Reducing Stent-Induced Blooming Artifacts Using Virtual Monoenergetic Imaging Reconstructions in a Superficial Femoral Artery Phantom: An In-Vitro Study.

IF 2.4 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
Lisa Rutten, Lennart van de Velde, Cornelis H Slump, Jasper Martens, Kartik Jain, Michel Versluis, Michel M P J Reijnen
{"title":"Reducing Stent-Induced Blooming Artifacts Using Virtual Monoenergetic Imaging Reconstructions in a Superficial Femoral Artery Phantom: An In-Vitro Study.","authors":"Lisa Rutten, Lennart van de Velde, Cornelis H Slump, Jasper Martens, Kartik Jain, Michel Versluis, Michel M P J Reijnen","doi":"10.1007/s13239-026-00848-8","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Spectral computed tomography derived virtual monoenergetic imaging (VMI) reconstructions have been shown to reduce stent-induced blooming artifacts thus minimizing in-stent lumen underestimation. We have investigated the use of VMI reconstructions to reduce the blooming artifact for stent markers, single and double stent configurations.</p><p><strong>Methods: </strong>Two partially overlapping covered stents were deployed in a superficial femoral artery phantom. A conventional and twelve VMI reconstructions (40-100 keV and 120-200 keV with 10 and 20 keV increments, respectively) were created. Contrast-to-noise ratio, signal-to-noise ratio, attenuation difference between native and in-stent vessel lumen and full width at half maximum (FWHM) of the stent peaks were evaluated. The in-stent lumen diameter was measured, and the stent evaluated using a 5-point scale by three interventional radiologists.</p><p><strong>Results: </strong>The 80-100 keV VMI reconstructions for the single stent configuration and the 140 keV VMI reconstruction for the other configurations resulted in the smallest lumen underestimation and best quality score on blooming. The underestimation was reduced from 21.5-29.1% for the conventional reconstruction to 14.1-18.6% for the optimal VMI reconstructions. No significant differences were observed in FWHM between the conventional reconstruction and the 50-200 keV VMI reconstructions.</p><p><strong>Conclusions: </strong>VMI was found to minimize stent-induced blooming artifacts and improve in-stent lumen assessment. More stent material requires higher-energy VMI reconstructions to reduce blooming. The absence of significant differences in quantitative parameters suggests that the number of pixels affected by the blooming artifact is not reduced using VMI reconstructions.</p>","PeriodicalId":54322,"journal":{"name":"Cardiovascular Engineering and Technology","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2026-07-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Engineering and Technology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s13239-026-00848-8","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Spectral computed tomography derived virtual monoenergetic imaging (VMI) reconstructions have been shown to reduce stent-induced blooming artifacts thus minimizing in-stent lumen underestimation. We have investigated the use of VMI reconstructions to reduce the blooming artifact for stent markers, single and double stent configurations.

Methods: Two partially overlapping covered stents were deployed in a superficial femoral artery phantom. A conventional and twelve VMI reconstructions (40-100 keV and 120-200 keV with 10 and 20 keV increments, respectively) were created. Contrast-to-noise ratio, signal-to-noise ratio, attenuation difference between native and in-stent vessel lumen and full width at half maximum (FWHM) of the stent peaks were evaluated. The in-stent lumen diameter was measured, and the stent evaluated using a 5-point scale by three interventional radiologists.

Results: The 80-100 keV VMI reconstructions for the single stent configuration and the 140 keV VMI reconstruction for the other configurations resulted in the smallest lumen underestimation and best quality score on blooming. The underestimation was reduced from 21.5-29.1% for the conventional reconstruction to 14.1-18.6% for the optimal VMI reconstructions. No significant differences were observed in FWHM between the conventional reconstruction and the 50-200 keV VMI reconstructions.

Conclusions: VMI was found to minimize stent-induced blooming artifacts and improve in-stent lumen assessment. More stent material requires higher-energy VMI reconstructions to reduce blooming. The absence of significant differences in quantitative parameters suggests that the number of pixels affected by the blooming artifact is not reduced using VMI reconstructions.

在股浅动脉幻像中使用虚拟单能成像重建减少支架诱导的盛开伪影:一项体外研究。
目的:光谱计算机断层扫描衍生的虚拟单能成像(VMI)重建已被证明可以减少支架诱导的盛开伪影,从而最大限度地减少支架内管腔低估。我们已经研究了使用VMI重建来减少开花伪影支架标记,单和双支架配置。方法:在股浅动脉假体内放置两个部分重叠的覆盖支架。创建了一个常规和12个VMI重建(40-100 keV和120-200 keV,分别以10和20 keV增量)。评估对比噪声比、信噪比、原生血管腔与支架内血管腔之间的衰减差以及支架峰的半最大值全宽度(FWHM)。测量支架内管腔直径,并由三位介入放射科医生用5分制对支架进行评估。结果:单支架构型的80-100 keV VMI重建和其他构型的140 keV VMI重建可获得最小的管腔低估和最佳的盛开质量评分。最佳VMI重建的低估计从传统重建的21.5 ~ 29.1%降低到14.1 ~ 18.6%。传统重建与50-200 keV VMI重建在FWHM上无显著差异。结论:VMI可最大限度地减少支架诱导的盛开伪影,并改善支架内腔评估。更多的支架材料需要更高能量的VMI重建来减少开花。定量参数没有显著差异表明,使用VMI重建并没有减少受盛开伪影影响的像素数量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Cardiovascular Engineering and Technology
Cardiovascular Engineering and Technology Engineering-Biomedical Engineering
CiteScore
4.00
自引率
0.00%
发文量
51
期刊介绍: Cardiovascular Engineering and Technology is a journal publishing the spectrum of basic to translational research in all aspects of cardiovascular physiology and medical treatment. It is the forum for academic and industrial investigators to disseminate research that utilizes engineering principles and methods to advance fundamental knowledge and technological solutions related to the cardiovascular system. Manuscripts spanning from subcellular to systems level topics are invited, including but not limited to implantable medical devices, hemodynamics and tissue biomechanics, functional imaging, surgical devices, electrophysiology, tissue engineering and regenerative medicine, diagnostic instruments, transport and delivery of biologics, and sensors. In addition to manuscripts describing the original publication of research, manuscripts reviewing developments in these topics or their state-of-art are also invited.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书