A case report of radiotherapy for head and neck cancer in a pregnant woman: dosimetric optimization and in vivo dosimetry

IF 2 Q1 Nursing
Axel Martin , David Thibouw , Laurent Delcoudert , Léone Aubignac , Karine Peignaux , Christelle Huet , Magali Edouard , Jérémy Baude , Igor Bessieres
{"title":"A case report of radiotherapy for head and neck cancer in a pregnant woman: dosimetric optimization and in vivo dosimetry","authors":"Axel Martin ,&nbsp;David Thibouw ,&nbsp;Laurent Delcoudert ,&nbsp;Léone Aubignac ,&nbsp;Karine Peignaux ,&nbsp;Christelle Huet ,&nbsp;Magali Edouard ,&nbsp;Jérémy Baude ,&nbsp;Igor Bessieres","doi":"10.1016/j.tipsro.2026.100387","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Radiotherapy during pregnancy is a rare and ethically complex requiring careful management of fetal exposure. Although peripheral doses from head and neck radiotherapy are typically low, fetal irradiation carries a potential developmental or carcinogenic risk. Accurate dose estimation and optimization are essential when radiotherapy cannot be postponed.</div></div><div><h3>Case presentation</h3><div>A 27-year-old woman, 26 weeks pregnant, required adjuvant radiotherapy for a left parotid adenoid cystic carcinoma (pT4a N3b, R1). To minimize fetal exposure, a one-month pre-treatment period was dedicated to dosimetric optimization. Three clinically equivalent VMAT plans were compared using different collimator angles (15°/345°, 80°/30°, and 80°/100°). Out-of-field dose measurements were assessed using a calibrated ionization chamber positioned in a slab phantom at varying depths and distances simulating fetal position and progression of pregnancy. All plan showed low out-of-field dose, below 100 mGy. The plan with 80°/100° collimator angles (Popt2) consistently showed the lowest out-of-field dose and was therefore selected for treatment. A lead apron was added during treatment. <em>In vivo</em> dosimetry was performed using radiophotoluminescent (RPL) detectors placed weekly under the lead apron at the uterine fundus projection. Over 10 monitored fractions, single-session doses averaged 1.25 mGy, and cumulative five-session doses reached 5.9–6.7 mGy. Total extrapolated fetal dose was approximately 41.0 mGy, closely matching the phantom-based estimate of 31.1 mGy.</div></div><div><h3>Conclusions</h3><div>This unprecedented case in our institution demonstrates the feasibility of safely delivering radiotherapy during pregnancy through multidisciplinary coordination and meticulous dose management. Phantom-based optimization and <em>in vivo</em> dosimetry confirmed that the fetal dose remained below the 100 mGy threshold set by international guidelines for deterministic effects. The patient completed treatment without complications and delivered a healthy infant at term. This experience provided the foundation for a standardized dosimetric approach within our institution for future cases involving pregnancy and radiotherapy.</div></div><div><h3>Background</h3><div>Radiotherapy during pregnancy is rare and presents significant ethical and technical challenges dure to fetal radiation exposure. The nature of these effects depends on both the dose level and the stage of pregnancy <span><span>[1]</span></span>. Although postponing treatment until after delivery is preferable whenever possible, radiotherapy can be administered during pregnancy under certain conditions. The International Commission on Radiological Protection (ICRP) has established that, for an absorbed dose below 100 <!--> <!-->mGy, no risk to fetal development is expected <span><span>[2]</span></span> and medical termination of pregnancy is not justified regarding radiation risks. Nevertheless, fetal dose optimization remains essential, as stochastic risks may exist even at low exposure levels.</div></div>","PeriodicalId":36328,"journal":{"name":"Technical Innovations and Patient Support in Radiation Oncology","volume":"38 ","pages":"Article 100387"},"PeriodicalIF":2.0000,"publicationDate":"2026-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Technical Innovations and Patient Support in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405632426000144","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/2/17 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"Nursing","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Radiotherapy during pregnancy is a rare and ethically complex requiring careful management of fetal exposure. Although peripheral doses from head and neck radiotherapy are typically low, fetal irradiation carries a potential developmental or carcinogenic risk. Accurate dose estimation and optimization are essential when radiotherapy cannot be postponed.

