Total Serum Calcium Spike-Recovery Experiments to Investigate Bones of Contention in Clinical Laboratories.

IF 2.3 Q3 MEDICAL LABORATORY TECHNOLOGY
Sarah L Davies, Rachel Marrington, Iryna Mikheenko, Khadan Mohamed, Finlay MacKenzie
{"title":"Total Serum Calcium Spike-Recovery Experiments to Investigate Bones of Contention in Clinical Laboratories.","authors":"Sarah L Davies, Rachel Marrington, Iryna Mikheenko, Khadan Mohamed, Finlay MacKenzie","doi":"10.1093/jalm/jfag059","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To help assess patient safety risks associated with ongoing differences in total calcium assays following bias shifts, we performed spike-recovery experiments within our clinical chemistry external quality assessment (EQA) schemes to evaluate bias and recovery against a reference measurement procedure (RMP).</p><p><strong>Methods: </strong>Aqueous calcium chloride dihydrate was spiked (0.30/0.60 mmol/L; 1.20/2.40 mg/dL) into pooled human serum. Bias was assessed against RMP-assigned targets and recovery against added spikes. Minimum analytical performance specifications (APS) for standard measurement uncertainty (1.36%) defined acceptable bias/recovery.</p><p><strong>Results: </strong>Commercially available analytical principles across analytical platforms did not achieve adequate bias against RMP-assigned targets and recovery against added spikes for total calcium in 3 EQA specimens. Roche Cobas (5-nitro-5'-methyl-1,2-bis(o-aminophenoxy)ethane-N,N,N',N'-tetraacetic acid [NM-BAPTA]), Roche Cobas Pro (NM-BAPTA), and Beckman AU (Arsenazo III) methods exhibited better bias and long-term stability. The Abbott Alinity (Arsenazo III) method exhibited notable negative bias (-2.33% to -3.10%), apparently coinciding with the introduction of Abbott Calcium2. Negative bias was present to a lesser extent in the Abbott Architect (Arsenazo III). The Siemens Atellica (cresolphthalein complexone) assay exhibited notable proportional over recovery (110.2% to 112.0%). There was some positive bias in the Siemens Atellica (Arsenazo III) and QuidelOrtho (dry slide/sensor) methods.</p><p><strong>Conclusions: </strong>Differences across analytical principles/platforms and difficulties achieving minimum APS for total calcium were shown within our EQA schemes, notably negative bias for Abbott Alinity and Architect (Arsenazo III), over-recovery for Siemens Atellica (cresolphthalein complexone) and positive bias for Siemens Atellica (Arsenazo III). Understanding the impact on clinical decision-making and patient outcomes is imperative for managing any patient safety risks.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":"1093-1107"},"PeriodicalIF":2.3000,"publicationDate":"2026-09-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/jalm/jfag059","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: To help assess patient safety risks associated with ongoing differences in total calcium assays following bias shifts, we performed spike-recovery experiments within our clinical chemistry external quality assessment (EQA) schemes to evaluate bias and recovery against a reference measurement procedure (RMP).

Methods: Aqueous calcium chloride dihydrate was spiked (0.30/0.60 mmol/L; 1.20/2.40 mg/dL) into pooled human serum. Bias was assessed against RMP-assigned targets and recovery against added spikes. Minimum analytical performance specifications (APS) for standard measurement uncertainty (1.36%) defined acceptable bias/recovery.

Results: Commercially available analytical principles across analytical platforms did not achieve adequate bias against RMP-assigned targets and recovery against added spikes for total calcium in 3 EQA specimens. Roche Cobas (5-nitro-5'-methyl-1,2-bis(o-aminophenoxy)ethane-N,N,N',N'-tetraacetic acid [NM-BAPTA]), Roche Cobas Pro (NM-BAPTA), and Beckman AU (Arsenazo III) methods exhibited better bias and long-term stability. The Abbott Alinity (Arsenazo III) method exhibited notable negative bias (-2.33% to -3.10%), apparently coinciding with the introduction of Abbott Calcium2. Negative bias was present to a lesser extent in the Abbott Architect (Arsenazo III). The Siemens Atellica (cresolphthalein complexone) assay exhibited notable proportional over recovery (110.2% to 112.0%). There was some positive bias in the Siemens Atellica (Arsenazo III) and QuidelOrtho (dry slide/sensor) methods.

Conclusions: Differences across analytical principles/platforms and difficulties achieving minimum APS for total calcium were shown within our EQA schemes, notably negative bias for Abbott Alinity and Architect (Arsenazo III), over-recovery for Siemens Atellica (cresolphthalein complexone) and positive bias for Siemens Atellica (Arsenazo III). Understanding the impact on clinical decision-making and patient outcomes is imperative for managing any patient safety risks.

临床实验室研究争用骨的血清总钙峰值恢复实验。
背景:为了帮助评估偏倚转移后总钙测定持续差异相关的患者安全风险,我们在临床化学外部质量评估(EQA)方案中进行了尖峰恢复实验,以评估参考测量程序(RMP)的偏倚和恢复。方法:将氯化钙水溶液(0.30/0.60 mmol/L; 1.20/2.40 mg/dL)加入人血清中。根据rmp指定的目标评估偏差,根据增加的峰值评估恢复。标准测量不确定度(1.36%)定义可接受偏差/回收率的最小分析性能规范(APS)。结果:商业上可用的分析原理在分析平台上对rmp分配的目标和对3个EQA标本中总钙添加峰的回收率没有达到足够的偏差。Roche Cobas(5-硝基-5′-甲基-1,2-双(邻氨基苯氧基)乙烷-N,N,N′,N′-四乙酸[NM-BAPTA])、Roche Cobas Pro (NM-BAPTA)和Beckman AU(偶氮胂III)方法具有较好的偏倚性和长期稳定性。Abbott Alinity (Arsenazo III)方法表现出显著的负偏倚(-2.33% ~ -3.10%),这与Abbott Calcium2的引入一致。在雅培建筑师(Arsenazo III)中,负面偏见的存在程度较小。西门子Atellica(甲酚酞络合酮)法的回收率为110.2% ~ 112.0%。在Siemens Atellica (Arsenazo III)和QuidelOrtho(干载片/传感器)方法中存在一些积极的偏差。结论:我们的EQA方案显示了分析原理/平台之间的差异以及实现总钙最小APS的困难,特别是Abbott Alinity和Architect(偶氮胂III)的负偏倚,Siemens Atellica(甲酚酞络合酮)的过度回收和Siemens Atellica(偶氮胂III)的正偏倚。了解对临床决策和患者结果的影响是管理任何患者安全风险的必要条件。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Applied Laboratory Medicine
Journal of Applied Laboratory Medicine MEDICAL LABORATORY TECHNOLOGY-
CiteScore
3.70
自引率
5.00%
发文量
137
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书