Impact of DOI Capability on Detector Performance: A Comparative Study of DOI and Non-DOI Detectors for High-Resolution and Sensitivity Organ-Specific PET Inserts

IF 3.8 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Lukas U. Rauscher;Magdelena S. Allen;Martin S. Judenhofer;Larry Byars;Michele Scipioni;José M. Udias;Ciprian Catana;Bernd J. Pichler;Fabian P. Schmidt
{"title":"Impact of DOI Capability on Detector Performance: A Comparative Study of DOI and Non-DOI Detectors for High-Resolution and Sensitivity Organ-Specific PET Inserts","authors":"Lukas U. Rauscher;Magdelena S. Allen;Martin S. Judenhofer;Larry Byars;Michele Scipioni;José M. Udias;Ciprian Catana;Bernd J. Pichler;Fabian P. Schmidt","doi":"10.1109/TRPMS.2025.3624609","DOIUrl":null,"url":null,"abstract":"High-sensitivity organ-specific positron emission tomography (PET) inserts employing long scintillation crystals in combination with small ring diameters require depth of interaction (DOI) encoding to ensure uniform spatial resolution across the field of view. In this study, we have developed high-resolution, DOI-capable detectors based on the Hamamatsu C13500 TOF-PET modules, using <inline-formula> <tex-math>$1.6\\times 1.6$ </tex-math></inline-formula> mm2 LSO crystals coupled to <inline-formula> <tex-math>$4\\times 4$ </tex-math></inline-formula> mm2 silicon photomultiplier. Detector configurations were designed either for a brain PET/MR insert (26 mm long crystals with a top light guide (TLG) for DOI encoding via light spread) or for a breast PET/MR insert (15 mm long with a TLG; 15 mm long with a stacked dual-layer configuration). Corresponding reference configurations without DOI capability were included to assess the impact of DOI implementation on detector performance. The 26 mm long brain detector configuration with TLG achieved an energy resolution of 12.1% and a coincidence time resolution (CTR) of 684 ps, with a degradation of 148 ps compared to the non-DOI variant. A newly developed DOI parameter calculation method improved DOI resolution from 9.1 ± 1.7 to 7.1 ± 1.8 mm. The 15 mm long single layer breast detector configuration with TLG yielded an energy resolution of 10.8%, a CTR of 460 ps, and a DOI resolution of 6.3 ± 1.8 mm. The 15 mm dual layer detector configuration provided comparable performance (11.5% energy resolution, 605 ps CTR) while eliminating the need for DOI calibration. For both the 15 and 26 mm single layer crystal block configurations, introducing a TLG preserved crystal identification and energy resolution. Overall, the DOI-capable designs demonstrated only minor performance tradeoffs, supporting the feasibility of both approaches for integration into organ-specific PET/MR inserts.","PeriodicalId":46807,"journal":{"name":"IEEE Transactions on Radiation and Plasma Medical Sciences","volume":"10 5","pages":"753-763"},"PeriodicalIF":3.8000,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Radiation and Plasma Medical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/11218221/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/10/27 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

High-sensitivity organ-specific positron emission tomography (PET) inserts employing long scintillation crystals in combination with small ring diameters require depth of interaction (DOI) encoding to ensure uniform spatial resolution across the field of view. In this study, we have developed high-resolution, DOI-capable detectors based on the Hamamatsu C13500 TOF-PET modules, using $1.6\times 1.6$ mm2 LSO crystals coupled to $4\times 4$ mm2 silicon photomultiplier. Detector configurations were designed either for a brain PET/MR insert (26 mm long crystals with a top light guide (TLG) for DOI encoding via light spread) or for a breast PET/MR insert (15 mm long with a TLG; 15 mm long with a stacked dual-layer configuration). Corresponding reference configurations without DOI capability were included to assess the impact of DOI implementation on detector performance. The 26 mm long brain detector configuration with TLG achieved an energy resolution of 12.1% and a coincidence time resolution (CTR) of 684 ps, with a degradation of 148 ps compared to the non-DOI variant. A newly developed DOI parameter calculation method improved DOI resolution from 9.1 ± 1.7 to 7.1 ± 1.8 mm. The 15 mm long single layer breast detector configuration with TLG yielded an energy resolution of 10.8%, a CTR of 460 ps, and a DOI resolution of 6.3 ± 1.8 mm. The 15 mm dual layer detector configuration provided comparable performance (11.5% energy resolution, 605 ps CTR) while eliminating the need for DOI calibration. For both the 15 and 26 mm single layer crystal block configurations, introducing a TLG preserved crystal identification and energy resolution. Overall, the DOI-capable designs demonstrated only minor performance tradeoffs, supporting the feasibility of both approaches for integration into organ-specific PET/MR inserts.
DOI能力对探测器性能的影响:用于高分辨率和灵敏度器官特异性PET插入物的DOI与非DOI探测器的比较研究
高灵敏度器官特异性正电子发射断层扫描(PET)插入采用长闪烁晶体与小环直径相结合,需要相互作用深度(DOI)编码,以确保整个视场的均匀空间分辨率。在这项研究中,我们开发了基于Hamamatsu C13500 TOF-PET模块的高分辨率,具有doi功能的探测器,使用1.6\ × 1.6$ mm2的LSO晶体耦合到4\ × 4$ mm2的硅光电倍增管。检测器配置被设计用于脑PET/MR插入物(26毫米长晶体,顶部光导(TLG),通过光传播进行DOI编码)或乳房PET/MR插入物(15毫米长,顶部光导,15毫米长,堆叠双层配置)。纳入不具备DOI能力的相应参考配置,以评估DOI实施对检测器性能的影响。带有TLG的26毫米长脑检测器配置实现了12.1%的能量分辨率和684 ps的符合时间分辨率(CTR),与非doi变体相比降低了148 ps。新开发的DOI参数计算方法将DOI分辨率从9.1±1.7 mm提高到7.1±1.8 mm。采用TLG的15 mm长单层乳房探测器结构,能量分辨率为10.8%,CTR为460 ps, DOI分辨率为6.3±1.8 mm。15毫米双层探测器配置提供了相当的性能(11.5%的能量分辨率,605 ps CTR),同时消除了DOI校准的需要。对于15和26毫米单层晶体块配置,引入了TLG保存晶体识别和能量分辨率。总体而言,具有doi功能的设计仅显示出较小的性能权衡,支持两种方法集成到特定器官的PET/MR插入件的可行性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
IEEE Transactions on Radiation and Plasma Medical Sciences
IEEE Transactions on Radiation and Plasma Medical Sciences RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
8.00
自引率
18.20%
发文量
109
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书