Impact of Detector Parameters and Image Resolution Modeling on Dedicated Brain PET Imaging

IF 3.8 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Min Gao;Samuel Matej;Joel S. Karp;Suleman Surti
{"title":"Impact of Detector Parameters and Image Resolution Modeling on Dedicated Brain PET Imaging","authors":"Min Gao;Samuel Matej;Joel S. Karp;Suleman Surti","doi":"10.1109/TRPMS.2025.3649140","DOIUrl":null,"url":null,"abstract":"High performance brain PET scanners suitable for accurate and precise measurement of uptake in small regions of the brain require high spatial resolution (1.5–2 mm) together with high system sensitivity. While small cross section crystals are needed to achieve high spatial resolution with pixelated detectors, effective system sensitivity requires long crystals and good time-of-flight (TOF) resolution. In addition, depending on crystal thickness, some level of depth-of-interaction (DOI) capability may also be needed to reduce parallax error. While various methods for DOI measurement can be utilized within the detector, the impact on reconstructed spatial resolution as a function of crystal thickness needs to be carefully evaluated. The focus of this work is to use Monte Carlo simulations to evaluate the impact of crystal size, thickness, and DOI resolution on the reconstructed spatial resolution in a high-resolution brain scanner design. Two crystal cross sections were chosen, 1.6 mm <inline-formula> <tex-math>$\\times 1$ </tex-math></inline-formula>.6 mm and 2 mm <inline-formula> <tex-math>$\\times $ </tex-math></inline-formula> 2 mm, and three different LSO crystal thickness (10, 15, and 20 mm) were evaluated. We evaluated these system designs without DOI capability and with DOI resolution (4–12 mm) for varying crystal thickness. In addition, we evaluated the use of image-space resolution modeling (IRM) techniques, which can mitigate losses in spatial resolution and positioning accuracy due to parallax error and Compton scatter, and to complement the advantages of DOI capability. A brain size elliptical phantom with small spheres was simulated to study the overall benefit in uptake measurements for small objects using the contrast recovery coefficient (CRC) metric. Our results show that a system with 2 mm <inline-formula> <tex-math>$\\times $ </tex-math></inline-formula> 2 mm crystals can provide comparable image quality in terms of CRC measurement than one with 1.6 mm <inline-formula> <tex-math>$\\times 1$ </tex-math></inline-formula>.6 mm crystals. However, crystal thickness has a bigger impact on spatial resolution and CRC with performance degrading with best performance achieved with 10 mm thick crystals. DOI measurement is necessary with longer crystals in order to maintain good imaging performance with high sensitivity. For the 15-mm thick crystals 9 mm, DOI resolution is needed to compensate for the loss in spatial resolution compared to 10-mm thick crystals without DOI, but still with lower CRC. A similar conclusion is reached for 20-mm crystals with 4-mm DOI resolution. Alternatively, IRM methods can provide improved and similar performance for all crystal thicknesses studied here, indicating a pathway for achieving our brain scanner design goals without additional cost, complexity, and performance tradeoffs. This work provides a systematic investigation for designing future brain PET systems.","PeriodicalId":46807,"journal":{"name":"IEEE Transactions on Radiation and Plasma Medical Sciences","volume":"10 7","pages":"1124-1131"},"PeriodicalIF":3.8000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Radiation and Plasma Medical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/11316579/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/12/26 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

