Rapid Fibrinogen Assessment With the qLabs FIB Point-of-Care System: Results From a Multicenter Clinical Validation Study.

International journal of laboratory hematology Pub Date : 2026-10-01 Epub Date: 2026-05-25 DOI:10.1111/ijlh.70133
Sandra Sanfilippo, Leslie Buisson, Blanche Delabie, Sophie Le Bellego, Marie-Eglantine Dujaric, Thibault Donnet
{"title":"Rapid Fibrinogen Assessment With the qLabs FIB Point-of-Care System: Results From a Multicenter Clinical Validation Study.","authors":"Sandra Sanfilippo, Leslie Buisson, Blanche Delabie, Sophie Le Bellego, Marie-Eglantine Dujaric, Thibault Donnet","doi":"10.1111/ijlh.70133","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Acquired hypofibrinogenemia is frequently associated with major bleeding and requires prompt correction. Rapid and accurate quantification of fibrinogen is essential to guide hemostatic therapy and improve patient outcomes. The qLabs FIB system provides fibrinogen results within minutes at the bedside, supporting timely diagnostic assessment and therapeutic decision-making. This study aimed to validate the device analytical performance in a prospective, observational, multicenter setting by comparing whole-blood qLabs FIB results with lab-based plasma fibrinogen.</p><p><strong>Methods: </strong>Fibrinogen concentrations from 463 citrated whole-blood samples were measured in parallel using the qLabs FIB and the Clauss method across 8 clinical sites.</p><p><strong>Results: </strong>Overall, a strong correlation was observed between the qLabs FIB and the Clauss laboratory method (r = 0.94). The median of relative differences did not differ significantly between clinical sites. Receiver operating characteristic analysis showed sensitivities and specificities in the 90% range for identifying fibrinogen concentrations below the clinical decision thresholds of 1.5 and 2.0 g/L.</p><p><strong>Conclusion: </strong>This study demonstrated a strong agreement of the qLabs FIB to the standard Clauss assay while providing rapid, near-patient fibrinogen measurements. This multicenter study confirms the initial performance of the system from a larger number of samples. The integration of qLabs FIB testing into routine clinical practice could support earlier, more targeted hemostatic interventions, thereby improving the safety and effectiveness of bleeding management.</p>","PeriodicalId":94050,"journal":{"name":"International journal of laboratory hematology","volume":" ","pages":"1173-1180"},"PeriodicalIF":0.0000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13555168/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of laboratory hematology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1111/ijlh.70133","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/5/25 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Acquired hypofibrinogenemia is frequently associated with major bleeding and requires prompt correction. Rapid and accurate quantification of fibrinogen is essential to guide hemostatic therapy and improve patient outcomes. The qLabs FIB system provides fibrinogen results within minutes at the bedside, supporting timely diagnostic assessment and therapeutic decision-making. This study aimed to validate the device analytical performance in a prospective, observational, multicenter setting by comparing whole-blood qLabs FIB results with lab-based plasma fibrinogen.

Methods: Fibrinogen concentrations from 463 citrated whole-blood samples were measured in parallel using the qLabs FIB and the Clauss method across 8 clinical sites.

Results: Overall, a strong correlation was observed between the qLabs FIB and the Clauss laboratory method (r = 0.94). The median of relative differences did not differ significantly between clinical sites. Receiver operating characteristic analysis showed sensitivities and specificities in the 90% range for identifying fibrinogen concentrations below the clinical decision thresholds of 1.5 and 2.0 g/L.

Conclusion: This study demonstrated a strong agreement of the qLabs FIB to the standard Clauss assay while providing rapid, near-patient fibrinogen measurements. This multicenter study confirms the initial performance of the system from a larger number of samples. The integration of qLabs FIB testing into routine clinical practice could support earlier, more targeted hemostatic interventions, thereby improving the safety and effectiveness of bleeding management.

使用qLabs FIB即时护理系统快速评估纤维蛋白原:来自多中心临床验证研究的结果。
背景:获得性低纤维蛋白原血症通常与大出血相关,需要及时纠正。快速准确地定量纤维蛋白原对指导止血治疗和改善患者预后至关重要。qLabs FIB系统在床边几分钟内提供纤维蛋白原结果,支持及时的诊断评估和治疗决策。本研究旨在通过比较全血qLabs FIB结果和实验室血浆纤维蛋白原,验证该设备在前瞻性、观察性、多中心环境下的分析性能。方法:采用qLabs FIB和Clauss方法在8个临床地点平行测量463份柠檬酸全血样本的纤维蛋白原浓度。结果:总体而言,qLabs FIB与Clauss实验室法之间存在很强的相关性(r = 0.94)。临床部位间相对差异的中位数无显著差异。受试者工作特征分析显示,在90%范围内,识别纤维蛋白原浓度低于临床判定阈值1.5和2.0 g/L的敏感性和特异性。结论:本研究证明了qLabs FIB在提供快速、近患者纤维蛋白原测量的同时,与标准的Clauss分析具有很强的一致性。这个多中心的研究从大量的样本中证实了系统的初始性能。将qLabs FIB检测整合到常规临床实践中可以支持更早、更有针对性的止血干预,从而提高出血管理的安全性和有效性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书