Chuang-Wei Wei, Run-Zhi Zhang, Yan Xu, Jia-Yi Liu, Nan Zhang, Lei Xu, Zhao-Ying Wen
{"title":"Non-contrast 3.0T coronary magnetic resonance angiography: Image quality comparison of coronal balanced-turbo-field-echo, axial turbo-field-echo, and modified-Dixon sequences.","authors":"Chuang-Wei Wei, Run-Zhi Zhang, Yan Xu, Jia-Yi Liu, Nan Zhang, Lei Xu, Zhao-Ying Wen","doi":"10.4329/wjr.v18.i2.115741","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Non-contrast whole-heart coronary magnetic resonance angiography (CMRA) remains underutilized in clinical practice due to limited visualization of distal vessels and prolonged acquisition times.</p><p><strong>Aim: </strong>To evaluate the performance of a coronal balanced turbo field echo (BTFE) sequence for CMRA at 3.0T in comparison with conventional axial turbo field echo (TFE) and modified Dixon (mDixon) sequences.</p><p><strong>Methods: </strong>Healthy young volunteers were prospectively enrolled from January 2025 to April 2025. All participants underwent coronary artery imaging using BTFE, TFE, and mDixon sequences. Subjective image quality was assessed based on the society of cardiovascular computed tomography 18-segment model using a four-point scale (1 = non-assessable to 4 = excellent) across the three major coronary arteries and their side branches. The assessability rate was defined as the percentage of segments receiving a score ≥ 2. Objective evaluation of the main coronary arteries included measurements of signal-to-noise ratio, contrast-to-noise ratio, vessel edge sharpness, and visible vessel length. Acquisition time was recorded for each sequence. Statistical comparisons among the three sequences were performed using the Friedman test and one-way repeated measures analysis of variance.</p><p><strong>Results: </strong>A total of 22 participants (mean age ± SD, 23 ± 2 years; 12 men) were included in the study. BTFE significantly outperformed both TFE and mDixon in subjective image quality scores across all coronary segments (all <i>P</i> < 0.05), except for the left main coronary artery, the proximal and mid segments of the left anterior descending artery (LAD), the proximal and mid segments of the left circumflex artery (LCX), the proximal segment of the right coronary artery, and ramus intermedius, for which no significant differences were observed between BTFE and mDixon (<i>P</i> > 0.05). The BTFE sequence achieved the highest overall assessability (99.3%, 282/284), surpassing mDixon (95.4%, 271/284) and TFE (87.0%, 247/284). BTFE yielded significantly higher signal-to-noise ratio, contrast-to-noise ratio, and vessel edge sharpness compared with both TFE and mDixon (all <i>P</i> < 0.001). The measured lengths of the LAD and right coronary artery were also significantly greater with BTFE (<i>P</i> < 0.05 for both). The LCX length was significantly longer on BTFE than on TFE (77.5 ± 10.5 mm <i>vs</i> 75.1 ± 8.3 mm; <i>P</i> = 0.017), but not significantly different from that on mDixon (77.5 ± 10.5 mm <i>vs</i> 75.7 ± 8.8 mm; <i>P</i> = 0.074). Additionally, BTFE demonstrated a shorter and more consistent acquisition time compared with both TFE and mDixon (6.13 ± 1.00 minutes <i>vs</i> 7.15 ± 2.02 minutes <i>vs</i> 7.17 ± 1.72 minutes; <i>P</i> < 0.05).</p><p><strong>Conclusion: </strong>Coronal BTFE-based non-contrast whole-heart CMRA at 3.0T offers superior image quality and reduced acquisition time compared with conventional axial TFE and mDixon sequences.</p>","PeriodicalId":23819,"journal":{"name":"World journal of radiology","volume":"18 2","pages":"115741"},"PeriodicalIF":1.2000,"publicationDate":"2026-02-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12968789/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"World journal of radiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4329/wjr.v18.i2.115741","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Non-contrast whole-heart coronary magnetic resonance angiography (CMRA) remains underutilized in clinical practice due to limited visualization of distal vessels and prolonged acquisition times.
Aim: To evaluate the performance of a coronal balanced turbo field echo (BTFE) sequence for CMRA at 3.0T in comparison with conventional axial turbo field echo (TFE) and modified Dixon (mDixon) sequences.
