Severe hyponatremia associated with unintended infusion of an unmixed neutral-pH peritoneal dialysis solution in an elderly patient with cognitive impairment.
{"title":"Severe hyponatremia associated with unintended infusion of an unmixed neutral-pH peritoneal dialysis solution in an elderly patient with cognitive impairment.","authors":"Minoru Ito, Ikuto Masakane, Kiyotaka Yabuki","doi":"10.1177/08968608261458138","DOIUrl":null,"url":null,"abstract":"<p><p>Neutral-pH, low-glucose degradation product peritoneal dialysis (PD) solutions are widely used because of their favorable biocompatibility. These solutions are typically packaged in dual-chamber bags and must be mixed immediately before use. Although the mixing procedure is simple, improper preparation may occur, particularly in patients with cognitive impairment, potentially resulting in unintended electrolyte exposure. We report a case of severe hyponatremia with altered mental status associated with inadvertent infusion of an unmixed neutral-pH PD solution in an elderly patient undergoing automated peritoneal dialysis (APD). A man in his eighth decade with diabetic nephropathy and progressive cognitive decline presented with somnolence and severe hyponatremia after a dual-chamber PD bag with an unopened septum was connected. Based on the clinical course and the finding that a PD solution bag with an unopened inter-chamber septum had been connected, hyponatremia due to exposure to an unmixed dialysate compartment was diagnosed. Management consisted of immediate discontinuation of the improperly prepared dialysate and correction using appropriately mixed PD solutions, without administration of hypertonic saline. Serum sodium levels improved gradually without neurological complications, accompanied by recovery of mental status. This case highlights a rare but serious complication related to improper preparation of dual-chamber PD solutions. Increased awareness of this risk and careful monitoring of vulnerable patients, particularly those with cognitive impairment, are essential to ensure safe home PD therapy.</p>","PeriodicalId":19969,"journal":{"name":"Peritoneal Dialysis International","volume":" ","pages":"8968608261458138"},"PeriodicalIF":1.8000,"publicationDate":"2026-06-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Peritoneal Dialysis International","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/08968608261458138","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"UROLOGY & NEPHROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Neutral-pH, low-glucose degradation product peritoneal dialysis (PD) solutions are widely used because of their favorable biocompatibility. These solutions are typically packaged in dual-chamber bags and must be mixed immediately before use. Although the mixing procedure is simple, improper preparation may occur, particularly in patients with cognitive impairment, potentially resulting in unintended electrolyte exposure. We report a case of severe hyponatremia with altered mental status associated with inadvertent infusion of an unmixed neutral-pH PD solution in an elderly patient undergoing automated peritoneal dialysis (APD). A man in his eighth decade with diabetic nephropathy and progressive cognitive decline presented with somnolence and severe hyponatremia after a dual-chamber PD bag with an unopened septum was connected. Based on the clinical course and the finding that a PD solution bag with an unopened inter-chamber septum had been connected, hyponatremia due to exposure to an unmixed dialysate compartment was diagnosed. Management consisted of immediate discontinuation of the improperly prepared dialysate and correction using appropriately mixed PD solutions, without administration of hypertonic saline. Serum sodium levels improved gradually without neurological complications, accompanied by recovery of mental status. This case highlights a rare but serious complication related to improper preparation of dual-chamber PD solutions. Increased awareness of this risk and careful monitoring of vulnerable patients, particularly those with cognitive impairment, are essential to ensure safe home PD therapy.
期刊介绍:
Peritoneal Dialysis International (PDI) is an international publication dedicated to peritoneal dialysis. PDI welcomes original contributions dealing with all aspects of peritoneal dialysis from scientists working in the peritoneal dialysis field around the world.
Peritoneal Dialysis International is included in Index Medicus and indexed in Current Contents/Clinical Practice, the Science Citation Index, and Excerpta Medica (Nephrology/Urology Core Journal). It is also abstracted and indexed in Chemical Abstracts (CA), as well as being indexed in Embase as a priority journal.