Laptop-style point-of-care ultrasound machines in the multiplace hyperbaric oxygen treatment chamber.

IF 0.7 4区 医学 Q4 MARINE & FRESHWATER BIOLOGY
Undersea and Hyperbaric Medicine Pub Date : 2026-01-01
Jenny Thacker, Marc Pullis, Thomas Masters, Christopher Logue
{"title":"Laptop-style point-of-care ultrasound machines in the multiplace hyperbaric oxygen treatment chamber.","authors":"Jenny Thacker, Marc Pullis, Thomas Masters, Christopher Logue","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Point-of-care ultrasound (POCUS) has diverse uses in the critical care hyperbaric (HBO₂) environment. The use of electronic devices in a multiplace HBO₂ chamber has traditionally been limited secondary to fire risk.</p><p><strong>Methods: </strong>The primary outcome was to assess the feasibility of modern laptop-style POCUS machines in a multiplace HBO₂ chamber. A secondary aim was to assess the reliability of POCUS images obtained at different pressures in a human subject in a multiplace chamber. Testing was initially conducted with one older machine (Sonosite 180 and Sonosite Micromaxx). Subsequently, two newer machines were studied (Sonosite Edge II and Mindray M9). Machine and transducer reliability were tested by performing a battery of POCUS imaging sequences: cardiac, FAST, inferior vena cava, and thoracic. The POCUS cardiac sequences evaluated included 2D, color Doppler, and M-mode.</p><p><strong>Results: </strong>Regarding feasibility, all machines functioned well at all depths/pressures. No instances of fire or spark were observed. No surface temperatures above 1200F were recorded. Regarding reliability, all machines performed adequately and completed the entire imaging sequence.</p><p><strong>Discussion: </strong>This study demonstrates the HBO₂ chamber safety of four laptop-style POCUS machines: Sonosite 180, Sonosite Micromaxx, Sonosite Edge II, and the Mindray M9. Our results also demonstrate excellent functionality of the newer Sonosite Edge II and Mindray machines. Secondarily, the reliability of POCUS images obtained at different pressures in a human subject has been verified.</p>","PeriodicalId":49396,"journal":{"name":"Undersea and Hyperbaric Medicine","volume":"53 1","pages":"51-58"},"PeriodicalIF":0.7000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Undersea and Hyperbaric Medicine","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MARINE & FRESHWATER BIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Point-of-care ultrasound (POCUS) has diverse uses in the critical care hyperbaric (HBO₂) environment. The use of electronic devices in a multiplace HBO₂ chamber has traditionally been limited secondary to fire risk.

Methods: The primary outcome was to assess the feasibility of modern laptop-style POCUS machines in a multiplace HBO₂ chamber. A secondary aim was to assess the reliability of POCUS images obtained at different pressures in a human subject in a multiplace chamber. Testing was initially conducted with one older machine (Sonosite 180 and Sonosite Micromaxx). Subsequently, two newer machines were studied (Sonosite Edge II and Mindray M9). Machine and transducer reliability were tested by performing a battery of POCUS imaging sequences: cardiac, FAST, inferior vena cava, and thoracic. The POCUS cardiac sequences evaluated included 2D, color Doppler, and M-mode.

Results: Regarding feasibility, all machines functioned well at all depths/pressures. No instances of fire or spark were observed. No surface temperatures above 1200F were recorded. Regarding reliability, all machines performed adequately and completed the entire imaging sequence.

Discussion: This study demonstrates the HBO₂ chamber safety of four laptop-style POCUS machines: Sonosite 180, Sonosite Micromaxx, Sonosite Edge II, and the Mindray M9. Our results also demonstrate excellent functionality of the newer Sonosite Edge II and Mindray machines. Secondarily, the reliability of POCUS images obtained at different pressures in a human subject has been verified.

多地点高压氧治疗室的笔记本电脑式即时超声机。
导语:即时超声(POCUS)在重症监护高压(HBO₂)环境中有多种用途。在多场所HBO 2室中使用电子设备传统上受到限制,其次是火灾风险。方法:主要目的是评估现代笔记本电脑式POCUS机器在多地点HBO 2室中的可行性。第二个目的是评估在不同压力下获得的POCUS图像的可靠性。测试最初是用一台较旧的机器(Sonosite 180和Sonosite Micromaxx)进行的。随后,研究了两台较新的机器(Sonosite Edge II和迈瑞M9)。通过执行一系列POCUS成像序列:心脏、FAST、下腔静脉和胸部,测试了机器和传感器的可靠性。POCUS心脏序列评估包括二维、彩色多普勒和m型。结果:就可行性而言,所有机器在所有深度/压力下均运行良好。没有观察到火灾或火花。没有地表温度超过1200华氏度的记录。在可靠性方面,所有机器都表现良好,完成了整个成像序列。讨论:本研究展示了四种笔记本式POCUS机器的HBO₂室安全性:Sonosite 180, Sonosite Micromaxx, Sonosite Edge II和Mindray M9。我们的结果也证明了较新的Sonosite Edge II和迈瑞机器的出色功能。其次,验证了在人体受试者不同压力下获得的POCUS图像的可靠性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Undersea and Hyperbaric Medicine
Undersea and Hyperbaric Medicine 医学-海洋与淡水生物学
CiteScore
1.60
自引率
11.10%
发文量
37
审稿时长
>12 weeks
期刊介绍: Undersea and Hyperbaric Medicine Journal accepts manuscripts for publication that are related to the areas of diving research and physiology, hyperbaric medicine and oxygen therapy, submarine medicine, naval medicine and clinical research related to the above topics. To be considered for UHM scientific papers must deal with significant and new research in an area related to biological, physical and clinical phenomena related to the above environments.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书