Methods: Healthy young volunteers were prospectively enrolled from January 2025 to April 2025. All participants underwent coronary artery imaging using BTFE, TFE, and mDixon sequences. Subjective image quality was assessed based on the society of cardiovascular computed tomography 18-segment model using a four-point scale (1 = non-assessable to 4 = excellent) across the three major coronary arteries and their side branches. The assessability rate was defined as the percentage of segments receiving a score ≥ 2. Objective evaluation of the main coronary arteries included measurements of signal-to-noise ratio, contrast-to-noise ratio, vessel edge sharpness, and visible vessel length. Acquisition time was recorded for each sequence. Statistical comparisons among the three sequences were performed using the Friedman test and one-way repeated measures analysis of variance.
Results: A total of 22 participants (mean age ± SD, 23 ± 2 years; 12 men) were included in the study. BTFE significantly outperformed both TFE and mDixon in subjective image quality scores across all coronary segments (all P < 0.05), except for the left main coronary artery, the proximal and mid segments of the left anterior descending artery (LAD), the proximal and mid segments of the left circumflex artery (LCX), the proximal segment of the right coronary artery, and ramus intermedius, for which no significant differences were observed between BTFE and mDixon (P > 0.05). The BTFE sequence achieved the highest overall assessability (99.3%, 282/284), surpassing mDixon (95.4%, 271/284) and TFE (87.0%, 247/284). BTFE yielded significantly higher signal-to-noise ratio, contrast-to-noise ratio, and vessel edge sharpness compared with both TFE and mDixon (all P < 0.001). The measured lengths of the LAD and right coronary artery were also significantly greater with BTFE (P < 0.05 for both). The LCX length was significantly longer on BTFE than on TFE (77.5 ± 10.5 mm vs 75.1 ± 8.3 mm; P = 0.017), but not significantly different from that on mDixon (77.5 ± 10.5 mm vs 75.7 ± 8.8 mm; P = 0.074). Additionally, BTFE demonstrated a shorter and more consistent acquisition time compared with both TFE and mDixon (6.13 ± 1.00 minutes vs 7.15 ± 2.02 minutes vs 7.17 ± 1.72 minutes; P < 0.05).
Conclusion: Coronal BTFE-based non-contrast whole-heart CMRA at 3.0T offers superior image quality and reduced acquisition time compared with conventional axial TFE and mDixon sequences.
背景:非对比全心冠状动脉磁共振血管造影(CMRA)在临床实践中仍未得到充分利用,因为远端血管的可视化有限,采集时间长。目的:评价3.0T时冠状平衡涡轮场回波序列(BTFE)与传统轴向涡轮场回波序列(TFE)和改进的Dixon序列(mDixon)在CMRA中的性能。方法:于2025年1月至2025年4月前瞻性招募健康青年志愿者。所有参与者均采用BTFE、TFE和mDixon序列进行冠状动脉成像。主观图像质量根据心血管计算机断层扫描学会18段模型进行评估,采用4分制(1 =不可评估至4 =优秀),横跨三条主要冠状动脉及其侧支。可评估率定义为得分≥2的部分的百分比。主要冠状动脉的客观评价包括测量信噪比、对比噪声比、血管边缘清晰度和可见血管长度。记录每个序列的采集时间。采用Friedman检验和单向重复测量方差分析对三个序列进行统计学比较。结果:共纳入22例受试者(平均年龄±SD, 23±2岁,男性12例)。BTFE在所有冠状动脉段的主观图像质量评分均显著优于TFE和mDixon(均P < 0.05),除了左冠状动脉主动脉、左前降支(LAD)近段和中段、左旋支(LCX)近段和中段、右冠状动脉近段和中间支外,BTFE和mDixon在所有冠状动脉段的主观图像质量评分上无显著差异(P < 0.05)。BTFE序列总体可评价性最高(99.3%,282/284),超过mDixon序列(95.4%,271/284)和TFE序列(87.0%,247/284)。与TFE和mDixon相比,BTFE的信噪比、噪声对比比和血管边缘清晰度显著更高(均P < 0.001)。BTFE组左冠状动脉和右冠状动脉长度均显著增加(P < 0.05)。BTFE组LCX长度明显长于TFE组(77.5±10.5 mm vs 75.1±8.3 mm, P = 0.017),但与mDixon组LCX长度无显著差异(77.5±10.5 mm vs 75.7±8.8 mm, P = 0.074)。此外,与TFE和mDixon相比,BTFE显示出更短且更一致的采集时间(6.13±1.00分钟vs 7.15±2.02分钟vs 7.17±1.72分钟;P < 0.05)。结论:与传统的轴向TFE和mDixon序列相比,基于冠状面btfe的3.0T非对比全心CMRA具有更好的图像质量和更短的采集时间。