Radiometric and spectral characterization of multipanel planar light-emitting devices: a generalizable irradiance mapping approach for dose estimation.
Emmanuel Gerelli, Georges Wagnières, Jaroslava Joniová
{"title":"Radiometric and spectral characterization of multipanel planar light-emitting devices: a generalizable irradiance mapping approach for dose estimation.","authors":"Emmanuel Gerelli, Georges Wagnières, Jaroslava Joniová","doi":"10.1117/1.JBO.31.5.055001","DOIUrl":null,"url":null,"abstract":"<p><strong>Significance: </strong>Accurate spectral and radiometric characterization of photobiomodulation (PBM) devices is essential for reliable dosimetry and reproducible therapeutic outcomes. Complex, multilight-emitting diode (LED) systems produce heterogeneous irradiance patterns that complicate dose estimation, underscoring the need for standardized quantitative methods.</p><p><strong>Aim: </strong>To develop and validate a generalizable methodology for the spectral and radiometric characterization of multipanel and multi-LED PBM devices, demonstrated using the ATP38® system.</p><p><strong>Approach: </strong>Spectral properties were measured using an optical fiber-based spectrometer. The radiometric properties were determined by measuring 2D irradiance maps acquired at a 4-cm working distance from a single panel at six wavelengths. A critical element of the method was correcting raw irradiance data for the detector's nonideal angular response, modeled by a <math> <mrow><msup><mi>cos</mi> <mi>m</mi></msup> <mo>(</mo> <mi>θ</mi> <mo>)</mo></mrow> </math> function, with <math><mrow><mi>m</mi></mrow> </math> determined for each wavelength. The corrected maps were fitted using double 2D Gaussian surfaces, enabling the generation of continuous irradiance distributions and their extrapolation to other working distances through homothety. This framework also allowed simulation of irradiance maps for multipanel configurations. To validate the physical characterization, <i>in vitro</i> experiments on normal human epidermal keratinocytes compared the biological effects of three illumination protocols applied at two panel locations. PBM effects were assessed by measuring the endogenously produced protoporphyrin IX fluorescence intensity and by the scratch test assay.</p><p><strong>Results: </strong>Six emission peaks were identified at 454, 525, 594, 627, 729, and 842 nm. The detector angular correction improved the accuracy of the radiometric characterization, and Gaussian modeling of the irradiance enabled its prediction at different working distances. Biological responses correlated with spatial irradiance variations.</p><p><strong>Conclusions: </strong>This framework enables quantitative characterization of the ATP38® device spectral and radiometric properties, supporting accurate light propagation modeling and dosimetry optimization.</p>","PeriodicalId":15264,"journal":{"name":"Journal of Biomedical Optics","volume":"31 5","pages":"055001"},"PeriodicalIF":3.3000,"publicationDate":"2026-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13186584/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Biomedical Optics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1117/1.JBO.31.5.055001","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/5/19 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
Significance: Accurate spectral and radiometric characterization of photobiomodulation (PBM) devices is essential for reliable dosimetry and reproducible therapeutic outcomes. Complex, multilight-emitting diode (LED) systems produce heterogeneous irradiance patterns that complicate dose estimation, underscoring the need for standardized quantitative methods.
Aim: To develop and validate a generalizable methodology for the spectral and radiometric characterization of multipanel and multi-LED PBM devices, demonstrated using the ATP38® system.
Approach: Spectral properties were measured using an optical fiber-based spectrometer. The radiometric properties were determined by measuring 2D irradiance maps acquired at a 4-cm working distance from a single panel at six wavelengths. A critical element of the method was correcting raw irradiance data for the detector's nonideal angular response, modeled by a function, with determined for each wavelength. The corrected maps were fitted using double 2D Gaussian surfaces, enabling the generation of continuous irradiance distributions and their extrapolation to other working distances through homothety. This framework also allowed simulation of irradiance maps for multipanel configurations. To validate the physical characterization, in vitro experiments on normal human epidermal keratinocytes compared the biological effects of three illumination protocols applied at two panel locations. PBM effects were assessed by measuring the endogenously produced protoporphyrin IX fluorescence intensity and by the scratch test assay.
Results: Six emission peaks were identified at 454, 525, 594, 627, 729, and 842 nm. The detector angular correction improved the accuracy of the radiometric characterization, and Gaussian modeling of the irradiance enabled its prediction at different working distances. Biological responses correlated with spatial irradiance variations.
Conclusions: This framework enables quantitative characterization of the ATP38® device spectral and radiometric properties, supporting accurate light propagation modeling and dosimetry optimization.
意义:光生物调节(PBM)装置准确的光谱和放射学表征对于可靠的剂量学和可重复的治疗结果至关重要。复杂的多发光二极管(LED)系统产生不均匀的辐照模式,使剂量估计复杂化,强调需要标准化的定量方法。目的:开发和验证一种通用的方法,用于多面板和多led PBM器件的光谱和辐射特性,使用ATP38®系统进行演示。方法:使用光纤光谱仪测量光谱特性。通过测量从单个面板在6个波长下在4厘米工作距离处获得的二维辐照度图来确定辐射特性。该方法的一个关键要素是校正探测器非理想角响应的原始辐照度数据,该数据由cos m (θ)函数建模,每个波长确定m。校正后的地图使用双二维高斯曲面进行拟合,从而能够生成连续的辐照度分布,并通过同质性将其外推到其他工作距离。该框架还允许模拟多面板配置的辐照度图。为了验证物理特性,在正常人表皮角质形成细胞的体外实验中,比较了在两个面板位置应用三种照明方案的生物学效应。通过测量内源性原卟啉IX的荧光强度和划痕试验来评估PBM效应。结果:在454、525、594、627、729、842 nm处鉴定出6个发射峰。探测器的角度校正提高了辐射表征的精度,高斯辐射度建模使其在不同工作距离下的预测成为可能。生物反应与空间辐照度变化相关。结论:该框架能够定量表征ATP38®器件的光谱和辐射特性,支持精确的光传播建模和剂量学优化。