The IDO-Isometer: a valuable alternative for measuring the strength of external rotation of the shoulder: a comparative study between two measurement devices.
J Dufour, D Vandenbosch, A Karelse, L DE Wilde, A VAN Tongel
{"title":"The IDO-Isometer: a valuable alternative for measuring the strength of external rotation of the shoulder: a comparative study between two measurement devices.","authors":"J Dufour, D Vandenbosch, A Karelse, L DE Wilde, A VAN Tongel","doi":"10.52628/91.4.14714","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The hand-held dynamometer (HHD) is widely used to assess isometric shoulder strength but may present some limitations. The present study aimed to evaluate the reliability of the IDO-Isometer as a potential alternative for measuring isometric external rotation strength of the shoulder at 90° abduction.</p><p><strong>Methods: </strong>Ninety asymptomatic adult volunteers (45 males, 45 females) underwent strength testing of both shoulders using an HHD and the IDO-Isometer. Measurements were performed in a standardized supine position with the shoulder abducted to 90° and the elbow flexed at 90°. Each device was used three times per shoulder with a 3-minute rest interval between devices. The sequence of testing was randomized. Intra- and inter-rater reliability of the IDO-Isometer were evaluated using Intraclass Correlation Coefficients (ICCs).</p><p><strong>Results: </strong>The IDO-Isometer demonstrated excellent agreement with the HHD, with ICC values of 0.90 (95% CI: 0.85-0.93) for both dominant and non-dominant shoulders. Intra-rater reliability yielded ICCs of 0.95 and 0.92, and inter-rater reliability showed ICCs of 0.97 and 0.96, respectively.</p><p><strong>Conclusion: </strong>The IDO-Isometer demonstrated excellent reliability for measuring isometric external rotation at 90° shoulder abduction. Given its lower cost and ease of use, it may represent a valuable alternative to the HHD in both clinical and research settings.</p>","PeriodicalId":7018,"journal":{"name":"Acta orthopaedica Belgica","volume":"91 4","pages":"547-552"},"PeriodicalIF":0.7000,"publicationDate":"2025-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta orthopaedica Belgica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.52628/91.4.14714","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The hand-held dynamometer (HHD) is widely used to assess isometric shoulder strength but may present some limitations. The present study aimed to evaluate the reliability of the IDO-Isometer as a potential alternative for measuring isometric external rotation strength of the shoulder at 90° abduction.
Methods: Ninety asymptomatic adult volunteers (45 males, 45 females) underwent strength testing of both shoulders using an HHD and the IDO-Isometer. Measurements were performed in a standardized supine position with the shoulder abducted to 90° and the elbow flexed at 90°. Each device was used three times per shoulder with a 3-minute rest interval between devices. The sequence of testing was randomized. Intra- and inter-rater reliability of the IDO-Isometer were evaluated using Intraclass Correlation Coefficients (ICCs).
Results: The IDO-Isometer demonstrated excellent agreement with the HHD, with ICC values of 0.90 (95% CI: 0.85-0.93) for both dominant and non-dominant shoulders. Intra-rater reliability yielded ICCs of 0.95 and 0.92, and inter-rater reliability showed ICCs of 0.97 and 0.96, respectively.
Conclusion: The IDO-Isometer demonstrated excellent reliability for measuring isometric external rotation at 90° shoulder abduction. Given its lower cost and ease of use, it may represent a valuable alternative to the HHD in both clinical and research settings.