Nadiya Johal, Badrinathan Chandrasekaran, Paul Foley
{"title":"T Wave Oversensing in Left Bundle Branch Optimised CRT.","authors":"Nadiya Johal, Badrinathan Chandrasekaran, Paul Foley","doi":"10.15420/aer.2025.52","DOIUrl":null,"url":null,"abstract":"<p><p>Five patients who underwent left bundle branch optimised (LOT)-CRT developed T wave oversensing (TWOS) between 2 months and 2 years post implant. LOT-CRT resulted in greater improvement in left ventricular size and function compared with historical case-matched controls. No patients experienced symptoms or inappropriate therapy as a result of TWOS. This report describes the occurrence, potential mechanisms and management of TWOS in patients undergoing LOT-CRT.</p>","PeriodicalId":8412,"journal":{"name":"Arrhythmia & Electrophysiology Review","volume":"15 ","pages":"e22"},"PeriodicalIF":4.3000,"publicationDate":"2026-06-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13284782/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arrhythmia & Electrophysiology Review","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15420/aer.2025.52","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Five patients who underwent left bundle branch optimised (LOT)-CRT developed T wave oversensing (TWOS) between 2 months and 2 years post implant. LOT-CRT resulted in greater improvement in left ventricular size and function compared with historical case-matched controls. No patients experienced symptoms or inappropriate therapy as a result of TWOS. This report describes the occurrence, potential mechanisms and management of TWOS in patients undergoing LOT-CRT.