{"title":"Safety and performance of the Hydra self-expanding THV: 6 months outcomes from the GENESIS-II study","authors":"Anmol Sonawane, Praveen Chandra, John Jose, Sandeep Bansal, Rahul Gupta, Krishnankutty Sudhir, Nagendra Boopathy Senguttuvan","doi":"10.1016/j.ihj.2026.03.004","DOIUrl":null,"url":null,"abstract":"<div><h3>Aim</h3><div>The Hydra THV incorporates an active release mechanism designed to facilitate the deployment of the supra-annular valve. The aim of the GENESIS-II study was to assess the safety and performance of Hydra THV at 6 months follow-up after treating severe aortic stenosis in patients at high surgical risk.</div></div><div><h3>Methods</h3><div>This prospective, multicenter, non-randomized, investigational study was conducted between November 2021 and November 2023. The study enrolled 40 patients with high surgical risk and symptomatic severe aortic stenosis from 19 sites across India. The primary safety endpoint of the study was cardiovascular mortality at 30 days, while the primary performance endpoint was device success, as defined by VARC-2.</div></div><div><h3>Results</h3><div>Among the 40 patients, the mean age was 74.4 ± 6.7 years, 65% were male, and the mean Society of Thoracic Surgeons (STS) risk score was 3.5 ± 1.6%. The device had a success rate of 95%. At 6 months, there was one cardiovascular death (2.5 %). All-cause death was observed in 5% patients. The incidence of new permanent pacemaker implantation was 7.9%. There was a progressive enhancement in effective orifice area, from 0.6 ± 0.2 cm<sup>2</sup> at baseline to 1.9 ± 0.4 cm<sup>2</sup> at 6 months (<em>p</em> < 0.001). Similarly, the mean aortic valve gradient significantly improved from 53.1 ± 12.5 mm Hg at baseline to 8.7 ± 3.3 mmHg at 6 months (<em>p</em> < 0.001) post-intervention.</div></div><div><h3>Conclusion</h3><div>The results of 6 months follow-up after implantation of the Hydra THV with an active release mechanism in the GENESIS-II study demonstrated a high device success rate with favorable safety and performance of the device.</div></div>","PeriodicalId":13384,"journal":{"name":"Indian heart journal","volume":"78 4","pages":"Pages 201-205"},"PeriodicalIF":2.0000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian heart journal","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0019483226000350","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/10 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Aim
The Hydra THV incorporates an active release mechanism designed to facilitate the deployment of the supra-annular valve. The aim of the GENESIS-II study was to assess the safety and performance of Hydra THV at 6 months follow-up after treating severe aortic stenosis in patients at high surgical risk.
Methods
This prospective, multicenter, non-randomized, investigational study was conducted between November 2021 and November 2023. The study enrolled 40 patients with high surgical risk and symptomatic severe aortic stenosis from 19 sites across India. The primary safety endpoint of the study was cardiovascular mortality at 30 days, while the primary performance endpoint was device success, as defined by VARC-2.
Results
Among the 40 patients, the mean age was 74.4 ± 6.7 years, 65% were male, and the mean Society of Thoracic Surgeons (STS) risk score was 3.5 ± 1.6%. The device had a success rate of 95%. At 6 months, there was one cardiovascular death (2.5 %). All-cause death was observed in 5% patients. The incidence of new permanent pacemaker implantation was 7.9%. There was a progressive enhancement in effective orifice area, from 0.6 ± 0.2 cm2 at baseline to 1.9 ± 0.4 cm2 at 6 months (p < 0.001). Similarly, the mean aortic valve gradient significantly improved from 53.1 ± 12.5 mm Hg at baseline to 8.7 ± 3.3 mmHg at 6 months (p < 0.001) post-intervention.
Conclusion
The results of 6 months follow-up after implantation of the Hydra THV with an active release mechanism in the GENESIS-II study demonstrated a high device success rate with favorable safety and performance of the device.
期刊介绍:
Indian Heart Journal (IHJ) is the official peer-reviewed open access journal of Cardiological Society of India and accepts articles for publication from across the globe. The journal aims to promote high quality research and serve as a platform for dissemination of scientific information in cardiology with particular focus on South Asia. The journal aims to publish cutting edge research in the field of clinical as well as non-clinical cardiology - including cardiovascular medicine and surgery. Some of the topics covered are Heart Failure, Coronary Artery Disease, Hypertension, Interventional Cardiology, Cardiac Surgery, Valvular Heart Disease, Pulmonary Hypertension and Infective Endocarditis. IHJ open access invites original research articles, research briefs, perspective, case reports, case vignette, cardiovascular images, cardiovascular graphics, research letters, correspondence, reader forum, and interesting photographs, for publication. IHJ open access also publishes theme-based special issues and abstracts of papers presented at the annual conference of the Cardiological Society of India.