{"title":"Super-resolution deep learning reconstruction achieves low-dose high-resolution imaging in pediatric cardiac computed tomography.","authors":"Haruka Obinata, Kiyohiro Takigiku, Kosuke Yonehara, Yuma Shibuya, Yohei Akazawa, Kohta Takei, Daisuke Miyazawa, Taro Sakashita","doi":"10.1007/s00247-026-06668-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Cardiac computed tomography (CT) is widely utilized in pediatric cardiology, but minimizing radiation exposure is essential. Recently, super-resolution deep learning reconstruction (SR-DLR) has emerged as a potential advancement over conventional deep learning reconstruction (C-DLR).</p><p><strong>Objective: </strong>To evaluate the potential of SR-DLR for further radiation dose reduction in pediatric cardiac CT through a comparison with C-DLR.</p><p><strong>Materials and methods: </strong>The phantom-based study assessed noise power spectrum and spatial resolution. For the clinical study, we compared nine C-DLR images (conventional radiation dose) and 11 SR-DLR images (lower radiation dose). Evaluation metrics included noise levels, contrast-to-noise ratio, and edge rise distance and edge rise slope.</p><p><strong>Results: </strong>In the phantom-based study, SR-DLR (250 mA) displayed equivalent noise characteristics as C-DLR (800 mA) while achieving superior spatial resolution. In clinical images, preliminary comparisons within the high-dose group revealed that SR-DLR significantly outperformed C-DLR in spatial resolution metrics, specifically edge rise distance and edge rise slope (P<0.001). Furthermore, despite the lower radiation dose used for SR-DLR than for C-DLR (volume CT dose index, 4.25±0.54 mGy vs. 13.3±6.43 mGy), there were no significant differences in noise, contrast resolution, and spatial resolution between the groups.</p><p><strong>Conclusion: </strong>SR-DLR maintains non-inferior image quality compared with C-DLR even at reduced radiation doses. These findings substantiate the potential for further dose reduction in pediatric cardiac CT without compromising image quality.</p>","PeriodicalId":19755,"journal":{"name":"Pediatric Radiology","volume":" ","pages":"2238-2248"},"PeriodicalIF":2.5000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pediatric Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00247-026-06668-1","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/5/28 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"PEDIATRICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Cardiac computed tomography (CT) is widely utilized in pediatric cardiology, but minimizing radiation exposure is essential. Recently, super-resolution deep learning reconstruction (SR-DLR) has emerged as a potential advancement over conventional deep learning reconstruction (C-DLR).
Objective: To evaluate the potential of SR-DLR for further radiation dose reduction in pediatric cardiac CT through a comparison with C-DLR.
Materials and methods: The phantom-based study assessed noise power spectrum and spatial resolution. For the clinical study, we compared nine C-DLR images (conventional radiation dose) and 11 SR-DLR images (lower radiation dose). Evaluation metrics included noise levels, contrast-to-noise ratio, and edge rise distance and edge rise slope.
Results: In the phantom-based study, SR-DLR (250 mA) displayed equivalent noise characteristics as C-DLR (800 mA) while achieving superior spatial resolution. In clinical images, preliminary comparisons within the high-dose group revealed that SR-DLR significantly outperformed C-DLR in spatial resolution metrics, specifically edge rise distance and edge rise slope (P<0.001). Furthermore, despite the lower radiation dose used for SR-DLR than for C-DLR (volume CT dose index, 4.25±0.54 mGy vs. 13.3±6.43 mGy), there were no significant differences in noise, contrast resolution, and spatial resolution between the groups.
Conclusion: SR-DLR maintains non-inferior image quality compared with C-DLR even at reduced radiation doses. These findings substantiate the potential for further dose reduction in pediatric cardiac CT without compromising image quality.
期刊介绍:
Official Journal of the European Society of Pediatric Radiology, the Society for Pediatric Radiology and the Asian and Oceanic Society for Pediatric Radiology
Pediatric Radiology informs its readers of new findings and progress in all areas of pediatric imaging and in related fields. This is achieved by a blend of original papers, complemented by reviews that set out the present state of knowledge in a particular area of the specialty or summarize specific topics in which discussion has led to clear conclusions. Advances in technology, methodology, apparatus and auxiliary equipment are presented, and modifications of standard techniques are described.
Manuscripts submitted for publication must contain a statement to the effect that all human studies have been reviewed by the appropriate ethics committee and have therefore been performed in accordance with the ethical standards laid down in an appropriate version of the 1964 Declaration of Helsinki. It should also be stated clearly in the text that all persons gave their informed consent prior to their inclusion in the study. Details that might disclose the identity of the subjects under study should be omitted.