Hydroxyurea interferes with point-of-care creatinine testing in children with sickle cell anemia.

IF 1.7 Q2 MULTIDISCIPLINARY SCIENCES
Ruth Namazzi, Robert O Opoka, Charles Ssuuna, Michael J Goings, Kathryn McElhinney, Dibyadyuti Datta, Russell E Ware, Chandy C John, Andrea L Conroy
{"title":"Hydroxyurea interferes with point-of-care creatinine testing in children with sickle cell anemia.","authors":"Ruth Namazzi, Robert O Opoka, Charles Ssuuna, Michael J Goings, Kathryn McElhinney, Dibyadyuti Datta, Russell E Ware, Chandy C John, Andrea L Conroy","doi":"10.1186/s13104-026-07916-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the performance of the point-of-care i-STAT device for creatinine measurement in Ugandan children with sickle cell anemia (SCA), and to determine the clinical impact of hydroxyurea-associated assay interference.</p><p><strong>Results description: </strong>This secondary analysis was nested within a randomized clinical trial involving 248 Ugandan children with SCA. The mean age at enrollment was 32 months, and 115/248 (46.2%) initiated hydroxyurea during 12 months of follow-up as part of routine clinical care. At enrollment, creatinine values measured by i-STAT and reference laboratory testing were clinically comparable. At follow-up, children receiving hydroxyurea had significantly higher creatinine values measured by i-STAT than by reference laboratory testing (mean difference + 0.19 mg/dL; p = 0.006), with several i-STAT results falling within the clinically abnormal range. No such discrepancy was observed among children not receiving hydroxyurea. In a subset of samples, serum hydroxyurea concentrations measured by high-performance liquid chromatography accounted for 73% of the variability in i-STAT creatinine values. These findings demonstrate clinically meaningful interference of hydroxyurea with enzymatic point-of-care creatinine testing, which may result in misclassification of kidney function and inappropriate clinical decision-making in children with SCA receiving hydroxyurea.</p><p><strong>Clinicaltrials: </strong>gov identifier: NCT03528434 (date registered: May 7<sup>th</sup>, 2018).</p>","PeriodicalId":9234,"journal":{"name":"BMC Research Notes","volume":" ","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2026-06-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13523223/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"BMC Research Notes","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s13104-026-07916-1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MULTIDISCIPLINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To evaluate the performance of the point-of-care i-STAT device for creatinine measurement in Ugandan children with sickle cell anemia (SCA), and to determine the clinical impact of hydroxyurea-associated assay interference.

Results description: This secondary analysis was nested within a randomized clinical trial involving 248 Ugandan children with SCA. The mean age at enrollment was 32 months, and 115/248 (46.2%) initiated hydroxyurea during 12 months of follow-up as part of routine clinical care. At enrollment, creatinine values measured by i-STAT and reference laboratory testing were clinically comparable. At follow-up, children receiving hydroxyurea had significantly higher creatinine values measured by i-STAT than by reference laboratory testing (mean difference + 0.19 mg/dL; p = 0.006), with several i-STAT results falling within the clinically abnormal range. No such discrepancy was observed among children not receiving hydroxyurea. In a subset of samples, serum hydroxyurea concentrations measured by high-performance liquid chromatography accounted for 73% of the variability in i-STAT creatinine values. These findings demonstrate clinically meaningful interference of hydroxyurea with enzymatic point-of-care creatinine testing, which may result in misclassification of kidney function and inappropriate clinical decision-making in children with SCA receiving hydroxyurea.

Clinicaltrials: gov identifier: NCT03528434 (date registered: May 7th, 2018).

羟基脲干扰镰状细胞性贫血患儿的即时肌酸酐检测。
目的:评估乌干达镰状细胞性贫血(SCA)患儿即时护理i-STAT装置肌酸酐测量的性能,并确定羟基脲相关检测干扰的临床影响。结果描述:该二次分析嵌套在一项随机临床试验中,涉及248名患有SCA的乌干达儿童。入组时的平均年龄为32个月,115/248(46.2%)患者在12个月的随访中开始使用羟基脲作为常规临床护理的一部分。入组时,i-STAT测定的肌酐值和参考实验室检测值在临床上具有可比性。在随访中,接受羟脲治疗的儿童的i-STAT测量肌酐值明显高于参考实验室检测(平均差值+ 0.19 mg/dL; p = 0.006),有几个i-STAT结果落在临床异常范围内。在未接受羟脲治疗的儿童中没有观察到这种差异。在一部分样品中,高效液相色谱测定的血清羟脲浓度占i-STAT肌酐值变异性的73%。这些研究结果表明,羟脲对酶促护理点肌酐检测的干扰具有临床意义,这可能导致接受羟脲治疗的SCA儿童肾功能的错误分类和不适当的临床决策。Clinicaltrials: gov识别码:NCT03528434(注册日期:2018年5月7日)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
BMC Research Notes
BMC Research Notes Biochemistry, Genetics and Molecular Biology-Biochemistry, Genetics and Molecular Biology (all)
CiteScore
3.60
自引率
0.00%
发文量
363
审稿时长
15 weeks
期刊介绍: BMC Research Notes publishes scientifically valid research outputs that cannot be considered as full research or methodology articles. We support the research community across all scientific and clinical disciplines by providing an open access forum for sharing data and useful information; this includes, but is not limited to, updates to previous work, additions to established methods, short publications, null results, research proposals and data management plans.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书