[Impact of Closed System Drug Transfer Device Architecture on Medical Waste Reduction - A Comparative Analysis of Filter-versus Barrier-Based Designs].
{"title":"[Impact of Closed System Drug Transfer Device Architecture on Medical Waste Reduction - A Comparative Analysis of Filter-versus Barrier-Based Designs].","authors":"Yasunari Okuda, Takahiro Saito, Yuki Mikame, Makoto Shinada, Daichi Arai, Erika Kanai, Izumi Tamura, Maaya Hasegawa, Shion Tanio, Saeko Usui, Ayano Katsura, Haruka Niimura, Tomohito Nakazawa, Yasushi Imai","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Closed system drug transfer devices (CSTDs) are recommended to minimize occupational exposure to antineoplastic drugs, yet their bulky components may increase medical waste. To compare waste volume/weight, greenhouse-gas (GHG) emissions, and disposal cost between filter-based and barrier-based CSTDs under real-world workload. A simulated preparation of 93 inpatient/outpatient chemotherapy cases (322 vials, 342 syringes) was performed using 2 filter CSTDs (Chemo-SafeLockTM and Chemfort®) and 1 barrier CSTD (PhaSealTM). Used components were shaken vertically in a 40-L container (2.5 Hz, 60 s) to determine packed volume and weight. Daily differences were extrapolated to the annual hospital activity (93 preparations × 246 days). Costs and GHG reductions were calculated with national disposal parameters. Filter CSTDs reduced waste volume by 27.9% (84.0 ± 2.3 L vs 114.4 L, p<0.01) without significant weight change. Full conversion would save 187 pails (7.5 kL), 1.06 million JPY, and 0.60 t-CO2e annually. Filter CSTDs provides equivalent containment while markedly decreasing waste and environmental impact, supporting their preferential adoption in sustainable oncology practice.</p>","PeriodicalId":35588,"journal":{"name":"Japanese Journal of Cancer and Chemotherapy","volume":"53 4","pages":"265-268"},"PeriodicalIF":0.0000,"publicationDate":"2026-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Japanese Journal of Cancer and Chemotherapy","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Closed system drug transfer devices (CSTDs) are recommended to minimize occupational exposure to antineoplastic drugs, yet their bulky components may increase medical waste. To compare waste volume/weight, greenhouse-gas (GHG) emissions, and disposal cost between filter-based and barrier-based CSTDs under real-world workload. A simulated preparation of 93 inpatient/outpatient chemotherapy cases (322 vials, 342 syringes) was performed using 2 filter CSTDs (Chemo-SafeLockTM and Chemfort®) and 1 barrier CSTD (PhaSealTM). Used components were shaken vertically in a 40-L container (2.5 Hz, 60 s) to determine packed volume and weight. Daily differences were extrapolated to the annual hospital activity (93 preparations × 246 days). Costs and GHG reductions were calculated with national disposal parameters. Filter CSTDs reduced waste volume by 27.9% (84.0 ± 2.3 L vs 114.4 L, p<0.01) without significant weight change. Full conversion would save 187 pails (7.5 kL), 1.06 million JPY, and 0.60 t-CO2e annually. Filter CSTDs provides equivalent containment while markedly decreasing waste and environmental impact, supporting their preferential adoption in sustainable oncology practice.
封闭系统药物转移装置(CSTDs)被推荐用于最大限度地减少抗肿瘤药物的职业暴露,但其笨重的部件可能会增加医疗浪费。在实际工作负荷下,比较基于过滤器和基于屏障的cstd的废物体积/重量、温室气体(GHG)排放量和处置成本。采用2个过滤式CSTD (Chemo-SafeLockTM和Chemfort®)和1个屏障式CSTD (PhaSealTM)对93例住院/门诊化疗患者(322瓶,342支注射器)进行模拟制备。用过的组件在40-L容器中(2.5 Hz, 60 s)垂直摇动,以确定包装体积和重量。每日差异外推到年度医院活动(93制剂× 246天)。成本和温室气体减量是根据国家处置参数计算的。过滤器CSTDs减少了27.9%的废物量(84.0±2.3 L vs 114.4 L, p