High performance brain PET scanners suitable for accurate and precise measurement of uptake in small regions of the brain require high spatial resolution (1.5–2 mm) together with high system sensitivity. While small cross section crystals are needed to achieve high spatial resolution with pixelated detectors, effective system sensitivity requires long crystals and good time-of-flight (TOF) resolution. In addition, depending on crystal thickness, some level of depth-of-interaction (DOI) capability may also be needed to reduce parallax error. While various methods for DOI measurement can be utilized within the detector, the impact on reconstructed spatial resolution as a function of crystal thickness needs to be carefully evaluated. The focus of this work is to use Monte Carlo simulations to evaluate the impact of crystal size, thickness, and DOI resolution on the reconstructed spatial resolution in a high-resolution brain scanner design. Two crystal cross sections were chosen, 1.6 mm $\times 1$ .6 mm and 2 mm $\times $ 2 mm, and three different LSO crystal thickness (10, 15, and 20 mm) were evaluated. We evaluated these system designs without DOI capability and with DOI resolution (4–12 mm) for varying crystal thickness. In addition, we evaluated the use of image-space resolution modeling (IRM) techniques, which can mitigate losses in spatial resolution and positioning accuracy due to parallax error and Compton scatter, and to complement the advantages of DOI capability. A brain size elliptical phantom with small spheres was simulated to study the overall benefit in uptake measurements for small objects using the contrast recovery coefficient (CRC) metric. Our results show that a system with 2 mm $\times $ 2 mm crystals can provide comparable image quality in terms of CRC measurement than one with 1.6 mm $\times 1$ .6 mm crystals. However, crystal thickness has a bigger impact on spatial resolution and CRC with performance degrading with best performance achieved with 10 mm thick crystals. DOI measurement is necessary with longer crystals in order to maintain good imaging performance with high sensitivity. For the 15-mm thick crystals 9 mm, DOI resolution is needed to compensate for the loss in spatial resolution compared to 10-mm thick crystals without DOI, but still with lower CRC. A similar conclusion is reached for 20-mm crystals with 4-mm DOI resolution. Alternatively, IRM methods can provide improved and similar performance for all crystal thicknesses studied here, indicating a pathway for achieving our brain scanner design goals without additional cost, complexity, and performance tradeoffs. This work provides a systematic investigation for designing future brain PET systems.
探测器参数和图像分辨率建模对专用脑PET成像的影响。
高性能脑PET扫描仪适用于精确和精确测量大脑小区域的摄取,需要高空间分辨率(1.5-2毫米)和高系统灵敏度。虽然像素化探测器需要小截面晶体来实现高空间分辨率,但有效的系统灵敏度需要长晶体和良好的飞行时间(TOF)分辨率。此外,根据晶体厚度的不同,可能还需要某种程度的相互作用深度(DOI)能力来减少视差误差。虽然在探测器内部可以使用各种DOI测量方法,但需要仔细评估晶体厚度对重建空间分辨率的影响。本工作的重点是利用蒙特卡罗模拟来评估高分辨率脑扫描仪设计中晶体尺寸、厚度和DOI分辨率对重建空间分辨率的影响。选取1.6 mm × 1.6 mm和2mm × 2mm两种晶体截面,分别对10 mm、15 mm和20 mm三种不同的晶体厚度进行评价。我们在没有DOI能力和DOI分辨率(4-12 mm)的情况下对不同晶体厚度的系统设计进行了评估。此外,我们评估了图像空间分辨率建模(IRM)技术的使用,该技术可以减轻视差误差和康普顿散射造成的空间分辨率和定位精度损失,并补充DOI能力的优势。为了研究使用对比恢复系数(CRC)度量对小物体摄取测量的总体效益,我们模拟了一个带有小球体的脑大小椭圆幻影。我们的研究结果表明,在CRC测量方面,具有2 mm × 2 mm晶体的系统可以提供与1.6 mm × 1.6 mm晶体相当的图像质量。然而,晶体厚度对空间分辨率和CRC的影响较大,性能下降,晶体厚度为10 mm时性能最佳。为了保持良好的成像性能和高灵敏度,对较长的晶体进行DOI测量是必要的。对于15mm厚的晶体,与没有DOI的10mm厚晶体相比,需要9mm DOI分辨率来补偿空间分辨率的损失,但CRC仍然较低。类似的结论达到了20毫米晶体与4毫米DOI分辨率。另外,IRM方法可以为这里研究的所有晶体厚度提供改进的和类似的性能,这表明了在没有额外成本、复杂性和性能权衡的情况下实现我们的大脑扫描仪设计目标的途径。本研究为未来脑PET系统的设计提供了系统的研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
IEEE Transactions on Radiation and Plasma Medical Sciences
IEEE Transactions on Radiation and Plasma Medical Sciences RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
8.00
自引率
18.20%
发文量
109
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书