Case presentation

A 27-year-old woman, 26 weeks pregnant, required adjuvant radiotherapy for a left parotid adenoid cystic carcinoma (pT4a N3b, R1). To minimize fetal exposure, a one-month pre-treatment period was dedicated to dosimetric optimization. Three clinically equivalent VMAT plans were compared using different collimator angles (15°/345°, 80°/30°, and 80°/100°). Out-of-field dose measurements were assessed using a calibrated ionization chamber positioned in a slab phantom at varying depths and distances simulating fetal position and progression of pregnancy. All plan showed low out-of-field dose, below 100 mGy. The plan with 80°/100° collimator angles (Popt2) consistently showed the lowest out-of-field dose and was therefore selected for treatment. A lead apron was added during treatment. In vivo dosimetry was performed using radiophotoluminescent (RPL) detectors placed weekly under the lead apron at the uterine fundus projection. Over 10 monitored fractions, single-session doses averaged 1.25 mGy, and cumulative five-session doses reached 5.9–6.7 mGy. Total extrapolated fetal dose was approximately 41.0 mGy, closely matching the phantom-based estimate of 31.1 mGy.

Conclusions

This unprecedented case in our institution demonstrates the feasibility of safely delivering radiotherapy during pregnancy through multidisciplinary coordination and meticulous dose management. Phantom-based optimization and in vivo dosimetry confirmed that the fetal dose remained below the 100 mGy threshold set by international guidelines for deterministic effects. The patient completed treatment without complications and delivered a healthy infant at term. This experience provided the foundation for a standardized dosimetric approach within our institution for future cases involving pregnancy and radiotherapy.

Background

Radiotherapy during pregnancy is rare and presents significant ethical and technical challenges dure to fetal radiation exposure. The nature of these effects depends on both the dose level and the stage of pregnancy [1]. Although postponing treatment until after delivery is preferable whenever possible, radiotherapy can be administered during pregnancy under certain conditions. The International Commission on Radiological Protection (ICRP) has established that, for an absorbed dose below 100  mGy, no risk to fetal development is expected [2] and medical termination of pregnancy is not justified regarding radiation risks. Nevertheless, fetal dose optimization remains essential, as stochastic risks may exist even at low exposure levels.
孕妇头颈癌放疗一例报告:剂量学优化与体内剂量学。
背景:妊娠期放射治疗是一种罕见且伦理复杂的疾病,需要对胎儿放射治疗进行谨慎管理。虽然头颈部放射治疗的外周剂量通常较低,但胎儿放射治疗具有潜在的发育或致癌风险。当放疗不能推迟时,准确的剂量估计和优化是必不可少的。病例介绍:27岁女性,怀孕26 周,左侧腮腺腺样囊性癌(pT4a N3b, R1)需要辅助放疗。为了最大限度地减少胎儿暴露,一个月的前治疗期是专门用于剂量学优化。采用不同的准直角度(15°/345°、80°/30°和80°/100°)比较了三种临床等效VMAT方案。外场剂量测量评估使用校准电离室放置在平板幻影在不同的深度和距离模拟胎儿的位置和妊娠进展。所有方案均显示低场外剂量,低于100 mGy。80°/100°准直角(Popt2)方案始终显示出最低的场外剂量,因此被选为治疗方案。治疗期间加铅围裙。使用放射性光致发光(RPL)探测器进行体内剂量测定,每周一次放置在子宫底投影处的铅围裙下。在10个监测馏分中,单次剂量平均为1.25 毫格瑞,五次累积剂量达到5.9-6.7 毫格瑞。外推的胎儿总剂量约为41.0 mGy,与基于幻影的31.1 mGy的估计非常吻合。结论:本病例在我院是前所未有的,通过多学科协调和细致的剂量管理,证明了妊娠期安全放射治疗的可行性。基于幻影的优化和体内剂量测定证实,胎儿剂量仍低于国际确定效应指南所设定的100 毫戈瑞阈值。患者完成治疗后无并发症,足月生下健康婴儿。这一经验为我们机构未来涉及妊娠和放疗的病例的标准化剂量学方法奠定了基础。背景:妊娠期间的放射治疗是罕见的,并且在胎儿辐射暴露期间提出了重大的伦理和技术挑战。这些影响的性质取决于剂量水平和妊娠阶段。虽然尽可能将治疗推迟到分娩后,但在某些情况下,可以在怀孕期间进行放射治疗。国际放射防护委员会(ICRP)已经确定,对于低于100 毫戈瑞的吸收剂量,预计不会对胎儿发育造成任何风险,并且就辐射风险而言,没有理由进行医疗终止妊娠。然而,胎儿剂量优化仍然是必要的,因为即使在低暴露水平下也可能存在随机风险。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.10
自引率
0.00%
发文量
48
审稿时长
67 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书