Evaluation of analytical performance and clinical applicability of the UriSed Mini, a semi-automated urine analyzer for cell counting in cerebrospinal fluid.

IF 1.8
Biochemia medica Pub Date : 2026-06-15 Epub Date: 2026-04-15 DOI:10.11613/BM.2026.020705
Oana Roxana Oprea, Ana-Maria Fotache, Ion Bogdan Mănescu, Minodora Dobreanu
{"title":"Evaluation of analytical performance and clinical applicability of the UriSed Mini, a semi-automated urine analyzer for cell counting in cerebrospinal fluid.","authors":"Oana Roxana Oprea, Ana-Maria Fotache, Ion Bogdan Mănescu, Minodora Dobreanu","doi":"10.11613/BM.2026.020705","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Cerebrospinal fluid (CSF) analysis is essential in diagnosing central nervous system pathologies. While manual cell counting by optical microscopy is the gold standard, it is labor-intensive and time-consuming, posing challenges particularly in urgent cases, automated hematology analyzers offer alternatives but come with limitations. The aim of the study was to evaluate the performance of the UriSed Mini instrument for CSF in accordance with ICSH guidelines.</p><p><strong>Materials and methods: </strong>A total of 77 CSF patient samples were analyzed by both methods from September 2023 to September 2024. Following the International Council for Standardization in Hematology guidelines, the UriSed Mini was evaluated for short- and long-term imprecision, limit of blank, limit of detection and analytical measurement range. Quantitative comparison between the two methods was conducted using Bland-Altman and Passing-Bablok analyses. Qualitative agreement was assessed through inter-rater agreement testing.</p><p><strong>Results: </strong>The UriSed Mini successfully met all performance criteria, both in comparison to the manufacturer's claims and in the context of practical clinical requirements. While statistically significant differences were observed in the quantitative comparison (0.6 cells/µL), these discrepancies are unlikely to have clinical significance at the medical decision-making level. The qualitative comparison demonstrated excellent agreement, with a weighted kappa coefficient of 0.96 (95% confidence interval: 0.89-1.00).</p><p><strong>Conclusions: </strong>This study demonstrates that the UriSed Mini meets analytical standards for CSF cell counting and aligns closely with optical microscopy at clinically relevant thresholds.</p>","PeriodicalId":94370,"journal":{"name":"Biochemia medica","volume":"36 2","pages":"020705"},"PeriodicalIF":1.8000,"publicationDate":"2026-06-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13067900/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biochemia medica","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11613/BM.2026.020705","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/4/15 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Cerebrospinal fluid (CSF) analysis is essential in diagnosing central nervous system pathologies. While manual cell counting by optical microscopy is the gold standard, it is labor-intensive and time-consuming, posing challenges particularly in urgent cases, automated hematology analyzers offer alternatives but come with limitations. The aim of the study was to evaluate the performance of the UriSed Mini instrument for CSF in accordance with ICSH guidelines.

Materials and methods: A total of 77 CSF patient samples were analyzed by both methods from September 2023 to September 2024. Following the International Council for Standardization in Hematology guidelines, the UriSed Mini was evaluated for short- and long-term imprecision, limit of blank, limit of detection and analytical measurement range. Quantitative comparison between the two methods was conducted using Bland-Altman and Passing-Bablok analyses. Qualitative agreement was assessed through inter-rater agreement testing.

Results: The UriSed Mini successfully met all performance criteria, both in comparison to the manufacturer's claims and in the context of practical clinical requirements. While statistically significant differences were observed in the quantitative comparison (0.6 cells/µL), these discrepancies are unlikely to have clinical significance at the medical decision-making level. The qualitative comparison demonstrated excellent agreement, with a weighted kappa coefficient of 0.96 (95% confidence interval: 0.89-1.00).

Conclusions: This study demonstrates that the UriSed Mini meets analytical standards for CSF cell counting and aligns closely with optical microscopy at clinically relevant thresholds.

用于脑脊液细胞计数的半自动尿液分析仪UriSed Mini的分析性能和临床适用性评估
脑脊液(CSF)分析是诊断中枢神经系统病变必不可少的。虽然通过光学显微镜进行手动细胞计数是金标准,但它是劳动密集型和耗时的,特别是在紧急情况下提出了挑战,自动化血液学分析仪提供了替代方案,但存在局限性。本研究的目的是根据ICSH指南评估UriSed迷你脑脊液仪器的性能。材料与方法:采用两种方法对2023年9月至2024年9月共77例脑脊液标本进行分析。根据国际血液学标准化委员会的指导方针,对UriSed Mini进行了短期和长期不精确性、空白限、检测限和分析测量范围的评估。采用Bland-Altman和Passing-Bablok分析对两种方法进行定量比较。通过分级间一致性检验评价定性一致性。结果:UriSed Mini成功地满足了所有性能标准,无论是与制造商的声明相比,还是在实际临床要求的背景下。虽然在定量比较中观察到统计学上的显著差异(0.6个细胞/µL),但这些差异在医疗决策层面不太可能具有临床意义。定性比较显示出极好的一致性,加权kappa系数为0.96(95%可信区间:0.89-1.00)。结论:本研究表明UriSed Mini符合脑脊液细胞计数的分析标准,并且在临床相关阈值上与光学显微镜密切一致。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书