Shaping the Flow: Impact of Blade Shape on Performance and Hemocompatibility in Rotodynamic Blood Pumps

IF 2.2 3区 医学 Q3 ENGINEERING, BIOMEDICAL
Artificial organs Pub Date : 2026-09-29 Epub Date: 2026-06-07 DOI:10.1111/aor.70178
Benjamin Torner, Simon Klocker, Niccolò Crescenti, Aldo J. Suria, Harald Reithofer, Michael Hübler, Daniel Zimpfer, Marcus Granegger
{"title":"Shaping the Flow: Impact of Blade Shape on Performance and Hemocompatibility in Rotodynamic Blood Pumps","authors":"Benjamin Torner,&nbsp;Simon Klocker,&nbsp;Niccolò Crescenti,&nbsp;Aldo J. Suria,&nbsp;Harald Reithofer,&nbsp;Michael Hübler,&nbsp;Daniel Zimpfer,&nbsp;Marcus Granegger","doi":"10.1111/aor.70178","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objectives</h3>\n \n <p>Rotodynamic blood pumps (RBPs) must operate reliably across a wide range of clinical conditions, including off-design operating regimes. Although blade shape is known to influence off-design performance in conventional turbopumps, its impact on hydraulic performance and hemocompatibility in RBPs has not been systematically investigated. This study combines in silico and in vitro analyses to assess the effects of leading-edge geometry and blade thickness on performance (pressure head-flow rate characteristics, efficiencies) and hemocompatibility (damage index, DI) across multiple RBPs and operating conditions.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Three representative RBPs, a cavopulmonary assist device, the CentriMag RBP, and a novel ventricular assist device, were investigated. Semi-circular, elliptical, and wedge-shaped leading edges were implemented, whereas three blade thickness configurations were analyzed.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Variations in leading-edge geometry resulted in minor differences in performance (&lt; 3 percent in efficiency) and hemocompatibility (&lt; 3% DI) across all devices and operating conditions. In contrast, blade thickness had a pronounced effect on performance under overload off-design operation, with thinner blades improving efficiency by up to 22%, while exhibiting only limited changes in hemocompatibility-related metrics (&lt; 7% DI). The observed efficiency differences were primarily attributed to decreased friction losses caused by less channel blockage with thinner impeller blades.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Overall, impeller blade shape plays a secondary role in hemocompatibility within the investigated pump types and operating conditions. This finding suggests that leading-edge optimization may be deprioritized relative to other hemocompatibility-critical geometry parameters in RBP design. Blade thickness, however, may lead to unfavorable channel blockage effects and therefore needs to be optimized with respect to manufacturing constraints and hydraulic performance.</p>\n </section>\n </div>","PeriodicalId":8450,"journal":{"name":"Artificial organs","volume":"50 10","pages":"1360-1371"},"PeriodicalIF":2.2000,"publicationDate":"2026-09-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/aor.70178","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Artificial organs","FirstCategoryId":"5","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/aor.70178","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/7 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives

Rotodynamic blood pumps (RBPs) must operate reliably across a wide range of clinical conditions, including off-design operating regimes. Although blade shape is known to influence off-design performance in conventional turbopumps, its impact on hydraulic performance and hemocompatibility in RBPs has not been systematically investigated. This study combines in silico and in vitro analyses to assess the effects of leading-edge geometry and blade thickness on performance (pressure head-flow rate characteristics, efficiencies) and hemocompatibility (damage index, DI) across multiple RBPs and operating conditions.

Methods

Three representative RBPs, a cavopulmonary assist device, the CentriMag RBP, and a novel ventricular assist device, were investigated. Semi-circular, elliptical, and wedge-shaped leading edges were implemented, whereas three blade thickness configurations were analyzed.

Results

Variations in leading-edge geometry resulted in minor differences in performance (< 3 percent in efficiency) and hemocompatibility (< 3% DI) across all devices and operating conditions. In contrast, blade thickness had a pronounced effect on performance under overload off-design operation, with thinner blades improving efficiency by up to 22%, while exhibiting only limited changes in hemocompatibility-related metrics (< 7% DI). The observed efficiency differences were primarily attributed to decreased friction losses caused by less channel blockage with thinner impeller blades.

Conclusions

Overall, impeller blade shape plays a secondary role in hemocompatibility within the investigated pump types and operating conditions. This finding suggests that leading-edge optimization may be deprioritized relative to other hemocompatibility-critical geometry parameters in RBP design. Blade thickness, however, may lead to unfavorable channel blockage effects and therefore needs to be optimized with respect to manufacturing constraints and hydraulic performance.

塑造流动:叶片形状对旋转动力血泵性能和血液相容性的影响。
目的:旋转动力血泵(rbp)必须在广泛的临床条件下可靠地运行,包括非设计操作方案。虽然已知叶片形状会影响常规涡轮泵的非设计性能,但其对rbp水力性能和血液相容性的影响尚未系统研究。该研究结合了计算机和体外分析,评估了前缘几何形状和叶片厚度对多个rbp和操作条件下的性能(压力头流速率特性,效率)和血液相容性(损伤指数,DI)的影响。方法:对三种具有代表性的RBP进行了研究,一种是腔室肺辅助装置,一种是CentriMag RBP,一种是新型心室辅助装置。设计了半圆形、椭圆形和楔形前缘,分析了三种叶片厚度结构。结果:前缘几何形状的变化导致性能的微小差异(结论:总体而言,叶轮叶片形状在所研究的泵类型和操作条件下对血液相容性起次要作用。这一发现表明,在RBP设计中,相对于其他血液相容性关键几何参数,前沿优化可能被优先考虑。然而,叶片厚度可能会导致不利的通道阻塞效应,因此需要根据制造限制和水力性能进行优化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Artificial organs
Artificial organs 工程技术-工程:生物医学
CiteScore
4.30
自引率
12.50%
发文量
303
审稿时长
4-8 weeks
期刊介绍: Artificial Organs is the official peer reviewed journal of The International Federation for Artificial Organs (Members of the Federation are: The American Society for Artificial Internal Organs, The European Society for Artificial Organs, and The Japanese Society for Artificial Organs), The International Faculty for Artificial Organs, the International Society for Rotary Blood Pumps, The International Society for Pediatric Mechanical Cardiopulmonary Support, and the Vienna International Workshop on Functional Electrical Stimulation. Artificial Organs publishes original research articles dealing with developments in artificial organs applications and treatment modalities and their clinical applications worldwide. Membership in the Societies listed above is not a prerequisite for publication. Articles are published without charge to the author except for color figures and excess page charges as noted.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书