F F Losev, T V Brailovskaya, F S Rusanov, I Ya Poyurovskaya, A F Sharganov
{"title":"[The effect of autoclaving sterilization on the physical-mechanical properties of samples made of photopolymer materials].","authors":"F F Losev, T V Brailovskaya, F S Rusanov, I Ya Poyurovskaya, A F Sharganov","doi":"10.17116/stomat20261050315","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>A comprehensive assessment of the effect of the standard autoclaving sterilization procedure on the physico-mechanical properties of samples made from domestic photopolymer material using stereolithographic 3D printing.</p><p><strong>Materials and methods: </strong>To form a representative sample, a batch of samples printed on a 3D stereolithographic system with a 0.05 mm pitch was used. The samples were divided into two groups: group No. 1 (<i>n</i>=10), which was not sterilized, and group No. 2 (<i>n</i>=10), which underwent a full cycle of autoclaving in the standard mode, the mechanical test procedure consisted of The values of the following parameters are: flexural strength (<i>σ</i>, MPa), flexural modulus of elasticity (Emod, MPa) and deformation (mm) to the moment of fracture.</p><p><strong>Results: </strong>A tendency to decrease bending strength by an average of 1.05% was revealed, from 140±11 MPa for non-autoclaved samples to 138±7 MPa for samples after autoclaving. The modulus of elasticity, which is the main indicator of the strength of the material, decreased by an average of 1.66% from 3242±228 to 3188±152 Mpa.</p><p><strong>Conclusion: </strong>The data obtained allow us to recommend the use of the studied domestic photopolymer material in combination with printing with a minimum step of 0.05 mm and subsequent sterilization by autoclaving in the standard mode as a technologically sound and clinically safe protocol for surgical dentistry.</p>","PeriodicalId":35887,"journal":{"name":"Stomatologiya","volume":"105 3","pages":"5-9"},"PeriodicalIF":0.0000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Stomatologiya","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.17116/stomat20261050315","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: A comprehensive assessment of the effect of the standard autoclaving sterilization procedure on the physico-mechanical properties of samples made from domestic photopolymer material using stereolithographic 3D printing.
Materials and methods: To form a representative sample, a batch of samples printed on a 3D stereolithographic system with a 0.05 mm pitch was used. The samples were divided into two groups: group No. 1 (n=10), which was not sterilized, and group No. 2 (n=10), which underwent a full cycle of autoclaving in the standard mode, the mechanical test procedure consisted of The values of the following parameters are: flexural strength (σ, MPa), flexural modulus of elasticity (Emod, MPa) and deformation (mm) to the moment of fracture.
Results: A tendency to decrease bending strength by an average of 1.05% was revealed, from 140±11 MPa for non-autoclaved samples to 138±7 MPa for samples after autoclaving. The modulus of elasticity, which is the main indicator of the strength of the material, decreased by an average of 1.66% from 3242±228 to 3188±152 Mpa.
Conclusion: The data obtained allow us to recommend the use of the studied domestic photopolymer material in combination with printing with a minimum step of 0.05 mm and subsequent sterilization by autoclaving in the standard mode as a technologically sound and clinically safe protocol for surgical dentistry.