Comparison of ELISA and MS/MS Methods for Urinary Glyphosate Analysis Reveals Overestimation and Poor Repeatability in Evaluated ELISA Methods

IF 7.3 2区 环境科学与生态学 Q1 ENGINEERING, ENVIRONMENTAL
Mari-Vorgan Louyer, Anne Gérard, Cécile Jabalot, Valérie Bouchart, Justine Lanson, Clémentine Dereumeaux, Marie Pécheux, Clémence Fillol, Sébastien Denys, Alison Connolly, Holger M. Koch, Fabien Mercier and Arthur David*, 
{"title":"Comparison of ELISA and MS/MS Methods for Urinary Glyphosate Analysis Reveals Overestimation and Poor Repeatability in Evaluated ELISA Methods","authors":"Mari-Vorgan Louyer,&nbsp;Anne Gérard,&nbsp;Cécile Jabalot,&nbsp;Valérie Bouchart,&nbsp;Justine Lanson,&nbsp;Clémentine Dereumeaux,&nbsp;Marie Pécheux,&nbsp;Clémence Fillol,&nbsp;Sébastien Denys,&nbsp;Alison Connolly,&nbsp;Holger M. Koch,&nbsp;Fabien Mercier and Arthur David*,&nbsp;","doi":"10.1021/acs.estlett.6c00374","DOIUrl":null,"url":null,"abstract":"<p >Glyphosate is one of the most widely used herbicides, and its human health effects remain debated. Risk assessment requires robust human biomonitoring (HBM) data generated using validated analytical methods. However, recent population studies employing commercial enzyme-linked immunosorbent assays (ELISA) have reported urinary glyphosate concentrations higher than those measured by mass spectrometry (MS), raising concerns about potential overestimation and data comparability. This study compares two MS/MS-based methods, including a novel ion chromatography-MS/MS approach, with two ELISA methods across external quality assessment (EQA) urine samples (n = 13), spiked urine samples (n = 3, 0.1–0.3 ng/mL), and real-world human urine samples (n = 40). The evaluated ELISA methods overestimated urinary glyphosate concentrations below 1 ng/mL in EQA and spiked samples and exhibited poor repeatability (often &gt;20%) relative to MS/MS-based methods. Analyses of real-world samples confirmed this trend, with ELISA yielding higher quantification frequencies (70–100% vs 45–63%), higher median concentrations (1.06–1.10 vs &lt;LOQ–0.11 ng/mL), respectively, and lower repeatability (&gt;30%) than MS/MS methods. These findings indicate that the evaluated ELISA methods lack sufficient reliability for quantifying urinary glyphosate at concentrations relevant to nonoccupational exposure, where matrix-related interferences may compromise analytical performance. These results support the preferential use of MS/MS-based methods for generating robust HBM data in research and regulatory contexts.</p>","PeriodicalId":37,"journal":{"name":"Environmental Science & Technology Letters Environ.","volume":"13 7","pages":"886–892"},"PeriodicalIF":7.3000,"publicationDate":"2026-07-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://pubs.acs.org/doi/pdf/10.1021/acs.estlett.6c00374","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Environmental Science & Technology Letters Environ.","FirstCategoryId":"1","ListUrlMain":"https://pubs.acs.org/doi/10.1021/acs.estlett.6c00374","RegionNum":2,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/30 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"ENGINEERING, ENVIRONMENTAL","Score":null,"Total":0}
引用次数: 0

Abstract

Glyphosate is one of the most widely used herbicides, and its human health effects remain debated. Risk assessment requires robust human biomonitoring (HBM) data generated using validated analytical methods. However, recent population studies employing commercial enzyme-linked immunosorbent assays (ELISA) have reported urinary glyphosate concentrations higher than those measured by mass spectrometry (MS), raising concerns about potential overestimation and data comparability. This study compares two MS/MS-based methods, including a novel ion chromatography-MS/MS approach, with two ELISA methods across external quality assessment (EQA) urine samples (n = 13), spiked urine samples (n = 3, 0.1–0.3 ng/mL), and real-world human urine samples (n = 40). The evaluated ELISA methods overestimated urinary glyphosate concentrations below 1 ng/mL in EQA and spiked samples and exhibited poor repeatability (often >20%) relative to MS/MS-based methods. Analyses of real-world samples confirmed this trend, with ELISA yielding higher quantification frequencies (70–100% vs 45–63%), higher median concentrations (1.06–1.10 vs <LOQ–0.11 ng/mL), respectively, and lower repeatability (>30%) than MS/MS methods. These findings indicate that the evaluated ELISA methods lack sufficient reliability for quantifying urinary glyphosate at concentrations relevant to nonoccupational exposure, where matrix-related interferences may compromise analytical performance. These results support the preferential use of MS/MS-based methods for generating robust HBM data in research and regulatory contexts.

ELISA法与MS/MS法在尿草甘膦分析中的比较表明ELISA法过高估计且重复性差
草甘膦是使用最广泛的除草剂之一,其对人类健康的影响仍存在争议。风险评估需要使用经过验证的分析方法生成可靠的人体生物监测(HBM)数据。然而,最近使用商业酶联免疫吸附试验(ELISA)的人群研究报告尿草甘膦浓度高于质谱(MS)测量的浓度,引起了对潜在高估和数据可比性的担忧。本研究比较了两种基于质谱/质谱的方法,包括一种新型离子色谱-质谱/质谱方法,以及两种ELISA方法在外部质量评估(EQA)尿液样本(n = 13)、加标尿液样本(n = 3, 0.1-0.3 ng/mL)和真实人类尿液样本(n = 40)中的应用。所评估的ELISA方法高估了EQA和加标样品中低于1 ng/mL的尿草甘膦浓度,并且与基于质谱/质谱的方法相比,重复性较差(通常为20%)。对真实样品的分析证实了这一趋势,与MS/MS方法相比,ELISA的定量频率更高(70-100% vs 45-63%),中位浓度更高(分别为1.06-1.10 vs < LOQ-0.11 ng/mL),重复性较低(>30%)。这些发现表明,所评估的ELISA方法在量化与非职业暴露相关的尿草甘膦浓度方面缺乏足够的可靠性,在非职业暴露中,基质相关的干扰可能会影响分析性能。这些结果支持在研究和监管环境中优先使用基于质谱/质谱的方法来生成可靠的HBM数据。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Environmental Science & Technology Letters Environ.
Environmental Science & Technology Letters Environ. ENGINEERING, ENVIRONMENTALENVIRONMENTAL SC-ENVIRONMENTAL SCIENCES
CiteScore
17.90
自引率
3.70%
发文量
163
期刊介绍: Environmental Science & Technology Letters serves as an international forum for brief communications on experimental or theoretical results of exceptional timeliness in all aspects of environmental science, both pure and applied. Published as soon as accepted, these communications are summarized in monthly issues. Additionally, the journal features short reviews on emerging topics in environmental science and technology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书