration through the edges of the pinholes and system shielding. The impact of photon penetration on spatial resolution, image contrast, and lesion signal-to-noise was assessed through simulation for the brain-dedicated multipinhole system AdaptiSPECT-C. In addition, minor changes to the system design to reduce penetration were considered; these were updating the materials of the pinhole aperture edges and shutters to a higher-density tungsten alloy and including platinum in the shielding around the pinhole apertures. A significant degradation in spatial resolution, image contrast, and signal-to-noise above approximately 200 keV was found with the current system design. It was shown that adapting the pinhole and shutter material to a higher-density tungsten alloy would greatly reduce this degradation and permit multipinhole imaging with 177Lu and 111In in the human brain. Incorporating a small amount of platinum around each aperture further improved the system point-spread function. However, results indicate high-resolution imaging of 131I would require a larger modification to the system design.","PeriodicalId":46807,"journal":{"name":"IEEE Transactions on Radiation and Plasma Medical Sciences","volume":"10 6","pages":"959-968"},"PeriodicalIF":3.8000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Radiation and Plasma Medical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/11329014/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/5 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Multipinhole single-photon emission computed tomography (SPECT) of the brain is capable of producing high-resolution images. However, pinhole imaging of high-energy photons remains challenging due to the tradeoff of greatly sacrificing sensitivity to limit penetration through the edges of the pinholes and system shielding. The impact of photon penetration on spatial resolution, image contrast, and lesion signal-to-noise was assessed through simulation for the brain-dedicated multipinhole system AdaptiSPECT-C. In addition, minor changes to the system design to reduce penetration were considered; these were updating the materials of the pinhole aperture edges and shutters to a higher-density tungsten alloy and including platinum in the shielding around the pinhole apertures. A significant degradation in spatial resolution, image contrast, and signal-to-noise above approximately 200 keV was found with the current system design. It was shown that adapting the pinhole and shutter material to a higher-density tungsten alloy would greatly reduce this degradation and permit multipinhole imaging with 177Lu and 111In in the human brain. Incorporating a small amount of platinum around each aperture further improved the system point-spread function. However, results indicate high-resolution imaging of 131I would require a larger modification to the system design.
脑专用多针孔SPECT系统对高能光子成像的模拟研究
大脑的多针孔单光子发射计算机断层扫描(SPECT)能够产生高分辨率的图像。然而,高能光子的针孔成像仍然具有挑战性,因为要在很大程度上牺牲灵敏度以限制穿透针孔边缘和系统屏蔽之间的权衡。通过对大脑专用多针孔系统AdaptiSPECT-C的模拟,评估光子穿透对空间分辨率、图像对比度和病变信噪比的影响。此外,还考虑对系统设计进行微小改动以减少穿透;他们将针孔孔边缘和百叶窗的材料更新为高密度的钨合金,并在针孔孔周围的屏蔽层中加入铂。在当前的系统设计中,空间分辨率、图像对比度和信噪比在大约200 keV以上都有显著下降。研究表明,将针孔和快门材料采用高密度钨合金将大大减少这种退化,并允许在人脑中使用177Lu和111In进行多针孔成像。在每个光圈周围加入少量的铂,进一步改善了系统的点扩展功能。然而,结果表明,131I的高分辨率成像需要对系统设计进行更大的修改。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
IEEE Transactions on Radiation and Plasma Medical Sciences
IEEE Transactions on Radiation and Plasma Medical Sciences RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
8.00
自引率
18.20%
发文量
109
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书