Histological Evaluation of In Vivo Dose Delivery in Respiration-gated Small Animal Radiotherapy.

IF 2.4 3区 医学 Q2 BIOLOGY
Qiwei Wu, Cheng Zheng, Zihao Liu, Xiaogang Yuan, Yunwen Huang, Ning Zhao, Hui Liu, Yidong Yang
{"title":"Histological Evaluation of In Vivo Dose Delivery in Respiration-gated Small Animal Radiotherapy.","authors":"Qiwei Wu, Cheng Zheng, Zihao Liu, Xiaogang Yuan, Yunwen Huang, Ning Zhao, Hui Liu, Yidong Yang","doi":"10.1667/RADE-25-00189.1","DOIUrl":null,"url":null,"abstract":"<p><p>Preclinical small-animal radiotherapy is essential for advancing radiobiological research. However, respiratory motion causes significant geometric uncertainty and dose blurring, reducing treatment accuracy. This study aims to develop and validate an integrated respiration-gated irradiation technique for mice and assess its effectiveness in decreasing motion-induced targeting errors. Respiratory motion was tracked in real time using an optical displacement sensor, and a mechanical shutter was employed to gate the X-ray beam. Both ungated and gated irradiation were delivered to the liver and lung (n = 8 per group). Ex vivo γH2AX immunofluorescence staining and an image analysis pipeline were used to evaluate the radiation field. The histological dimension in the irradiated tissue was adjusted using tissue-specific shrinkage factors to account for histology-related shrinkage, and the corrected measurements were compared between the ungated and gated groups. γH2AX analysis showed that respiratory gating significantly reduced the histological dimension along the superior-inferior axis in both the liver (ungated: 3.00 ± 0.85 mm vs. gated: 1.75 ± 0.24 mm; P < 0.01) and the lung (ungated: 2.55 ± 0.64 mm vs. gated: 1.33 ± 0.19 mm; P < 0.001). This resulted in a 1.25 mm reduction in the motion-related geometric margin for the liver and a 1.22 mm reduction for the lung. This study provides direct biological evidence that respiratory gating effectively reduces motion-induced dose blurring. The proposed gating method can significantly reduce treatment margins caused by motion, thereby enhancing targeting accuracy and sparing normal tissue.</p>","PeriodicalId":20903,"journal":{"name":"Radiation research","volume":"205 6","pages":"607-614"},"PeriodicalIF":2.4000,"publicationDate":"2026-04-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1667/RADE-25-00189.1","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"BIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Preclinical small-animal radiotherapy is essential for advancing radiobiological research. However, respiratory motion causes significant geometric uncertainty and dose blurring, reducing treatment accuracy. This study aims to develop and validate an integrated respiration-gated irradiation technique for mice and assess its effectiveness in decreasing motion-induced targeting errors. Respiratory motion was tracked in real time using an optical displacement sensor, and a mechanical shutter was employed to gate the X-ray beam. Both ungated and gated irradiation were delivered to the liver and lung (n = 8 per group). Ex vivo γH2AX immunofluorescence staining and an image analysis pipeline were used to evaluate the radiation field. The histological dimension in the irradiated tissue was adjusted using tissue-specific shrinkage factors to account for histology-related shrinkage, and the corrected measurements were compared between the ungated and gated groups. γH2AX analysis showed that respiratory gating significantly reduced the histological dimension along the superior-inferior axis in both the liver (ungated: 3.00 ± 0.85 mm vs. gated: 1.75 ± 0.24 mm; P < 0.01) and the lung (ungated: 2.55 ± 0.64 mm vs. gated: 1.33 ± 0.19 mm; P < 0.001). This resulted in a 1.25 mm reduction in the motion-related geometric margin for the liver and a 1.22 mm reduction for the lung. This study provides direct biological evidence that respiratory gating effectively reduces motion-induced dose blurring. The proposed gating method can significantly reduce treatment margins caused by motion, thereby enhancing targeting accuracy and sparing normal tissue.

呼吸门控小动物放射治疗中体内给药的组织学评价。
临床前小动物放射治疗对推进放射生物学研究至关重要。然而,呼吸运动引起显著的几何不确定性和剂量模糊,降低了治疗的准确性。本研究旨在开发和验证一种综合呼吸门控照射小鼠技术,并评估其在减少运动诱导的靶向错误方面的有效性。使用光学位移传感器实时跟踪呼吸运动,并使用机械百叶窗限制x射线束。对肝脏和肺部进行非门控和门控照射(每组n = 8)。采用体外γ - h2ax免疫荧光染色和图像分析管道评价辐射场。使用组织特异性收缩因子调整辐照组织的组织学尺寸,以解释组织学相关的收缩,并比较未门控组和门控组之间的校正测量结果。γ - h2ax分析显示,呼吸门控显著降低了肝脏(未门控:3.00±0.85 mm vs.门控:1.75±0.24 mm, P < 0.01)和肺(未门控:2.55±0.64 mm vs.门控:1.33±0.19 mm, P < 0.001)上下轴的组织学尺寸。这导致肝脏运动相关几何边缘减少1.25 mm,肺部减少1.22 mm。这项研究提供了直接的生物学证据,呼吸门控有效地减少了运动引起的剂量模糊。所提出的门控方法可以显著减少运动引起的治疗余量,从而提高靶向精度并保留正常组织。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiation research
Radiation research 医学-核医学
CiteScore
5.10
自引率
8.80%
发文量
179
审稿时长
1 months
期刊介绍: Radiation Research publishes original articles dealing with radiation effects and related subjects in the areas of physics, chemistry, biology and medicine, including epidemiology and translational research. The term radiation is used in its broadest sense and includes specifically ionizing radiation and ultraviolet, visible and infrared light as well as microwaves, ultrasound and heat. Effects may be physical, chemical or biological. Related subjects include (but are not limited to) dosimetry methods and instrumentation, isotope techniques and studies with chemical agents contributing to the understanding of radiation effects.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书