Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020).
Minghui Gong, Yang Yu, Yingying Chen, Jing Li, Jiayu Su, Yannan Huang, Kai Zhong, Jing Zhang, Yuta Teruyama, Ying Zhang, Yinong Jiang, Wei Song
{"title":"Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020).","authors":"Minghui Gong, Yang Yu, Yingying Chen, Jing Li, Jiayu Su, Yannan Huang, Kai Zhong, Jing Zhang, Yuta Teruyama, Ying Zhang, Yinong Jiang, Wei Song","doi":"10.1080/08037051.2026.2682038","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Validation of the accuracy of the Omron HEM-790XT1 automatic oscillometric upper-arm blood pressure (BP) monitor in the general adults according to the international AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020) standard.</p><p><strong>Methods: </strong>Eligible adult participants meeting the protocol-specific criteria for age, gender, BP range and arm circumference distribution were recruited. The BP was measured by two trained observers using the same-arm sequential method. The test device measured BP during the inflation phase and was compared with readings from a standard mercury sphygmomanometer. The test device was used with its standard cuff, validated for an arm circumference range of 22-32 cm.</p><p><strong>Results: </strong>A total of 85 participants were included in the final analysis after excluding 2 participants. The average age was 35 ± 13.7 years, with males accounting for 35.29% and females for 64.71%. According to Criterion 1 (all individual paired readings), the mean difference ± standard deviation (SD) between the test and reference device was -1.09 ± 4.94 mmHg for systolic blood pressure (SBP) and -2.49 ± 4.97 mmHg for diastolic blood pressure (DBP), meeting the requirement of ≤5 ± 8 mmHg. According to Criterion 2 (average per participant), the mean difference ± SD between the test and reference device was -1.09 ± 3.81 mmHg for SBP and -2.49 ± 4.12 mmHg for DBP. Both the SDs for SBP and DBP were below the derived thresholds of 6.86 and 6.47 mmHg, respectively. The Omron HEM-790XT1 automatic oscillometric BP monitor demonstrated satisfactory consistency with the mercury sphygmomanometer in both SBP and DBP measurements.</p><p><strong>Conclusions: </strong>The Omron HEM-790XT1 automatic oscillometric upper-arm BP monitor satisfied all validation criteria of the AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020) standard and can be recommended for accurate BP measurement in adults within the validated arm circumference range.</p>","PeriodicalId":9000,"journal":{"name":"Blood Pressure","volume":" ","pages":"2682038"},"PeriodicalIF":2.3000,"publicationDate":"2026-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/08037051.2026.2682038","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/17 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Validation of the accuracy of the Omron HEM-790XT1 automatic oscillometric upper-arm blood pressure (BP) monitor in the general adults according to the international AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020) standard.
Methods: Eligible adult participants meeting the protocol-specific criteria for age, gender, BP range and arm circumference distribution were recruited. The BP was measured by two trained observers using the same-arm sequential method. The test device measured BP during the inflation phase and was compared with readings from a standard mercury sphygmomanometer. The test device was used with its standard cuff, validated for an arm circumference range of 22-32 cm.
Results: A total of 85 participants were included in the final analysis after excluding 2 participants. The average age was 35 ± 13.7 years, with males accounting for 35.29% and females for 64.71%. According to Criterion 1 (all individual paired readings), the mean difference ± standard deviation (SD) between the test and reference device was -1.09 ± 4.94 mmHg for systolic blood pressure (SBP) and -2.49 ± 4.97 mmHg for diastolic blood pressure (DBP), meeting the requirement of ≤5 ± 8 mmHg. According to Criterion 2 (average per participant), the mean difference ± SD between the test and reference device was -1.09 ± 3.81 mmHg for SBP and -2.49 ± 4.12 mmHg for DBP. Both the SDs for SBP and DBP were below the derived thresholds of 6.86 and 6.47 mmHg, respectively. The Omron HEM-790XT1 automatic oscillometric BP monitor demonstrated satisfactory consistency with the mercury sphygmomanometer in both SBP and DBP measurements.
Conclusions: The Omron HEM-790XT1 automatic oscillometric upper-arm BP monitor satisfied all validation criteria of the AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020) standard and can be recommended for accurate BP measurement in adults within the validated arm circumference range.
Blood PressureMedicine-Cardiology and Cardiovascular Medicine
CiteScore
3.20
自引率
5.60%
发文量
41
期刊介绍:
For outstanding coverage of the latest advances in hypertension research, turn to Blood Pressure, a primary source for authoritative and timely information on all aspects of hypertension research and management.
Features include:
• Physiology and pathophysiology of blood pressure regulation
• Primary and secondary hypertension
• Cerebrovascular and cardiovascular complications of hypertension
• Detection, treatment and follow-up of hypertension
• Non pharmacological and pharmacological management
• Large outcome trials in hypertension.