Veronika Tibljas, Simona Francese, Marjory Da Costa Abreu, Charlotte Lefebvre, Robert Bradshaw
{"title":"A Comparative Study of MALDI MSI Versus DESI MSI Applied to Questioned Document Examination","authors":"Veronika Tibljas, Simona Francese, Marjory Da Costa Abreu, Charlotte Lefebvre, Robert Bradshaw","doi":"10.1002/rcm.70135","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Rationale</h3>\n \n <p>In forensic investigations, questioned document examination (QDE) provides critical information for document authentication and the detection of forgeries. Conventional QDE approaches often require separate analyses of inks and of any potential fingerprints. This study compares two mass spectrometry imaging (MSI) techniques for QDE, enabling the simultaneous analysis of inks and fingerprints deposited onto paper, providing chemical and biometric information from a single sample.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Two mock forgery samples containing two ballpoint pen inks, printed ink and latent fingerprints were each analysed using either desorption electrospray ionisation mass spectrometry imaging (DESI MSI) or matrix-assisted laser desorption/ionisation mass spectrometry imaging (MALDI MSI) and the results were compared. Analyses were conducted using state-of-the-art multi reflection time-of-flight (MRT) mass spectrometry instrumentation, providing high resolving power and sub-ppm mass accuracy.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Both techniques successfully differentiated between two blue ballpoint pen inks by detecting ink components present in one pen but absent from the other (despite them being optically indistinguishable). Chemical information associated with the deposited fingerprints was also successfully detected and mapped using both DESI MSI and MALDI MSI. However, MALDI MSI outperformed DESI MSI in terms of the clarity of biometric information obtained.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Overall, the strengths and limitations of DESI MSI and MALDI MSI for QDE were evaluated, demonstrating the potential of each technique for QDE. This assessment helps identify the most suitable approach for the analysis of this type of forensic casework by showing the benefits and reliability of each technique.</p>\n </section>\n </div>","PeriodicalId":225,"journal":{"name":"Rapid Communications in Mass Spectrometry","volume":"40 19","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2026-07-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/rcm.70135","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rapid Communications in Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/rcm.70135","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
Rationale
In forensic investigations, questioned document examination (QDE) provides critical information for document authentication and the detection of forgeries. Conventional QDE approaches often require separate analyses of inks and of any potential fingerprints. This study compares two mass spectrometry imaging (MSI) techniques for QDE, enabling the simultaneous analysis of inks and fingerprints deposited onto paper, providing chemical and biometric information from a single sample.
Methods
Two mock forgery samples containing two ballpoint pen inks, printed ink and latent fingerprints were each analysed using either desorption electrospray ionisation mass spectrometry imaging (DESI MSI) or matrix-assisted laser desorption/ionisation mass spectrometry imaging (MALDI MSI) and the results were compared. Analyses were conducted using state-of-the-art multi reflection time-of-flight (MRT) mass spectrometry instrumentation, providing high resolving power and sub-ppm mass accuracy.
Results
Both techniques successfully differentiated between two blue ballpoint pen inks by detecting ink components present in one pen but absent from the other (despite them being optically indistinguishable). Chemical information associated with the deposited fingerprints was also successfully detected and mapped using both DESI MSI and MALDI MSI. However, MALDI MSI outperformed DESI MSI in terms of the clarity of biometric information obtained.
Conclusions
Overall, the strengths and limitations of DESI MSI and MALDI MSI for QDE were evaluated, demonstrating the potential of each technique for QDE. This assessment helps identify the most suitable approach for the analysis of this type of forensic casework by showing the benefits and reliability of each technique.
期刊介绍:
Rapid Communications in Mass Spectrometry is a journal whose aim is the rapid publication of original research results and ideas on all aspects of the science of gas-phase ions; it covers all the associated scientific disciplines. There is no formal limit on paper length ("rapid" is not synonymous with "brief"), but papers should be of a length that is commensurate with the importance and complexity of the results being reported. Contributions may be theoretical or practical in nature; they may deal with methods, techniques and applications, or with the interpretation of results; they may cover any area in science that depends directly on measurements made upon gaseous ions or that is associated with such measurements.