{"title":"Noninvasive Estimation of Intracranial Pressure Using Optic-Disc Venous Pulsation Thresholds.","authors":"Thomas J Berdahl, Brian M Shafer, John P Berdahl, Andrew G Lee, Mario I Zamora, Daniel C Terveen","doi":"10.1097/WNO.0000000000002487","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Pulsations and pressure changes in the retinal vein are known to be implicated in glaucoma and are sensitive to changes in both intraocular and intracranial pressures (IOP and ICP). This study evaluates the feasibility and physiological concordance of estimating ICP from spontaneous retinal venous pulsation (SVP) thresholds using noninvasive negative-pressure (NP) goggles and comparing these estimates with reference ICPs measured through manometry during lumbar puncture (LP).</p><p><strong>Methods: </strong>This prospective, single-center pilot study evaluated the capacity of the Ocular Pressure Adjusting Pump (OPAP) to aid in estimating ICP. Adults who were scheduled for clinically indicated LP underwent noninvasive IOP modulation with sealed negative-pressure goggles (OPAP) within 24 hours before LP. An increasing magnitude of NP was applied until SVP was extinguished. Masked examiners determined IOP at the last presence of SVP and first absence of SVP at the optic disc, as measured through the OPAP goggles. Predicted ICP was defined as the midpoint between those 2 IOP values. OPAP measurements were performed twice for each eye: once in an upright seated position and again in the left lateral decubitus (LLD) position. Agreements between OPAP-predicted and LP-measured ICPs were evaluated with a Bland-Altman analysis using 80% limits of agreement (LoA).</p><p><strong>Results: </strong>Of 15 participants, 11 had visible SVPs that could be extinguished during NP application with OPAP and were included in the analysis. Four participants did not exhibit SVPs at baseline; all had LP-measured ICP exceeding IOP (mean ± SD: 17.6 ± 5.3 and 13.5 ± 1.1 mm Hg, respectively; P < 0.001). For eyes with measurable SVP thresholds, the mean OPAP-predicted ICP was 12.7 ± 3.2 mm Hg in an upright seated position and 17.5 ± 3.1 mm Hg in a lateral position. In the upright analysis, mean bias (predicted-reference ICP) was -0.17 mm Hg with 80% LoA ±6.19 Hg. In the lateral analysis, mean bias was +4.66 mm Hg with 80% LoA ±3.56 mm Hg.</p><p><strong>Conclusions: </strong>In this pilot feasibility study, SVP-threshold-derived ICP estimates demonstrated minimal bias when OPAP measurements were taken with patients in an upright position and systematic overestimation when OPAP measurements were taken with patients in the LLD position. Conversely, OPAP's ICP estimations demonstrated better consistency (narrower 80% LoAs) with measurements taken in the LLD as compared with upright measurements (wider 80% LoAs). These pilot study findings are exploratory in nature and are not intended to guide individual patient care. Larger studies with standardized positioning and contemporaneous reference measurements are required to define accuracy, identify sources of bias, and demonstrate clinical utility.</p>","PeriodicalId":16485,"journal":{"name":"Journal of Neuro-Ophthalmology","volume":" ","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2026-06-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Neuro-Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/WNO.0000000000002487","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Pulsations and pressure changes in the retinal vein are known to be implicated in glaucoma and are sensitive to changes in both intraocular and intracranial pressures (IOP and ICP). This study evaluates the feasibility and physiological concordance of estimating ICP from spontaneous retinal venous pulsation (SVP) thresholds using noninvasive negative-pressure (NP) goggles and comparing these estimates with reference ICPs measured through manometry during lumbar puncture (LP).
Methods: This prospective, single-center pilot study evaluated the capacity of the Ocular Pressure Adjusting Pump (OPAP) to aid in estimating ICP. Adults who were scheduled for clinically indicated LP underwent noninvasive IOP modulation with sealed negative-pressure goggles (OPAP) within 24 hours before LP. An increasing magnitude of NP was applied until SVP was extinguished. Masked examiners determined IOP at the last presence of SVP and first absence of SVP at the optic disc, as measured through the OPAP goggles. Predicted ICP was defined as the midpoint between those 2 IOP values. OPAP measurements were performed twice for each eye: once in an upright seated position and again in the left lateral decubitus (LLD) position. Agreements between OPAP-predicted and LP-measured ICPs were evaluated with a Bland-Altman analysis using 80% limits of agreement (LoA).
