Comparison Between Respiratory-Triggered and Free-Breathing Signal Acquisition Methods for Single-Voxel Phosphorus Magnetic Resonance Spectroscopy of the Liver on a Clinical System.

IF 3.5 2区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Kumi Ozaki, Jihun Kwon, Marc Van Cauteren, Yasutomo Katsumata, Yukichi Tanahashi, Satoshi Goshima
{"title":"Comparison Between Respiratory-Triggered and Free-Breathing Signal Acquisition Methods for Single-Voxel Phosphorus Magnetic Resonance Spectroscopy of the Liver on a Clinical System.","authors":"Kumi Ozaki, Jihun Kwon, Marc Van Cauteren, Yasutomo Katsumata, Yukichi Tanahashi, Satoshi Goshima","doi":"10.1002/jmri.70352","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The optimal protocol for clinical liver <sup>31</sup>P-magnetic resonance spectroscopy (MRS) remains unclear. Single-voxel <sup>31</sup>P-MRS using image-selected in vivo spectroscopy (ISIS) employs respiratory-triggering (RT) or free-breathing (FB) acquisition. RT provides robust data but prolongs scan duration; FB allows faster acquisition but may suffer from low signal-to-noise ratio (SNR). Yet, direct comparison using a regulatory-approved coil in a clinical setting has not been reported.</p><p><strong>Purpose: </strong>To compare <sup>31</sup>P-MRS data stability and robustness between RT and FB.</p><p><strong>Study type: </strong>Prospective.</p><p><strong>Population: </strong>24 volunteers (19 male/5 female).</p><p><strong>Field strength/sequence: </strong>3 T MRI; single-voxel <sup>31</sup>P-MRS using ISIS (free induction decay-based) with approved <sup>31</sup>P coil.</p><p><strong>Methods: </strong><sup>31</sup>P-MRS was performed using RT and FB techniques (128 and 192 signal averages, respectively; expected scan duration ~13 min each). Spectra were analyzed using jMRUI. SNR and peak areas for PME, Pi, PDE, α-ATP, and NADPH, normalized using γ-ATP, were compared. PME/PDE and NADPH/(PME + PDE) ratios were also compared.</p><p><strong>Statistical test: </strong>Paired t-tests and Bland-Altman analysis were used. A p value < 0.05 was considered significant.</p><p><strong>Results: </strong>Scan duration was significantly longer for RT (15 min 44 s) than FB (12 min 56 s). No significant differences were observed for SNR (p = 0.570), NADPH/(PME + PDE) (p = 0.931), PME/γ-ATP (p = 0.556), Pi/γ-ATP (p = 0.931), α-ATP/γ-ATP (p = 0.332), or NADPH/γ-ATP (p = 0.394). Significant differences were noted for PDE/γ-ATP (RT 1.68 vs. FB 1.35, p = 0.003) and PME/PDE (RT 0.434 vs. FB 0.489, p = 0.046). Bland-Altman analysis showed near-zero fixed biases and no proportional bias, with limits of agreement from -0.53 to 0.62 (PME), -0.30 to 0.30 (Pi), -0.41 to 1.07 (PDE), -0.45 to 0.43 (α-ATP), and -0.80 to 0.90 (NADPH).</p><p><strong>Data conclusion: </strong><sup>31</sup>P-MRS of the liver showed equivalent stability and robustness for RT and FB. FB yielded comparable data within a shorter, predictable scan duration.</p><p><strong>Evidence level: </strong>2.</p><p><strong>Technical efficacy: </strong>1.</p>","PeriodicalId":16140,"journal":{"name":"Journal of Magnetic Resonance Imaging","volume":" ","pages":""},"PeriodicalIF":3.5000,"publicationDate":"2026-05-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Magnetic Resonance Imaging","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/jmri.70352","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The optimal protocol for clinical liver 31P-magnetic resonance spectroscopy (MRS) remains unclear. Single-voxel 31P-MRS using image-selected in vivo spectroscopy (ISIS) employs respiratory-triggering (RT) or free-breathing (FB) acquisition. RT provides robust data but prolongs scan duration; FB allows faster acquisition but may suffer from low signal-to-noise ratio (SNR). Yet, direct comparison using a regulatory-approved coil in a clinical setting has not been reported.

Purpose: To compare 31P-MRS data stability and robustness between RT and FB.

