Flow Control When Testing Dry-Powder Inhalers.

IF 2.2 4区 医学 Q3 RESPIRATORY SYSTEM
Daryl L Roberts
{"title":"Flow Control When Testing Dry-Powder Inhalers.","authors":"Daryl L Roberts","doi":"10.1177/19412711261448446","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>International compendia specify setting the flow for testing dry powder inhalers (DPIs) with the pressure ratio across the flow control valve ≤0.5. This guidance derives from thermodynamic theory, but there is no requirement that the flow control equipment demonstrates adherence to this theory. There is therefore a need to define and test for acceptable flow control valve performance, enabling DPI manufacturers to determine whether their test equipment is achieving the expected flow control. Herein, we present flow control data for two valves common in the inhaler testing community and describe the need for qualification specifications for flow control equipment with such valves.</p><p><strong>Methods: </strong>We devised a simple laboratory apparatus and method for determining the mass flow rate through two common flow control valves in their commercial configuration. In the test method, the pressure ratio across the valve is lowered from ∼0.7 to 0.3. This range of pressure ratio covers the standard test requirement that the pressure ratio must be 0.5 or less and maps the transition of compressible flow through the valve to the point where the air flow theoretically remains constant.</p><p><strong>Results: </strong>A common industrial valve, mentioned specifically in Pharm. Eur. 2.9.18 and in earlier editions of United States Pharmacopeia 601, exhibits nearly constant flow when the pressure ratio is <0.5. Another valve in widespread use exhibits a continued slight increase in mass flow rate even as the pressure ratio is decreased below 0.5.</p><p><strong>Conclusions: </strong>Inhaler testing laboratories should test whether their equipment exhibits the constant flow condition that is often assumed when the pressure ratio across the flow control valve is 0.5 or smaller. Quantitative specifications, yet to be developed, for qualifying commercial flow control equipment must account for both the transient start-up and the steady-state flow in DPI testing.</p>","PeriodicalId":14940,"journal":{"name":"Journal of Aerosol Medicine and Pulmonary Drug Delivery","volume":" ","pages":"228-235"},"PeriodicalIF":2.2000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Aerosol Medicine and Pulmonary Drug Delivery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/19412711261448446","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/5/5 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 0

Abstract

Background: International compendia specify setting the flow for testing dry powder inhalers (DPIs) with the pressure ratio across the flow control valve ≤0.5. This guidance derives from thermodynamic theory, but there is no requirement that the flow control equipment demonstrates adherence to this theory. There is therefore a need to define and test for acceptable flow control valve performance, enabling DPI manufacturers to determine whether their test equipment is achieving the expected flow control. Herein, we present flow control data for two valves common in the inhaler testing community and describe the need for qualification specifications for flow control equipment with such valves.

Methods: We devised a simple laboratory apparatus and method for determining the mass flow rate through two common flow control valves in their commercial configuration. In the test method, the pressure ratio across the valve is lowered from ∼0.7 to 0.3. This range of pressure ratio covers the standard test requirement that the pressure ratio must be 0.5 or less and maps the transition of compressible flow through the valve to the point where the air flow theoretically remains constant.

Results: A common industrial valve, mentioned specifically in Pharm. Eur. 2.9.18 and in earlier editions of United States Pharmacopeia 601, exhibits nearly constant flow when the pressure ratio is <0.5. Another valve in widespread use exhibits a continued slight increase in mass flow rate even as the pressure ratio is decreased below 0.5.

Conclusions: Inhaler testing laboratories should test whether their equipment exhibits the constant flow condition that is often assumed when the pressure ratio across the flow control valve is 0.5 or smaller. Quantitative specifications, yet to be developed, for qualifying commercial flow control equipment must account for both the transient start-up and the steady-state flow in DPI testing.

测试干粉吸入器时的流量控制。
背景:国际纲要规定了流量控制阀两端压力比≤0.5的干粉吸入器检测流量。这一指导来自热力学理论,但没有要求流量控制设备证明遵守这一理论。因此,需要定义和测试可接受的流量控制阀性能,使DPI制造商能够确定他们的测试设备是否达到预期的流量控制。在此,我们提供了吸入器测试界常见的两种阀门的流量控制数据,并描述了对带有此类阀门的流量控制设备的资格规范的需求。方法:我们设计了一种简单的实验装置和方法,通过两个常见的流量控制阀在其商业配置中测定质量流量。在这种测试方法中,阀门两端的压力比从0.7降低到0.3。该压力比范围涵盖了压力比必须小于等于0.5的标准测试要求,并映射了通过阀门的可压缩流量的过渡到理论上空气流量保持恒定的点。结果:一种常见的工业阀门,专门在药学中提到。结论:当流量控制阀上的压力比为0.5或更小时,吸入器测试实验室应该测试他们的设备是否显示恒定的流量条件。商业流量控制设备的定量规格尚未制定,必须考虑到DPI测试中的瞬态启动和稳态流量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.70
自引率
2.90%
发文量
34
审稿时长
>12 weeks
期刊介绍: Journal of Aerosol Medicine and Pulmonary Drug Delivery is the only peer-reviewed journal delivering innovative, authoritative coverage of the health effects of inhaled aerosols and delivery of drugs through the pulmonary system. The Journal is a forum for leading experts, addressing novel topics such as aerosolized chemotherapy, aerosolized vaccines, methods to determine toxicities, and delivery of aerosolized drugs in the intubated patient. Journal of Aerosol Medicine and Pulmonary Drug Delivery coverage includes: Pulmonary drug delivery Airway reactivity and asthma treatment Inhalation of particles and gases in the respiratory tract Toxic effects of inhaled agents Aerosols as tools for studying basic physiologic phenomena.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书