Bench testing a contact sensing tracheal tube for monitoring the cuff-trachea interface.

Tamaralayefa B Agbiki, Sandor Erdody, Ricardo Correia, Sergiy Korposh, David W Hewson, Andrew M Norris, Barrie R Hayes-Gill, Stephen P Morgan
{"title":"Bench testing a contact sensing tracheal tube for monitoring the cuff-trachea interface.","authors":"Tamaralayefa B Agbiki, Sandor Erdody, Ricardo Correia, Sergiy Korposh, David W Hewson, Andrew M Norris, Barrie R Hayes-Gill, Stephen P Morgan","doi":"10.1186/s42490-026-00113-y","DOIUrl":null,"url":null,"abstract":"<p><p>Cuffed tracheal tubes enable a secure airway during mechanical ventilation; however, improper cuff inflation can lead to complications such as inspiratory gas leaks, pulmonary ingress of orogastric secretions, or tracheal injury. This study explores a sensor-based approach to optimise cuff inflation using fibre Bragg grating (FBG) optical sensors embedded within the cuff to detect contact with the tracheal wall. Testing was conducted on multiple trachea models, including cylindrical and bio-inspired models, as well as ex vivo porcine trachea samples, to assess cuff-trachea contact. Seal performance was evaluated at both standard inflation pressures and contact-guided inflation pressures. The contact-sensing tracheal tube successfully detected cuff-trachea contact in all models and ex vivo samples tested. Seal tests showed that larger diameter tracheal models exhibited lower leakage rates at contact-guided inflation pressures compared with standard inflation, indicating improved sealing. In contrast, smaller diameter models showed higher leakage rates at contact-guided pressures due to cuff folding, which formed leakage pathways. These results demonstrate that contact sensing improved cuff performance in larger tracheal models by enhancing sealing effectiveness. The contact-sensing tracheal tube differentiated tracheal model sizes, indicating potential for personalised cuff inflation to accommodate variations in tracheal size and shape. Future work will focus on further sensor miniaturisation and integration within lower-volume cuffs to enable effective contact sensing across a wider range of tracheal geometries and support translation to in vivo use.</p>","PeriodicalId":72425,"journal":{"name":"BMC biomedical engineering","volume":"8 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2026-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13224478/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"BMC biomedical engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s42490-026-00113-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Cuffed tracheal tubes enable a secure airway during mechanical ventilation; however, improper cuff inflation can lead to complications such as inspiratory gas leaks, pulmonary ingress of orogastric secretions, or tracheal injury. This study explores a sensor-based approach to optimise cuff inflation using fibre Bragg grating (FBG) optical sensors embedded within the cuff to detect contact with the tracheal wall. Testing was conducted on multiple trachea models, including cylindrical and bio-inspired models, as well as ex vivo porcine trachea samples, to assess cuff-trachea contact. Seal performance was evaluated at both standard inflation pressures and contact-guided inflation pressures. The contact-sensing tracheal tube successfully detected cuff-trachea contact in all models and ex vivo samples tested. Seal tests showed that larger diameter tracheal models exhibited lower leakage rates at contact-guided inflation pressures compared with standard inflation, indicating improved sealing. In contrast, smaller diameter models showed higher leakage rates at contact-guided pressures due to cuff folding, which formed leakage pathways. These results demonstrate that contact sensing improved cuff performance in larger tracheal models by enhancing sealing effectiveness. The contact-sensing tracheal tube differentiated tracheal model sizes, indicating potential for personalised cuff inflation to accommodate variations in tracheal size and shape. Future work will focus on further sensor miniaturisation and integration within lower-volume cuffs to enable effective contact sensing across a wider range of tracheal geometries and support translation to in vivo use.

台架测试用于监测袖带-气管界面的接触式传感气管管。
套管气管在机械通气过程中确保气道安全;然而,不适当的袖带充气会导致并发症,如吸入气体泄漏、口胃分泌物进入肺部或气管损伤。本研究探索了一种基于传感器的方法,利用嵌入袖带的光纤布拉格光栅(FBG)光学传感器来检测与气管壁的接触,从而优化袖带充气。在多种气管模型上进行了测试,包括圆柱形模型和仿生模型,以及离体猪气管样本,以评估袖套与气管的接触。在标准膨胀压力和接触导向膨胀压力下,对密封性能进行了评估。接触式感应气管管在所有模型和离体样本中成功检测到套管与气管的接触。密封测试表明,与标准充气相比,大直径气管模型在接触导向充气压力下的泄漏率更低,表明密封性得到改善。相比之下,较小直径的模型在接触引导压力下由于袖带折叠形成泄漏通道而显示出更高的泄漏率。这些结果表明,接触传感通过提高密封效果改善了较大气管模型中的袖带性能。接触式感应气管管可区分气管模型尺寸,表明个性化袖口充气的潜力,以适应气管尺寸和形状的变化。未来的工作将集中在传感器的进一步小型化和集成上,以实现在更大范围的气管几何形状上有效的接触传感,并支持转化为体内使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
审稿时长
19 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书