Study type: Prospective.

Population: 24 volunteers (19 male/5 female).

Field strength/sequence: 3 T MRI; single-voxel 31P-MRS using ISIS (free induction decay-based) with approved 31P coil.

Methods: 31P-MRS was performed using RT and FB techniques (128 and 192 signal averages, respectively; expected scan duration ~13 min each). Spectra were analyzed using jMRUI. SNR and peak areas for PME, Pi, PDE, α-ATP, and NADPH, normalized using γ-ATP, were compared. PME/PDE and NADPH/(PME + PDE) ratios were also compared.

Statistical test: Paired t-tests and Bland-Altman analysis were used. A p value < 0.05 was considered significant.

Results: Scan duration was significantly longer for RT (15 min 44 s) than FB (12 min 56 s). No significant differences were observed for SNR (p = 0.570), NADPH/(PME + PDE) (p = 0.931), PME/γ-ATP (p = 0.556), Pi/γ-ATP (p = 0.931), α-ATP/γ-ATP (p = 0.332), or NADPH/γ-ATP (p = 0.394). Significant differences were noted for PDE/γ-ATP (RT 1.68 vs. FB 1.35, p = 0.003) and PME/PDE (RT 0.434 vs. FB 0.489, p = 0.046). Bland-Altman analysis showed near-zero fixed biases and no proportional bias, with limits of agreement from -0.53 to 0.62 (PME), -0.30 to 0.30 (Pi), -0.41 to 1.07 (PDE), -0.45 to 0.43 (α-ATP), and -0.80 to 0.90 (NADPH).

Data conclusion: 31P-MRS of the liver showed equivalent stability and robustness for RT and FB. FB yielded comparable data within a shorter, predictable scan duration.

Evidence level: 2.

Technical efficacy: 1.

肝脏单体素磷磁共振波谱检测呼吸触发与自由呼吸信号采集方法的比较
背景:临床肝脏31p磁共振波谱(MRS)的最佳方案尚不清楚。使用图像选择体内光谱(ISIS)的单体素31P-MRS采用呼吸触发(RT)或自由呼吸(FB)采集。RT提供了可靠的数据,但延长了扫描时间;FB允许更快的采集,但可能受到低信噪比(SNR)的影响。然而,在临床环境中使用监管批准的线圈进行直接比较尚未有报道。目的:比较RT和FB之间31P-MRS数据的稳定性和稳健性。研究类型:前瞻性。人口:24名志愿者(19名男性/5名女性)。场强/序列:3t MRI;单体素31P- mrs使用ISIS(基于自由感应衰减)和批准的31P线圈。方法:采用RT和FB技术进行31P-MRS(分别为128和192个信号平均值,预计扫描时间约为13 min)。用jMRUI对光谱进行分析。用γ-ATP归一化后,比较PME、Pi、PDE、α-ATP和NADPH的信噪比和峰面积。比较PME/PDE和NADPH/(PME + PDE)比值。统计检验:采用配对t检验和Bland-Altman分析。结果:RT组扫描时间(15 min 44 s)明显长于FB组(12 min 56 s)。信噪比差异无显著意义(p = 0.570), NADPH /(中外+ PDE) (p = 0.931),中外/γatp (p = 0.556),π/γatp (p = 0.931),αatp /γatp (p = 0.332),或NADPH /γatp (p = 0.394)。PDE/γ-ATP (RT = 1.68, FB = 1.35, p = 0.003)和PME/PDE (RT = 0.434, FB = 0.489, p = 0.046)差异有统计学意义。Bland-Altman分析显示固定偏倚接近零,无比例偏倚,一致性限为-0.53 ~ 0.62 (PME), -0.30 ~ 0.30 (Pi), -0.41 ~ 1.07 (PDE), -0.45 ~ 0.43 (α-ATP), -0.80 ~ 0.90 (NADPH)。数据结论:肝脏31P-MRS对RT和FB具有同等的稳定性和稳健性。FB在较短的、可预测的扫描时间内产生了可比的数据。证据等级:2。技术功效:
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
9.70
自引率
6.80%
发文量
494
审稿时长
2 months
期刊介绍: The Journal of Magnetic Resonance Imaging (JMRI) is an international journal devoted to the timely publication of basic and clinical research, educational and review articles, and other information related to the diagnostic applications of magnetic resonance.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书