{"title":"Does the Apple airpods pro 2 hearing aid feature meet prescribed targets for standardized audiograms?","authors":"Haiping Huang, Taylor Dalzell, Erin M Picou","doi":"10.1080/14992027.2026.2685112","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>We aimed to record deviation of Apple AirPods Pro 2 Hearing Aid Feature (HAF) output from prescriptive targets (i.e., NAL-NL2) using two manually entered, standardized audiograms. We also aimed to record the acoustic effects of the HAF tuning controls and their utility for fine tuning HAF outputs.</p><p><strong>Design and study sample: </strong>HAF output was quantified by real-ear aided responses measured bilaterally in an acoustic manikin using 55, 65, and 75 dB SPL recorded speech inputs. Repeated measurements were made for left and right earbuds, from two sets of AirPods Pro 2 devices. Two manually entered, standardized audiograms (mild- and moderate-sloping hearing losses) were used to generate prescriptive targets. Fine tuning by a researcher towards validated prescriptive targets was explored using \"Amplification\" and \"Tone\" sliders on an iPhone.</p><p><strong>Results: </strong>For both manually-entered audiograms, under-amplification relative to targets was observed at HAF's default setting; however, match to targets could be achieved after fine tuning. The \"amplification\" slider gives ∼10 dB of broadband level changes, while \"Tone\" slider gives ∼5-10 dB of spectral tilt.</p><p><strong>Conclusions: </strong>We provide preliminary data supporting the potential of HAF for fitting patients with manually-entered mild- and moderate-sloping audiograms. HAF tuning controls could be used to meet prescriptive targets.</p>","PeriodicalId":13759,"journal":{"name":"International Journal of Audiology","volume":" ","pages":"1-8"},"PeriodicalIF":2.0000,"publicationDate":"2026-06-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Audiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/14992027.2026.2685112","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: We aimed to record deviation of Apple AirPods Pro 2 Hearing Aid Feature (HAF) output from prescriptive targets (i.e., NAL-NL2) using two manually entered, standardized audiograms. We also aimed to record the acoustic effects of the HAF tuning controls and their utility for fine tuning HAF outputs.
Design and study sample: HAF output was quantified by real-ear aided responses measured bilaterally in an acoustic manikin using 55, 65, and 75 dB SPL recorded speech inputs. Repeated measurements were made for left and right earbuds, from two sets of AirPods Pro 2 devices. Two manually entered, standardized audiograms (mild- and moderate-sloping hearing losses) were used to generate prescriptive targets. Fine tuning by a researcher towards validated prescriptive targets was explored using "Amplification" and "Tone" sliders on an iPhone.
Results: For both manually-entered audiograms, under-amplification relative to targets was observed at HAF's default setting; however, match to targets could be achieved after fine tuning. The "amplification" slider gives ∼10 dB of broadband level changes, while "Tone" slider gives ∼5-10 dB of spectral tilt.
Conclusions: We provide preliminary data supporting the potential of HAF for fitting patients with manually-entered mild- and moderate-sloping audiograms. HAF tuning controls could be used to meet prescriptive targets.
目的:我们旨在使用两个手动输入的标准化听力图记录Apple AirPods Pro 2助听器功能(HAF)输出与规定目标(即NAL-NL2)的偏差。我们还旨在记录HAF调谐控制的声学效果及其用于微调HAF输出的效用。设计和研究样本:HAF输出通过在声学人体模型中使用55,65和75db声压级录制语音输入双侧测量的实耳辅助响应来量化。对两套AirPods Pro 2设备的左右耳塞进行了重复测量。两个手动输入的标准化听力图(轻度和中度倾斜性听力损失)被用来生成规定的目标。研究人员利用iPhone上的“放大”和“音调”滑块对经过验证的规定目标进行微调。结果:对于两种手动输入的听力图,在HAF的默认设置下观察到相对于目标的放大不足;但是,经过微调后,可以实现与目标的匹配。“放大”滑块提供~ 10 dB的宽带电平变化,而“音调”滑块提供~ 5-10 dB的频谱倾斜。结论:我们提供的初步数据支持HAF用于拟合手动输入轻度和中度倾斜听力图的患者的潜力。HAF调优控制可以用来满足规定的目标。
期刊介绍:
International Journal of Audiology is committed to furthering development of a scientifically robust evidence base for audiology. The journal is published by the British Society of Audiology, the International Society of Audiology and the Nordic Audiological Society.