Towards a phantom-based quantitative assessment of dose reduction and image-quality trade-offs in half- and full-rotation dental CBCT protocols: a thermoluminescent dosimetry approach with clinical insight.
Hakan Amasya, Osman Günay, Fahrettin Fatih Kesmezacar, Emine Tuna Akdoğan, Duygu Tunçman Kayaokay, Songül Çavdar Karaçam, Nami Yeyin, Mustafa Demir, Kaan Orhan, Ghada ALMisned, H O Tekin
{"title":"Towards a phantom-based quantitative assessment of dose reduction and image-quality trade-offs in half- and full-rotation dental CBCT protocols: a thermoluminescent dosimetry approach with clinical insight.","authors":"Hakan Amasya, Osman Günay, Fahrettin Fatih Kesmezacar, Emine Tuna Akdoğan, Duygu Tunçman Kayaokay, Songül Çavdar Karaçam, Nami Yeyin, Mustafa Demir, Kaan Orhan, Ghada ALMisned, H O Tekin","doi":"10.1007/s00411-026-01228-5","DOIUrl":null,"url":null,"abstract":"<p><p>This study aimed to compare full- (360°) and half-rotation (180°) dental cone-beam computed tomography (CBCT) protocols in terms of effective dose (ED) and quantitative image quality for the maxilla and mandible regions. An Alderson Radiation Therapy phantom was imaged using a Hyperion X9 Pro (Cefla, Imola) CBCT device with a 10 × 6 cm field of view. Both anatomical regions were acquired using full- and half-rotation protocols. A total of 67 thermoluminescence dosimeters were positioned in the phantom for ED calculations. Quantitative image-quality assessment was performed using axial slices extracted from each volume, and signal-to-noise ratio (SNRs) and contrast-to-noise ratio (CNRs) were calculated. The EDs were 406.33 µSv (full-rotation) and 208.29 µSv (half-rotation) for the maxilla, and 248.94 µSv (full-rotation) and 73.63 µSv (half-rotation) for the mandible. These doses corresponded to dose reductions of 48.74% in the maxilla and 70.42% in the mandible. SNRs and CNRs decreased by 36.69% and 34.74% in the maxilla, and by 49.95% and 50.28% in the mandible, respectively. Dose reduction was more pronounced in the mandible than in the maxilla, and this was accompanied by a greater loss in SNR and CNR. It is concluded that protocol selection should be guided by diagnostic requirements and adherence to the \"As Low as Diagnostically Acceptable being Indication-oriented and Patient-specific (ALADAIP) principle.</p>","PeriodicalId":21002,"journal":{"name":"Radiation and Environmental Biophysics","volume":" ","pages":"863-871"},"PeriodicalIF":2.4000,"publicationDate":"2026-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation and Environmental Biophysics","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1007/s00411-026-01228-5","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/8 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"BIOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
This study aimed to compare full- (360°) and half-rotation (180°) dental cone-beam computed tomography (CBCT) protocols in terms of effective dose (ED) and quantitative image quality for the maxilla and mandible regions. An Alderson Radiation Therapy phantom was imaged using a Hyperion X9 Pro (Cefla, Imola) CBCT device with a 10 × 6 cm field of view. Both anatomical regions were acquired using full- and half-rotation protocols. A total of 67 thermoluminescence dosimeters were positioned in the phantom for ED calculations. Quantitative image-quality assessment was performed using axial slices extracted from each volume, and signal-to-noise ratio (SNRs) and contrast-to-noise ratio (CNRs) were calculated. The EDs were 406.33 µSv (full-rotation) and 208.29 µSv (half-rotation) for the maxilla, and 248.94 µSv (full-rotation) and 73.63 µSv (half-rotation) for the mandible. These doses corresponded to dose reductions of 48.74% in the maxilla and 70.42% in the mandible. SNRs and CNRs decreased by 36.69% and 34.74% in the maxilla, and by 49.95% and 50.28% in the mandible, respectively. Dose reduction was more pronounced in the mandible than in the maxilla, and this was accompanied by a greater loss in SNR and CNR. It is concluded that protocol selection should be guided by diagnostic requirements and adherence to the "As Low as Diagnostically Acceptable being Indication-oriented and Patient-specific (ALADAIP) principle.
期刊介绍:
This journal is devoted to fundamental and applied issues in radiation research and biophysics. The topics may include:
Biophysics of ionizing radiation: radiation physics and chemistry, radiation dosimetry, radiobiology, radioecology, biophysical foundations of medical applications of radiation, and radiation protection.
Biological effects of radiation: experimental or theoretical work on molecular or cellular effects; relevance of biological effects for risk assessment; biological effects of medical applications of radiation; relevance of radiation for biosphere and in space; modelling of ecosystems; modelling of transport processes of substances in biotic systems.
Risk assessment: epidemiological studies of cancer and non-cancer effects; quantification of risk including exposures to radiation and confounding factors
Contributions to these topics may include theoretical-mathematical and experimental material, as well as description of new techniques relevant for the study of these issues. They can range from complex radiobiological phenomena to issues in health physics and environmental protection.