Shaojie Chang, Emily K Koons, Hao Gong, Jamison E Thorne, Eric E Williamson, Thomas A Foley, Cynthia H McCollough, Shuai Leng
{"title":"Deep Learning-based Monoenergetic Imaging for Calcified Coronary Stenosis Assessment at Energy-integrating Detector CT.","authors":"Shaojie Chang, Emily K Koons, Hao Gong, Jamison E Thorne, Eric E Williamson, Thomas A Foley, Cynthia H McCollough, Shuai Leng","doi":"10.1148/ryct.250230","DOIUrl":null,"url":null,"abstract":"<p><p>Purpose To develop and evaluate a Deep learnIng-bAsed MONoenergetic imaging at Different energies (DIAMOND) framework for generating virtual monoenergetic images (VMIs) from conventional energy-integrating detector (EID) CT, aiming to reduce blooming artifacts and improve stenosis assessment in coronary CT angiography (CCTA) with heavily calcified plaques. Materials and Methods This study (August 2022-September 2023) used a combination of retrospective and prospective imaging data. DIAMOND was trained using a simplified U-Net architecture on a retrospective dataset of 10 CCTA examinations performed with ultrahigh-resolution (UHR) photon-counting detector (PCD) CT; 70-keV PCD VMIs (energy equivalent to 120-kV single-energy EID CT) served as inputs and 100-keV PCD VMIs as targets. The trained model was then applied to a prospective dataset of participants with heavily calcified plaques who underwent EID CT at 120 kV followed by same-day PCD CT. Percent diameter stenosis (PDS) was quantified for a phantom and participants by using commercial software and compared across EID CT, DIAMOND, and PCD CT using Bland-Altman analysis. Changes in stenosis severity categorization based on PDS were evaluated. Results DIAMOND reduced blooming artifacts and improved lumen visualization, with image quality resembling PCD CT. In 23 participants (mean age, 69 years ± 8 [SD]; 18 male), average PDS decreased from 35.65% (EID CT) to 25.19% (DIAMOND, <i>P</i> < .05), approaching 24.27% with UHR PCD CT (<i>P</i> < .05). Relative to EID CT, DIAMOND led to Coronary Artery Disease Reporting and Data System reclassification in 11 of 26 (42%) lesions, mainly in mild to moderate stenosis ranges. Processing time was approximately 0.21 second per axial section on a standard graphics processing unit. Conclusion This study demonstrated the feasibility of using DIAMOND to generate high-kiloelectron volt VMIs from single-energy EID CT, providing artifact-reduced coronary imaging and improved stenosis quantification for heavily calcified plaques comparable to PCD CT without hardware upgrades. <b>Keywords:</b> Coronary CT Angiography, Coronary Artery Stenosis, Energy-Integrating Detector CT, Photon-Counting Detector CT, Deep Learning, CT-Photon Counting, Angiography, Coronary Arteries <i>Supplemental material is available for this article.</i> © RSNA, 2026.</p>","PeriodicalId":21168,"journal":{"name":"Radiology. Cardiothoracic imaging","volume":"8 2","pages":"e250230"},"PeriodicalIF":4.5000,"publicationDate":"2026-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13004168/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiology. Cardiothoracic imaging","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1148/ryct.250230","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose To develop and evaluate a Deep learnIng-bAsed MONoenergetic imaging at Different energies (DIAMOND) framework for generating virtual monoenergetic images (VMIs) from conventional energy-integrating detector (EID) CT, aiming to reduce blooming artifacts and improve stenosis assessment in coronary CT angiography (CCTA) with heavily calcified plaques. Materials and Methods This study (August 2022-September 2023) used a combination of retrospective and prospective imaging data. DIAMOND was trained using a simplified U-Net architecture on a retrospective dataset of 10 CCTA examinations performed with ultrahigh-resolution (UHR) photon-counting detector (PCD) CT; 70-keV PCD VMIs (energy equivalent to 120-kV single-energy EID CT) served as inputs and 100-keV PCD VMIs as targets. The trained model was then applied to a prospective dataset of participants with heavily calcified plaques who underwent EID CT at 120 kV followed by same-day PCD CT. Percent diameter stenosis (PDS) was quantified for a phantom and participants by using commercial software and compared across EID CT, DIAMOND, and PCD CT using Bland-Altman analysis. Changes in stenosis severity categorization based on PDS were evaluated. Results DIAMOND reduced blooming artifacts and improved lumen visualization, with image quality resembling PCD CT. In 23 participants (mean age, 69 years ± 8 [SD]; 18 male), average PDS decreased from 35.65% (EID CT) to 25.19% (DIAMOND, P < .05), approaching 24.27% with UHR PCD CT (P < .05). Relative to EID CT, DIAMOND led to Coronary Artery Disease Reporting and Data System reclassification in 11 of 26 (42%) lesions, mainly in mild to moderate stenosis ranges. Processing time was approximately 0.21 second per axial section on a standard graphics processing unit. Conclusion This study demonstrated the feasibility of using DIAMOND to generate high-kiloelectron volt VMIs from single-energy EID CT, providing artifact-reduced coronary imaging and improved stenosis quantification for heavily calcified plaques comparable to PCD CT without hardware upgrades. Keywords: Coronary CT Angiography, Coronary Artery Stenosis, Energy-Integrating Detector CT, Photon-Counting Detector CT, Deep Learning, CT-Photon Counting, Angiography, Coronary Arteries Supplemental material is available for this article. © RSNA, 2026.
基于深度学习的单能成像在能量积分检测器CT上评估钙化冠状动脉狭窄。
目的开发和评估基于深度学习的不同能量单能成像(DIAMOND)框架,用于从传统能量积分检测器(EID) CT生成虚拟单能图像(VMIs),旨在减少盛开伪影,改善严重钙化斑块冠状动脉CT血管造影(CCTA)的狭窄评估。材料与方法本研究(2022年8月- 2023年9月)采用回顾性和前瞻性影像资料相结合的方法。DIAMOND使用简化的U-Net架构对10个使用超高分辨率(UHR)光子计数检测器(PCD) CT进行的CCTA检查的回顾性数据集进行训练;输入为70 kev PCD vmi(能量相当于120 kv单能量EID CT),目标为100 kev PCD vmi。然后将训练好的模型应用于重度钙化斑块参与者的前瞻性数据集,这些参与者接受了120千伏的EID CT,随后进行了当天的PCD CT。使用商业软件量化假体和参与者的直径狭窄百分比(PDS),并使用Bland-Altman分析比较EID CT、DIAMOND和PCD CT。评估基于PDS的狭窄严重程度分类的变化。结果DIAMOND减少了盛开伪影,改善了流明可视化,图像质量与PCD CT相似。23名参与者(平均年龄69岁±8岁[SD]; 18名男性),平均PDS从35.65% (EID CT)下降到25.19% (DIAMOND, P < 0.05), UHR PCD CT接近24.27% (P < 0.05)。相对于EID CT, DIAMOND导致26个病变中的11个(42%)的冠状动脉疾病报告和数据系统重新分类,主要是在轻度到中度狭窄范围内。在标准图形处理单元上,每个轴向截面的处理时间约为0.21秒。本研究证明了使用DIAMOND从单能EID CT产生高千电子伏VMIs的可行性,与无需硬件升级的PCD CT相比,可以减少伪影的冠状动脉成像和改善严重钙化斑块的狭窄量化。关键词:冠状动脉CT血管造影,冠状动脉狭窄,能量积分检测器CT,光子计数检测器CT,深度学习,CT光子计数,血管造影,冠状动脉©rsna, 2026。
本文章由计算机程序翻译,如有差异,请以英文原文为准。