A non-inferiority trial comparing manual and automated high-level disinfection for reprocessing of transabdominal and transvaginal ultrasound probes.

IF 5.2 2区 医学 Q1 INFECTIOUS DISEASES
Hannah Siwe, Maryse Van Landegem, Shari De Vriese, Astrid Luypaert, Merima Hecimovic, Steven Weyers, Isabelle Dehaene, Erik Van Laecke, Philip Meuleman, Piet Cools
{"title":"A non-inferiority trial comparing manual and automated high-level disinfection for reprocessing of transabdominal and transvaginal ultrasound probes.","authors":"Hannah Siwe, Maryse Van Landegem, Shari De Vriese, Astrid Luypaert, Merima Hecimovic, Steven Weyers, Isabelle Dehaene, Erik Van Laecke, Philip Meuleman, Piet Cools","doi":"10.1186/s13756-026-01777-w","DOIUrl":null,"url":null,"abstract":"<p><p>Manual chemical methods are commonly used for ultrasound probe disinfection but are known to be operator dependent and subject to variability. Automated ultraviolet-C (UV-C) light-emitting diode (LED) disinfection represents a chemical and mercury free alternative, however, clinical evidence comparing UV-C LED to conventional disinfectants for high-level disinfection of ultrasound probes remains limited. In this single-centre non-inferiority study conducted in the departments of gynaecology, obstetrics and urology of a Belgian university hospital, automated UV-C LED disinfection was compared to manual chlorine dioxide disinfection for the reprocessing of transabdominal and transvaginal ultrasound probes. Following clinical use, probes were decontaminated by healthcare practitioners alternating between both reprocessing methods. Samples were collected after treatment and processed for microbial analysis to assess the effectiveness of the decontamination process. A total of 296 probes were included, comprising 86 transabdominal and 210 transvaginal probes. UV-C LED disinfection achieved effectiveness rates of 92.86% for transabdominal probes and 97.17% for transvaginal probes, compared with 79.49% and 93.27%, respectively for chlorine dioxide disinfection. Positive samples following UV-C disinfection yielded single colonies whereas chlorine dioxide treated probes showed greater variability in colony counts and species diversity. Although the total reprocessing time for UV-C disinfection was longer, manual handling times were comparable between both methods. Here, we demonstrated that UV-C LED disinfection was non-inferior to chlorine dioxide disinfection for reprocessing of transabdominal and transvaginal ultrasound probes under routine conditions.</p>","PeriodicalId":7950,"journal":{"name":"Antimicrobial Resistance and Infection Control","volume":" ","pages":""},"PeriodicalIF":5.2000,"publicationDate":"2026-06-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13479550/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Antimicrobial Resistance and Infection Control","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s13756-026-01777-w","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
引用次数: 0

Abstract

Manual chemical methods are commonly used for ultrasound probe disinfection but are known to be operator dependent and subject to variability. Automated ultraviolet-C (UV-C) light-emitting diode (LED) disinfection represents a chemical and mercury free alternative, however, clinical evidence comparing UV-C LED to conventional disinfectants for high-level disinfection of ultrasound probes remains limited. In this single-centre non-inferiority study conducted in the departments of gynaecology, obstetrics and urology of a Belgian university hospital, automated UV-C LED disinfection was compared to manual chlorine dioxide disinfection for the reprocessing of transabdominal and transvaginal ultrasound probes. Following clinical use, probes were decontaminated by healthcare practitioners alternating between both reprocessing methods. Samples were collected after treatment and processed for microbial analysis to assess the effectiveness of the decontamination process. A total of 296 probes were included, comprising 86 transabdominal and 210 transvaginal probes. UV-C LED disinfection achieved effectiveness rates of 92.86% for transabdominal probes and 97.17% for transvaginal probes, compared with 79.49% and 93.27%, respectively for chlorine dioxide disinfection. Positive samples following UV-C disinfection yielded single colonies whereas chlorine dioxide treated probes showed greater variability in colony counts and species diversity. Although the total reprocessing time for UV-C disinfection was longer, manual handling times were comparable between both methods. Here, we demonstrated that UV-C LED disinfection was non-inferior to chlorine dioxide disinfection for reprocessing of transabdominal and transvaginal ultrasound probes under routine conditions.

一项非劣效性试验,比较人工和自动高水平消毒对经腹和经阴道超声探头的再处理。
人工化学方法通常用于超声探头消毒,但已知是操作员依赖和受可变性。自动紫外- c (UV-C)发光二极管(LED)消毒代表了一种无化学和无汞的替代方案,然而,将UV-C LED与传统消毒剂进行高水平超声探头消毒的临床证据仍然有限。在比利时一所大学医院妇科、产科和泌尿科进行的这项单中心非劣效性研究中,对经腹和经阴道超声探头再处理的自动UV-C LED消毒与人工二氧化氯消毒进行了比较。临床使用后,由医疗保健从业者交替使用两种再处理方法对探针进行净化。处理后收集样品并进行微生物分析,以评估去污过程的有效性。共纳入296个探针,其中经腹探针86个,经阴道探针210个。UV-C LED消毒对经腹探针和阴道探针的有效率分别为92.86%和97.17%,而二氧化氯消毒的有效率分别为79.49%和93.27%。UV-C消毒后的阳性样品产生单个菌落,而二氧化氯处理的探针在菌落计数和物种多样性方面表现出更大的变化。虽然UV-C消毒的总再处理时间较长,但两种方法的人工处理时间相当。在这里,我们证明了在常规条件下,UV-C LED消毒不逊于二氧化氯消毒对经腹和经阴道超声探头的再处理。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Antimicrobial Resistance and Infection Control
Antimicrobial Resistance and Infection Control PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH -INFECTIOUS DISEASES
CiteScore
9.70
自引率
3.60%
发文量
140
审稿时长
13 weeks
期刊介绍: Antimicrobial Resistance and Infection Control is a global forum for all those working on the prevention, diagnostic and treatment of health-care associated infections and antimicrobial resistance development in all health-care settings. The journal covers a broad spectrum of preeminent practices and best available data to the top interventional and translational research, and innovative developments in the field of infection control.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书