Improvement of Lung Nodule Volumetric Accuracy with Photon-counting Computed Tomography Over Energy-integrating Computed Tomography in Low-dose Screening: A Phantom Study.

IF 7.9 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Joost F Hop, Marcel J W Greuter, Mark van Dijk, Geertruida H de Bock, Rozemarijn Vliegenthart
{"title":"Improvement of Lung Nodule Volumetric Accuracy with Photon-counting Computed Tomography Over Energy-integrating Computed Tomography in Low-dose Screening: A Phantom Study.","authors":"Joost F Hop, Marcel J W Greuter, Mark van Dijk, Geertruida H de Bock, Rozemarijn Vliegenthart","doi":"10.1097/RLI.0000000000001299","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To compare lung nodule volumetric accuracy and precision between photon-counting detector (PCD) computed tomography (CT) and energy-integrating detector (EID) CT using low-dose lung cancer screening protocols, and to optimize reconstruction parameters for lung nodule volumetry.</p><p><strong>Materials and methods: </strong>An anthropomorphic chest phantom with 12 artificial lung nodules of varying size, shape, and radiodensity was scanned using EID-CT and PCD-CT with reference and optimized low-dose lung cancer screening protocols. PCD-CT reconstruction parameters (slice thickness, matrix size, kernel, iterative reconstruction, and virtual monoenergetic imaging energy) were varied. Each protocol was scanned 3 times with nodule repositioning. Nodule volumes were independently measured semiautomatically by 2 observers. Interobserver agreement and test-retest reliability were assessed using the intraclass correlation coefficient (ICC) and Bland-Altman plots. Volumetric accuracy and precision were calculated relative to ground-truth volumes. Volumetric accuracy was compared between PCD-CT and EID-CT using one-way analysis of variance, and across PCD-CT reconstructions using univariable linear regression. Volumetric precision was assessed based on the SD of mean volume differences. Noise was compared across scanners and reconstructions using one-way analysis of variance.</p><p><strong>Results: </strong>A total of 1224 nodule measurements demonstrated excellent volumetric interobserver agreement (ICC: 0.99) and test-retest reliability (ICCs of 0.96 for both observers). Volumetric accuracy improved from -16.6% and -15.2% with the reference and optimized EID-CT protocols to -12.5% and -10.2% with the reference and optimized PCD-CT protocols (P < 0.05). Volumetric precision remained comparable between reference and optimized EID-CT (6.3 and 8.2 mm3) and PCD-CT (6.4 and 6.2 mm3) protocols. On PCD-CT, ultra-thin slices (0.2 mm) and an ultra-sharp kernel (Qr76) worsened volumetric accuracy by 2.9% and 2.8%, respectively (P < 0.05). Image noise was lower on PCD-CT than on EID-CT (P < 0.05) and varied significantly across reconstruction settings on PCD-CT.</p><p><strong>Conclusions: </strong>PCD-CT improved lung nodule volumetric accuracy and reduced volume underestimation by up to 6% compared with EID-CT using low-dose screening protocols, while maintaining similar volumetric precision. Ultra-thin slices and an ultra-sharp kernel worsened volumetric accuracy. By reducing volume underestimation, PCD-CT may shift a larger proportion of nodules to higher baseline risk categories, potentially increasing the number of screening participants requiring clinical referral or short-term follow-up CT.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":""},"PeriodicalIF":7.9000,"publicationDate":"2026-06-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001299","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: To compare lung nodule volumetric accuracy and precision between photon-counting detector (PCD) computed tomography (CT) and energy-integrating detector (EID) CT using low-dose lung cancer screening protocols, and to optimize reconstruction parameters for lung nodule volumetry.

Materials and methods: An anthropomorphic chest phantom with 12 artificial lung nodules of varying size, shape, and radiodensity was scanned using EID-CT and PCD-CT with reference and optimized low-dose lung cancer screening protocols. PCD-CT reconstruction parameters (slice thickness, matrix size, kernel, iterative reconstruction, and virtual monoenergetic imaging energy) were varied. Each protocol was scanned 3 times with nodule repositioning. Nodule volumes were independently measured semiautomatically by 2 observers. Interobserver agreement and test-retest reliability were assessed using the intraclass correlation coefficient (ICC) and Bland-Altman plots. Volumetric accuracy and precision were calculated relative to ground-truth volumes. Volumetric accuracy was compared between PCD-CT and EID-CT using one-way analysis of variance, and across PCD-CT reconstructions using univariable linear regression. Volumetric precision was assessed based on the SD of mean volume differences. Noise was compared across scanners and reconstructions using one-way analysis of variance.

