Investigation of a Low-cost Portable Inline Dynamometer for Measuring Isometric Knee Extensor Strength.

IF 2.4 Q3 SPORT SCIENCES
International Journal of Sports Physical Therapy Pub Date : 2026-07-01 eCollection Date: 2026-01-01 DOI:10.26603/001c.163287
Amantej Sran, Elira Leake, Sydney Yott, Julie E Brandon, Geoffrey A Power, Scott C E Brandon
{"title":"Investigation of a Low-cost Portable Inline Dynamometer for Measuring Isometric Knee Extensor Strength.","authors":"Amantej Sran, Elira Leake, Sydney Yott, Julie E Brandon, Geoffrey A Power, Scott C E Brandon","doi":"10.26603/001c.163287","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and purpose: </strong>Isometric strength testing is an important tool for detecting deficits in injured populations and monitoring strength changes during rehabilitation and training. However, it is unclear whether strength measurements obtained from low-cost, inline force sensors match measurements from large, expensive, lab-based dynamometers. Therefore, the purpose of this study was to investigate agreement in maximum voluntary isometric knee extension torque measurements between a lab-based dynamometer, a low-cost wireless inline force sensor, and a wired inline force sensor.</p><p><strong>Design: </strong>Prospective cross-sectional study.</p><p><strong>Setting: </strong>Laboratory.</p><p><strong>Methods: </strong>Fourteen healthy adult participants (age >18 years) were recruited from the university community using posters, email, and social media advertisements. Peak isometric knee extension torque was measured for two repetitions at each of three knee angles (90°, 45° and 15°) using a lab-based dynamometer (Humac, CSMI Medical Solutions, USA). Then, all six trials were repeated while simultaneously recording strength using a low-cost wireless inline force sensor (Tindeq, Trondheim, Norway) and a wired inline force sensor wired (MLP, Transducer Techniques, CA, USA). Peak electromyography (EMG) magnitude was recorded for knee extensors and flexors, and electrical stimulation was used to quantify the level of voluntary activation (VA) for knee extensors. Agreement between torque measurement systems was assessed using Pearson's Correlation (r), Interclass Correlation Coefficients (ICC3,1, absolute agreement), Minimum Detectable Change, and Bland-Altman Limits of Agreement (LOA). Additionally, peak torque was compared at each knee angle using a repeated measures ANOVA, and peak EMG outputs were compared using paired t-tests.</p><p><strong>Results: </strong>Fourteen participants completed the study (3 male, 11 female; 26 ± 8.97 years; 1.71 ± 0.11 m; 66.75 ± 15.9 kg). The mean (95% CI) difference in torque between Tindeq and Humac dynamometers was 0.16 Nm/kg (-0.07 to 0.26 Nm/kg) at a knee angle of 15 degrees, -0.02 Nm/kg (-0.19 to 0.14 Nm/kg) at 45 degrees, and -0.10 Nm/kg (-0.32 to 0.13 Nm/kg) at 90 degrees. Bland-Altman LOA were -0.86 to 0.67 Nm/kg at 15 degrees, -0.60 to 0.55 Nm/kg at 45 degrees, and -0.86 to 0.67 Nm/kg at 90 degrees. Bias and LOA were similar for the MLP device. All participants achieved >90% VA. Results were more similar between Tindeq and MLP inline force sensors (r = 1.0, ICC3,1 = 0.94-1.00) than between inline force sensors and the Humac device (r <0.81, ICC3,1 = 0.37-0.80).</p><p><strong>Conclusions: </strong>Although sample size was small and only a single testing session was performed, the Tindeq device showed small measurement bias, underestimating the Humac torque by less than 0.1 Nm at 45 and 90 degrees. However, relatively large LOA indicated that individual measurement errors could exceed 25%. Low-cost wireless force sensors (e.g. Tindeq) show promise for clinical use, but individual measurements should be interpreted with caution.</p><p><strong>Level of evidence: </strong>2 (measurement agreement and validity study).</p>","PeriodicalId":47892,"journal":{"name":"International Journal of Sports Physical Therapy","volume":"21 7","pages":"644-654"},"PeriodicalIF":2.4000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13327836/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Sports Physical Therapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.26603/001c.163287","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"SPORT SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Background and purpose: Isometric strength testing is an important tool for detecting deficits in injured populations and monitoring strength changes during rehabilitation and training. However, it is unclear whether strength measurements obtained from low-cost, inline force sensors match measurements from large, expensive, lab-based dynamometers. Therefore, the purpose of this study was to investigate agreement in maximum voluntary isometric knee extension torque measurements between a lab-based dynamometer, a low-cost wireless inline force sensor, and a wired inline force sensor.

Design: Prospective cross-sectional study.

Setting: Laboratory.

