Improvement of Lung Nodule Volumetric Accuracy with Photon-counting Computed Tomography Over Energy-integrating Computed Tomography in Low-dose Screening: A Phantom Study.
IF 7.9 1区 医学Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Joost F Hop, Marcel J W Greuter, Mark van Dijk, Geertruida H de Bock, Rozemarijn Vliegenthart
{"title":"Improvement of Lung Nodule Volumetric Accuracy with Photon-counting Computed Tomography Over Energy-integrating Computed Tomography in Low-dose Screening: A Phantom Study.","authors":"Joost F Hop, Marcel J W Greuter, Mark van Dijk, Geertruida H de Bock, Rozemarijn Vliegenthart","doi":"10.1097/RLI.0000000000001299","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To compare lung nodule volumetric accuracy and precision between photon-counting detector (PCD) computed tomography (CT) and energy-integrating detector (EID) CT using low-dose lung cancer screening protocols, and to optimize reconstruction parameters for lung nodule volumetry.</p><p><strong>Materials and methods: </strong>An anthropomorphic chest phantom with 12 artificial lung nodules of varying size, shape, and radiodensity was scanned using EID-CT and PCD-CT with reference and optimized low-dose lung cancer screening protocols. PCD-CT reconstruction parameters (slice thickness, matrix size, kernel, iterative reconstruction, and virtual monoenergetic imaging energy) were varied. Each protocol was scanned 3 times with nodule repositioning. Nodule volumes were independently measured semiautomatically by 2 observers. Interobserver agreement and test-retest reliability were assessed using the intraclass correlation coefficient (ICC) and Bland-Altman plots. Volumetric accuracy and precision were calculated relative to ground-truth volumes. Volumetric accuracy was compared between PCD-CT and EID-CT using one-way analysis of variance, and across PCD-CT reconstructions using univariable linear regression. Volumetric precision was assessed based on the SD of mean volume differences. Noise was compared across scanners and reconstructions using one-way analysis of variance.</p><p><strong>Results: </strong>A total of 1224 nodule measurements demonstrated excellent volumetric interobserver agreement (ICC: 0.99) and test-retest reliability (ICCs of 0.96 for both observers). Volumetric accuracy improved from -16.6% and -15.2% with the reference and optimized EID-CT protocols to -12.5% and -10.2% with the reference and optimized PCD-CT protocols (P < 0.05). Volumetric precision remained comparable between reference and optimized EID-CT (6.3 and 8.2 mm3) and PCD-CT (6.4 and 6.2 mm3) protocols. On PCD-CT, ultra-thin slices (0.2 mm) and an ultra-sharp kernel (Qr76) worsened volumetric accuracy by 2.9% and 2.8%, respectively (P < 0.05). Image noise was lower on PCD-CT than on EID-CT (P < 0.05) and varied significantly across reconstruction settings on PCD-CT.</p><p><strong>Conclusions: </strong>PCD-CT improved lung nodule volumetric accuracy and reduced volume underestimation by up to 6% compared with EID-CT using low-dose screening protocols, while maintaining similar volumetric precision. Ultra-thin slices and an ultra-sharp kernel worsened volumetric accuracy. By reducing volume underestimation, PCD-CT may shift a larger proportion of nodules to higher baseline risk categories, potentially increasing the number of screening participants requiring clinical referral or short-term follow-up CT.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":""},"PeriodicalIF":7.9000,"publicationDate":"2026-06-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001299","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: To compare lung nodule volumetric accuracy and precision between photon-counting detector (PCD) computed tomography (CT) and energy-integrating detector (EID) CT using low-dose lung cancer screening protocols, and to optimize reconstruction parameters for lung nodule volumetry.
Materials and methods: An anthropomorphic chest phantom with 12 artificial lung nodules of varying size, shape, and radiodensity was scanned using EID-CT and PCD-CT with reference and optimized low-dose lung cancer screening protocols. PCD-CT reconstruction parameters (slice thickness, matrix size, kernel, iterative reconstruction, and virtual monoenergetic imaging energy) were varied. Each protocol was scanned 3 times with nodule repositioning. Nodule volumes were independently measured semiautomatically by 2 observers. Interobserver agreement and test-retest reliability were assessed using the intraclass correlation coefficient (ICC) and Bland-Altman plots. Volumetric accuracy and precision were calculated relative to ground-truth volumes. Volumetric accuracy was compared between PCD-CT and EID-CT using one-way analysis of variance, and across PCD-CT reconstructions using univariable linear regression. Volumetric precision was assessed based on the SD of mean volume differences. Noise was compared across scanners and reconstructions using one-way analysis of variance.
Results: A total of 1224 nodule measurements demonstrated excellent volumetric interobserver agreement (ICC: 0.99) and test-retest reliability (ICCs of 0.96 for both observers). Volumetric accuracy improved from -16.6% and -15.2% with the reference and optimized EID-CT protocols to -12.5% and -10.2% with the reference and optimized PCD-CT protocols (P < 0.05). Volumetric precision remained comparable between reference and optimized EID-CT (6.3 and 8.2 mm3) and PCD-CT (6.4 and 6.2 mm3) protocols. On PCD-CT, ultra-thin slices (0.2 mm) and an ultra-sharp kernel (Qr76) worsened volumetric accuracy by 2.9% and 2.8%, respectively (P < 0.05). Image noise was lower on PCD-CT than on EID-CT (P < 0.05) and varied significantly across reconstruction settings on PCD-CT.
Conclusions: PCD-CT improved lung nodule volumetric accuracy and reduced volume underestimation by up to 6% compared with EID-CT using low-dose screening protocols, while maintaining similar volumetric precision. Ultra-thin slices and an ultra-sharp kernel worsened volumetric accuracy. By reducing volume underestimation, PCD-CT may shift a larger proportion of nodules to higher baseline risk categories, potentially increasing the number of screening participants requiring clinical referral or short-term follow-up CT.
期刊介绍:
Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.