Results: A total of 1224 nodule measurements demonstrated excellent volumetric interobserver agreement (ICC: 0.99) and test-retest reliability (ICCs of 0.96 for both observers). Volumetric accuracy improved from -16.6% and -15.2% with the reference and optimized EID-CT protocols to -12.5% and -10.2% with the reference and optimized PCD-CT protocols (P < 0.05). Volumetric precision remained comparable between reference and optimized EID-CT (6.3 and 8.2 mm3) and PCD-CT (6.4 and 6.2 mm3) protocols. On PCD-CT, ultra-thin slices (0.2 mm) and an ultra-sharp kernel (Qr76) worsened volumetric accuracy by 2.9% and 2.8%, respectively (P < 0.05). Image noise was lower on PCD-CT than on EID-CT (P < 0.05) and varied significantly across reconstruction settings on PCD-CT.

Conclusions: PCD-CT improved lung nodule volumetric accuracy and reduced volume underestimation by up to 6% compared with EID-CT using low-dose screening protocols, while maintaining similar volumetric precision. Ultra-thin slices and an ultra-sharp kernel worsened volumetric accuracy. By reducing volume underestimation, PCD-CT may shift a larger proportion of nodules to higher baseline risk categories, potentially increasing the number of screening participants requiring clinical referral or short-term follow-up CT.

在低剂量筛查中,光子计数计算机断层扫描比能量积分计算机断层扫描提高肺结节体积准确性:一项幻象研究。
目的:比较低剂量肺癌筛查方案下光子计数检测器(PCD) CT和能量积分检测器(EID) CT肺结节体积测量的准确性和精密度,并优化肺结节体积测量的重建参数。材料与方法:采用EID-CT和PCD-CT扫描拟人胸假体,植入12个不同大小、形状和放射密度的人工肺结节,参照优化的低剂量肺癌筛查方案。不同的PCD-CT重建参数(切片厚度、矩阵大小、核、迭代重建和虚拟单能成像能量)不同。每个方案扫描3次,重新定位结节。由2名观察员独立半自动测量结节体积。使用类内相关系数(ICC)和Bland-Altman图评估观察者间一致性和重测信度。相对于地面真值体积计算体积精度和精度。采用单向方差分析比较了PCD-CT和EID-CT的体积精度,并采用单变量线性回归比较了PCD-CT重建的体积精度。根据平均体积差的标准差评估体积精度。使用单向方差分析比较扫描仪和重建的噪声。结果:总共1224个结节测量显示出良好的体积观察者间一致性(ICC: 0.99)和测试-重测信度(ICC为0.96)。体积准确度由参考和优化的EID-CT方案的-16.6%和-15.2%提高到参考和优化的PCD-CT方案的-12.5%和-10.2% (P < 0.05)。参考和优化的EID-CT(6.3和8.2 mm3)和PCD-CT(6.4和6.2 mm3)方案的体积精度保持相当。在PCD-CT上,超薄切片(0.2 mm)和超锋利核(Qr76)分别使体积准确性下降2.9%和2.8% (P < 0.05)。图像噪声在PCD-CT上明显低于EID-CT (P < 0.05),且在不同的PCD-CT重建设置上差异显著。结论:与使用低剂量筛查方案的EID-CT相比,PCD-CT提高了肺结节的体积准确性,减少了高达6%的体积低估,同时保持了相似的体积精度。超薄的切片和超锋利的内核恶化了体积的准确性。通过减少体积低估,PCD-CT可能将更大比例的结节转移到基线风险更高的类别,潜在地增加了需要临床转诊或短期随访CT的筛查参与者的数量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Investigative Radiology
Investigative Radiology 医学-核医学
CiteScore
15.10
自引率
16.40%
发文量
188
审稿时长
4-8 weeks
期刊介绍: Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书