{"title":"Evaluation of the BT-50 Automatic Quality Control Unit on the XN9000 Hematology Analyzer","authors":"Congyu Zou, Yan Wei, Yuyin Liu, Xiaoqian Zeng","doi":"10.1111/ijlh.70184","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objective</h3>\n \n <p>This study aims to explore the application of BT50-based automatic quality control in the XN9000 hematology analyzer line, comparing its benefits and limitations with manual QC procedures, and to identify potential improvements, thereby providing a reference for the development of automated QC in clinical laboratories.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Both manual and automatic QC procedures were performed for blood cell analysis. Manual QC data (January 1, 2024, to July 31, 2024) and automatic QC data (August 1, 2024, to February 28, 2025) were collected from the Department of Clinical Laboratory, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology. SPSS 18.0 software was used for the statistical analysis. The parameters analyzed included the coefficient of variation (CV), daily QC completion time, first sample review time, and staff workload.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>(1) Compared to manual QC, no statistically significant differences were observed in the CV values of any parameters between the two methods (Mann–Whitney <i>U</i> test, <i>p</i> > 0.05 for all). (2) The median completion time for automatic QC was 42 min earlier than that for manual QC, with automatic QC finishing at 07:48:47 (IQR: 07:47:55, 07:48:46) and manual QC finishing at 08:30:34 (IQR: 08:24:17, 08:37:32). (3) With a daily work start time of 08:00, the median first sample review time was advanced by 20 min under automatic QC, with manual QC reviewed at 08:48:00 (IQR: 08:36:00, 08:50:30) and automatic QC reviewed at 08:28:00 (IQR: 08:21:30, 09:07:30). (4) The automatic QC system significantly reduced the number of manual steps, including those for storage, rewarming, mixing, and post-analysis storage.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>Implementing the automatic QC system facilitates earlier report review, reduces workload, enhances operational efficiency, and maintains CV values within acceptable laboratory standards.</p>\n </section>\n </div>","PeriodicalId":14120,"journal":{"name":"International Journal of Laboratory Hematology","volume":"48 5","pages":"1062-1068"},"PeriodicalIF":2.2000,"publicationDate":"2026-09-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Laboratory Hematology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/ijlh.70184","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/29 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
This study aims to explore the application of BT50-based automatic quality control in the XN9000 hematology analyzer line, comparing its benefits and limitations with manual QC procedures, and to identify potential improvements, thereby providing a reference for the development of automated QC in clinical laboratories.
Methods
Both manual and automatic QC procedures were performed for blood cell analysis. Manual QC data (January 1, 2024, to July 31, 2024) and automatic QC data (August 1, 2024, to February 28, 2025) were collected from the Department of Clinical Laboratory, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology. SPSS 18.0 software was used for the statistical analysis. The parameters analyzed included the coefficient of variation (CV), daily QC completion time, first sample review time, and staff workload.
Results
(1) Compared to manual QC, no statistically significant differences were observed in the CV values of any parameters between the two methods (Mann–Whitney U test, p > 0.05 for all). (2) The median completion time for automatic QC was 42 min earlier than that for manual QC, with automatic QC finishing at 07:48:47 (IQR: 07:47:55, 07:48:46) and manual QC finishing at 08:30:34 (IQR: 08:24:17, 08:37:32). (3) With a daily work start time of 08:00, the median first sample review time was advanced by 20 min under automatic QC, with manual QC reviewed at 08:48:00 (IQR: 08:36:00, 08:50:30) and automatic QC reviewed at 08:28:00 (IQR: 08:21:30, 09:07:30). (4) The automatic QC system significantly reduced the number of manual steps, including those for storage, rewarming, mixing, and post-analysis storage.
Conclusion
Implementing the automatic QC system facilitates earlier report review, reduces workload, enhances operational efficiency, and maintains CV values within acceptable laboratory standards.
期刊介绍:
The International Journal of Laboratory Hematology provides a forum for the communication of new developments, research topics and the practice of laboratory haematology.
The journal publishes invited reviews, full length original articles, and correspondence.
The International Journal of Laboratory Hematology is the official journal of the International Society for Laboratory Hematology, which addresses the following sub-disciplines: cellular analysis, flow cytometry, haemostasis and thrombosis, molecular diagnostics, haematology informatics, haemoglobinopathies, point of care testing, standards and guidelines.
The journal was launched in 2006 as the successor to Clinical and Laboratory Hematology, which was first published in 1979. An active and positive editorial policy ensures that work of a high scientific standard is reported, in order to bridge the gap between practical and academic aspects of laboratory haematology.