Katrina L Falk, Michael A Speidel, Grace M Minesinger, Gretchen M Foltz, Fred T Lee, Timothy J Ziemlewicz, Martin G Wagner, Paul F Laeseke
{"title":"Injectable radiopaque targets for cone-beam CT guided histotripsy.","authors":"Katrina L Falk, Michael A Speidel, Grace M Minesinger, Gretchen M Foltz, Fred T Lee, Timothy J Ziemlewicz, Martin G Wagner, Paul F Laeseke","doi":"10.1080/02656736.2026.2642172","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Cone-beam CT (CBCT) guided histotripsy is being developed as a supplemental image guidance technique for histotripsy. The goal of this work is to present a percutaneously injectable, radiopaque target that can be used in preclinical testing of CBCT-guided histotripsy.</p><p><strong>Methods: </strong>Livers of five <i>in vivo</i> swine were percutaneously injected with a mixture of barium paste diluted with saline or autologous blood to a concentration of 4:1. Post-injection CBCT images were used to analyze radiographic differences in injected volume; temporal stability of the injection (intensity, volume, and sphericity); and presence of image artifacts. Four injections were targeted and treated <i>in vivo</i> using a clinical histotripsy system. Injectate volume and intensity, and treatment zone size and contrast-to-noise ratio (CNR) were compared to control treatments.</p><p><strong>Results: </strong>There was no significant difference in imaged volume or intensity between the 100 or 200 µL(<i>p</i> = 0.384) injection. Over the analyzed 5-206 min, the barium-saline and barium-blood injections had similar ranges of imaged volume and intensity, both changing during the first 30 min then were stable. The image artifact was reported as none to moderate in 5/6 cases for both injections. The CNR of the treatment zones and difference in diameters (planned versus actual) were 5.9 ± 1.8, 3.7 ± 1.3, and 5.9 ± 1.0 and 8.1 ± 4.0 mm, 15.9 ± 11.4 mm, and 7.6 ± 2.2 mm for the barium-saline, barium-blood and control groups respectively. The barium-saline injectate was not visible after treatment while the barium-blood was visible but dispersed.</p><p><strong>Conclusion: </strong>Percutaneously injectable, radiopaque barium-based targets were acutely safe, targetable and treatable on CBCT and did not alter histotripsy treatment zone size <i>in vivo</i>.</p>","PeriodicalId":520653,"journal":{"name":"International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group","volume":"43 1","pages":"2642172"},"PeriodicalIF":3.0000,"publicationDate":"2026-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13072709/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/02656736.2026.2642172","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/12 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Cone-beam CT (CBCT) guided histotripsy is being developed as a supplemental image guidance technique for histotripsy. The goal of this work is to present a percutaneously injectable, radiopaque target that can be used in preclinical testing of CBCT-guided histotripsy.
Methods: Livers of five in vivo swine were percutaneously injected with a mixture of barium paste diluted with saline or autologous blood to a concentration of 4:1. Post-injection CBCT images were used to analyze radiographic differences in injected volume; temporal stability of the injection (intensity, volume, and sphericity); and presence of image artifacts. Four injections were targeted and treated in vivo using a clinical histotripsy system. Injectate volume and intensity, and treatment zone size and contrast-to-noise ratio (CNR) were compared to control treatments.
Results: There was no significant difference in imaged volume or intensity between the 100 or 200 µL(p = 0.384) injection. Over the analyzed 5-206 min, the barium-saline and barium-blood injections had similar ranges of imaged volume and intensity, both changing during the first 30 min then were stable. The image artifact was reported as none to moderate in 5/6 cases for both injections. The CNR of the treatment zones and difference in diameters (planned versus actual) were 5.9 ± 1.8, 3.7 ± 1.3, and 5.9 ± 1.0 and 8.1 ± 4.0 mm, 15.9 ± 11.4 mm, and 7.6 ± 2.2 mm for the barium-saline, barium-blood and control groups respectively. The barium-saline injectate was not visible after treatment while the barium-blood was visible but dispersed.
Conclusion: Percutaneously injectable, radiopaque barium-based targets were acutely safe, targetable and treatable on CBCT and did not alter histotripsy treatment zone size in vivo.