Dosimetric analysis of VMAT-TBI: measurement of organ and junction dose using a whole body anthropomorphic phantom.

IF 2.4 4区 环境科学与生态学 Q3 BIOLOGY
Radiation and Environmental Biophysics Pub Date : 2026-05-01 Epub Date: 2026-04-15 DOI:10.1007/s00411-026-01213-y
Nitin R Kakade, Rajesh Kumar, Mahendra More, Shatabdi Chakrabarty, S D Sharma, B K Sapra
{"title":"Dosimetric analysis of VMAT-TBI: measurement of organ and junction dose using a whole body anthropomorphic phantom.","authors":"Nitin R Kakade, Rajesh Kumar, Mahendra More, Shatabdi Chakrabarty, S D Sharma, B K Sapra","doi":"10.1007/s00411-026-01213-y","DOIUrl":null,"url":null,"abstract":"<p><p>Total body irradiation (TBI) is an essential component of conditioning regimens for hematopoietic stem cell transplantation. Volumetric Modulated Arc Therapy-TBI (VMAT-TBI) has recently emerged as an innovative alternative to conventional TBI that offers improved dose conformity. However, VMAT-TBI introduces junction regions between adjacent fields, making dosimetric verification crucial for ensuring uniform dose delivery. In addition, measurement of organ dose in organs such as brain, eyes, heart, liver, lung, kidney, and spinal cord is important for minimizing acute and late toxicities. The present study aimed at measuring the organ and junction doses using a whole body anthropomorphic phantom (ATOM 701-D, CIRS, Norfolk, Virginia). This phantom includes capsule-containing thermoluminescence dosimeters (TLDs) at multiple positions within organ and field junction. Computed Tomography (CT) images of the phantom were acquired to delineate different organs and the TLD locations. A treatment plan comprising four sub-plans with three junction fields was generated using the Eclipse treatment planning system (Version 17.0, Varian Medical Systems, Palo Alto, CA) to deliver 95% of the prescription dose to 95% of the planning target volume (PTV). Conformal dose coverage of the PTV was achieved and organ doses were within acceptable tolerances. The percentage deviation between doses measured with the TLDs and those calculated with the treatment planning system (TPS) with respect to prescription dose was evaluated at 38 positions (26 within various organs and 12 within junction regions). Percentage deviations at the 12 positions in the junction regions and 21 positions within organs were within ± 5% with a slightly higher differences ranging from − 6.72% to 7.43% at remaining five positions. The results demonstrate the effectiveness of VMAT-TBI as a promising alternative to conventional TBI, for hematopoietic stem cell transplantation.</p>","PeriodicalId":21002,"journal":{"name":"Radiation and Environmental Biophysics","volume":" ","pages":"793-801"},"PeriodicalIF":2.4000,"publicationDate":"2026-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation and Environmental Biophysics","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1007/s00411-026-01213-y","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/4/15 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"BIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Total body irradiation (TBI) is an essential component of conditioning regimens for hematopoietic stem cell transplantation. Volumetric Modulated Arc Therapy-TBI (VMAT-TBI) has recently emerged as an innovative alternative to conventional TBI that offers improved dose conformity. However, VMAT-TBI introduces junction regions between adjacent fields, making dosimetric verification crucial for ensuring uniform dose delivery. In addition, measurement of organ dose in organs such as brain, eyes, heart, liver, lung, kidney, and spinal cord is important for minimizing acute and late toxicities. The present study aimed at measuring the organ and junction doses using a whole body anthropomorphic phantom (ATOM 701-D, CIRS, Norfolk, Virginia). This phantom includes capsule-containing thermoluminescence dosimeters (TLDs) at multiple positions within organ and field junction. Computed Tomography (CT) images of the phantom were acquired to delineate different organs and the TLD locations. A treatment plan comprising four sub-plans with three junction fields was generated using the Eclipse treatment planning system (Version 17.0, Varian Medical Systems, Palo Alto, CA) to deliver 95% of the prescription dose to 95% of the planning target volume (PTV). Conformal dose coverage of the PTV was achieved and organ doses were within acceptable tolerances. The percentage deviation between doses measured with the TLDs and those calculated with the treatment planning system (TPS) with respect to prescription dose was evaluated at 38 positions (26 within various organs and 12 within junction regions). Percentage deviations at the 12 positions in the junction regions and 21 positions within organs were within ± 5% with a slightly higher differences ranging from − 6.72% to 7.43% at remaining five positions. The results demonstrate the effectiveness of VMAT-TBI as a promising alternative to conventional TBI, for hematopoietic stem cell transplantation.

VMAT-TBI的剂量学分析:使用全身拟人模型测量器官和接点剂量。
全身照射(TBI)是造血干细胞移植调理方案的重要组成部分。体积调制电弧治疗-TBI (VMAT-TBI)最近成为传统TBI的一种创新替代方案,提供了更好的剂量一致性。然而,VMAT-TBI引入了相邻场之间的结区,使得剂量学验证对于确保均匀剂量递送至关重要。此外,测量脑、眼、心、肝、肺、肾和脊髓等器官的剂量对于尽量减少急性和晚期毒性很重要。本研究旨在使用全身拟人化假体(ATOM 701-D, CIRS, Norfolk, Virginia)测量器官和接点剂量。该幻影包括在器官和野交界处多个位置的含热释光剂量计(tld)。计算机断层扫描(CT)的幻影图像描绘不同的器官和TLD的位置。使用Eclipse治疗计划系统(Version 17.0, Varian Medical Systems, Palo Alto, CA)生成包含四个子计划和三个连接域的治疗计划,以将95%的处方剂量提供到95%的计划目标体积(PTV)。达到了PTV的适形剂量覆盖,器官剂量在可接受的耐受范围内。用tld测量的剂量与用治疗计划系统(TPS)计算的剂量相对于处方剂量在38个位置(26个在不同器官内,12个在连接区域内)进行了百分比偏差的评估。结区12个位置和器官内21个位置的百分比偏差在±5%以内,其余5个位置的差异略大,在−6.72% ~ 7.43%之间。结果表明,VMAT-TBI作为传统TBI的一种有希望的替代方案,在造血干细胞移植中是有效的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.00
自引率
5.90%
发文量
53
审稿时长
>36 weeks
期刊介绍: This journal is devoted to fundamental and applied issues in radiation research and biophysics. The topics may include: Biophysics of ionizing radiation: radiation physics and chemistry, radiation dosimetry, radiobiology, radioecology, biophysical foundations of medical applications of radiation, and radiation protection. Biological effects of radiation: experimental or theoretical work on molecular or cellular effects; relevance of biological effects for risk assessment; biological effects of medical applications of radiation; relevance of radiation for biosphere and in space; modelling of ecosystems; modelling of transport processes of substances in biotic systems. Risk assessment: epidemiological studies of cancer and non-cancer effects; quantification of risk including exposures to radiation and confounding factors Contributions to these topics may include theoretical-mathematical and experimental material, as well as description of new techniques relevant for the study of these issues. They can range from complex radiobiological phenomena to issues in health physics and environmental protection.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书