Methods: Fourteen healthy adult participants (age >18 years) were recruited from the university community using posters, email, and social media advertisements. Peak isometric knee extension torque was measured for two repetitions at each of three knee angles (90°, 45° and 15°) using a lab-based dynamometer (Humac, CSMI Medical Solutions, USA). Then, all six trials were repeated while simultaneously recording strength using a low-cost wireless inline force sensor (Tindeq, Trondheim, Norway) and a wired inline force sensor wired (MLP, Transducer Techniques, CA, USA). Peak electromyography (EMG) magnitude was recorded for knee extensors and flexors, and electrical stimulation was used to quantify the level of voluntary activation (VA) for knee extensors. Agreement between torque measurement systems was assessed using Pearson's Correlation (r), Interclass Correlation Coefficients (ICC3,1, absolute agreement), Minimum Detectable Change, and Bland-Altman Limits of Agreement (LOA). Additionally, peak torque was compared at each knee angle using a repeated measures ANOVA, and peak EMG outputs were compared using paired t-tests.

Results: Fourteen participants completed the study (3 male, 11 female; 26 ± 8.97 years; 1.71 ± 0.11 m; 66.75 ± 15.9 kg). The mean (95% CI) difference in torque between Tindeq and Humac dynamometers was 0.16 Nm/kg (-0.07 to 0.26 Nm/kg) at a knee angle of 15 degrees, -0.02 Nm/kg (-0.19 to 0.14 Nm/kg) at 45 degrees, and -0.10 Nm/kg (-0.32 to 0.13 Nm/kg) at 90 degrees. Bland-Altman LOA were -0.86 to 0.67 Nm/kg at 15 degrees, -0.60 to 0.55 Nm/kg at 45 degrees, and -0.86 to 0.67 Nm/kg at 90 degrees. Bias and LOA were similar for the MLP device. All participants achieved >90% VA. Results were more similar between Tindeq and MLP inline force sensors (r = 1.0, ICC3,1 = 0.94-1.00) than between inline force sensors and the Humac device (r <0.81, ICC3,1 = 0.37-0.80).

Conclusions: Although sample size was small and only a single testing session was performed, the Tindeq device showed small measurement bias, underestimating the Humac torque by less than 0.1 Nm at 45 and 90 degrees. However, relatively large LOA indicated that individual measurement errors could exceed 25%. Low-cost wireless force sensors (e.g. Tindeq) show promise for clinical use, but individual measurements should be interpreted with caution.

Level of evidence: 2 (measurement agreement and validity study).

一种低成本便携式膝伸肌强度测量仪的研究。
背景与目的:等长肌力测试是发现受伤人群缺陷和监测康复训练过程中肌力变化的重要工具。然而,目前尚不清楚低成本内联力传感器获得的强度测量结果是否与大型、昂贵的实验室测功机测量结果相匹配。因此,本研究的目的是研究实验室测力仪、低成本无线内联力传感器和有线内联力传感器在最大自主等距膝关节伸展扭矩测量中的一致性。设计:前瞻性横断面研究。设置:实验室。方法:通过海报、电子邮件和社交媒体广告从大学社区招募14名健康成人参与者(年龄0 ~ 18岁)。使用实验室测功仪(Humac, CSMI Medical Solutions, USA)在三个膝关节角度(90°,45°和15°)中的每一个重复两次测量峰值等距膝关节伸展扭矩。然后,使用低成本无线内联力传感器(Tindeq, Trondheim, Norway)和有线内联力传感器(MLP, Transducer Techniques, CA, USA),重复所有六次试验,同时记录强度。记录膝关节伸肌和屈肌的峰值肌电图(EMG)幅度,并使用电刺激来量化膝关节伸肌的自愿激活(VA)水平。使用Pearson’s Correlation (r)、Interclass Correlation Coefficients (ICC3,1,绝对一致性)、Minimum Detectable Change和Bland-Altman limit of Agreement (LOA)来评估扭矩测量系统之间的一致性。此外,使用重复测量方差分析比较每个膝关节角度的峰值扭矩,并使用配对t检验比较峰值肌电图输出。结果:共14例受试者完成研究,男3例,女11例,年龄26±8.97岁,身高1.71±0.11米,体重66.75±15.9公斤。Tindeq和Humac测力机之间的扭矩平均值(95% CI)差异为:膝关节角为15度时为0.16 Nm/kg(-0.07至0.26 Nm/kg),膝关节角为45度时为-0.02 Nm/kg(-0.19至0.14 Nm/kg),膝关节角为90度时为-0.10 Nm/kg(-0.32至0.13 Nm/kg)。Bland-Altman LOA在15°时为-0.86 ~ 0.67 Nm/kg,在45°时为-0.60 ~ 0.55 Nm/kg,在90°时为-0.86 ~ 0.67 Nm/kg。MLP装置的偏置和LOA相似。结果显示,Tindeq和MLP内联力传感器(r = 1.0, ICC3,1 = 0.94-1.00)比内联力传感器和Humac装置更相似(r结论:虽然样本量小,只进行了一次测试,但Tindeq装置的测量偏差很小,在45度和90度处低估了小于0.1 Nm的Humac扭矩。然而,较大的LOA表明单个测量误差可能超过25%。低成本无线力传感器(如Tindeq)显示出临床应用的前景,但个体测量应谨慎解释。证据水平:2(测量一致性和效度研究)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.50
自引率
5.90%
发文量
124
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书