Results: Of 15 participants, 11 had visible SVPs that could be extinguished during NP application with OPAP and were included in the analysis. Four participants did not exhibit SVPs at baseline; all had LP-measured ICP exceeding IOP (mean ± SD: 17.6 ± 5.3 and 13.5 ± 1.1 mm Hg, respectively; P < 0.001). For eyes with measurable SVP thresholds, the mean OPAP-predicted ICP was 12.7 ± 3.2 mm Hg in an upright seated position and 17.5 ± 3.1 mm Hg in a lateral position. In the upright analysis, mean bias (predicted-reference ICP) was -0.17 mm Hg with 80% LoA ±6.19 Hg. In the lateral analysis, mean bias was +4.66 mm Hg with 80% LoA ±3.56 mm Hg.
Conclusions: In this pilot feasibility study, SVP-threshold-derived ICP estimates demonstrated minimal bias when OPAP measurements were taken with patients in an upright position and systematic overestimation when OPAP measurements were taken with patients in the LLD position. Conversely, OPAP's ICP estimations demonstrated better consistency (narrower 80% LoAs) with measurements taken in the LLD as compared with upright measurements (wider 80% LoAs). These pilot study findings are exploratory in nature and are not intended to guide individual patient care. Larger studies with standardized positioning and contemporaneous reference measurements are required to define accuracy, identify sources of bias, and demonstrate clinical utility.
背景:已知视网膜静脉的脉动和压力变化与青光眼有关,并且对眼内和颅内压(IOP和ICP)的变化很敏感。本研究评估了使用无创负压(NP)护目镜从自发视网膜静脉脉动(SVP)阈值估计ICP的可行性和生理一致性,并将这些估计值与腰椎穿刺(LP)测压测量的参考ICP进行比较。方法:本前瞻性、单中心试点研究评估眼压调节泵(OPAP)的容量,以帮助估计ICP。计划进行临床指征LP的成人在LP前24小时内使用密封负压护目镜(OPAP)进行无创IOP调节。施加越来越大的NP,直到SVP熄灭。蒙面检查人员通过OPAP护目镜测量视盘上最后一次SVP存在和第一次SVP不存在时的IOP。预测ICP定义为这两个IOP值之间的中点。每只眼睛进行两次OPAP测量:一次在直立坐姿,一次在左侧卧位(LLD)。opap预测和lp测量的icp之间的一致性采用Bland-Altman分析,使用80%的一致性限(LoA)进行评估。结果:在15名参与者中,11名有可见的svp,可以在使用OPAP进行NP应用时被消除,并被纳入分析。4名参与者在基线时没有表现出svp;所有患者的lp测量ICP均超过IOP(平均±SD分别为17.6±5.3和13.5±1.1 mm Hg, P < 0.001)。对于具有可测量SVP阈值的眼睛,opap预测的平均ICP在直立坐姿为12.7±3.2 mm Hg,在侧卧位为17.5±3.1 mm Hg。在直立分析中,平均偏倚(预测参考ICP)为-0.17 mm Hg, 80% LoA±6.19 Hg。在横向分析中,平均偏倚为+4.66 mm Hg, 80% LoA±3.56 mm Hg。结论:在这项试点可行性研究中,svp阈值推导的ICP估计显示,当患者处于直立位置时进行OPAP测量时偏差最小,而当患者处于LLD位置时进行OPAP测量时系统高估。相反,与垂直测量(更宽的80% LoAs)相比,OPAP的ICP估计与LLD测量显示出更好的一致性(更窄的80% LoAs)。这些初步研究结果在本质上是探索性的,并不打算指导个别病人的护理。需要更大规模的标准化定位和同期参考测量的研究来确定准确性,确定偏差来源,并证明临床实用性。