Advanced spectrofluorimetric determination of ascorbic acid via serotonin hydrochloride quenching using a twin solar cell device with eight-angled blue radiation sources
{"title":"Advanced spectrofluorimetric determination of ascorbic acid via serotonin hydrochloride quenching using a twin solar cell device with eight-angled blue radiation sources","authors":"Ghufran K. Allawi , N.S. Turkey","doi":"10.1016/j.rechem.2026.103207","DOIUrl":null,"url":null,"abstract":"<div><div>A novel spectrofluorimetric flow injection analysis (FIA) method was developed for the selective quantification of ascorbic acid via fluorescence quenching of serotonin hydrochloride. The system employs a custom-designed photometric array comprising 16 irradiation sources arranged in a dual-axis matrix—eight aligned horizontally and eight orthogonally, enabling multi-angle excitation and enhanced spectral resolution. Fluorescence signals were captured using a twin-pair solar cell detector, offering high sensitivity and minimal optical interference. The method exhibited a linear calibration range of 0.1–30 μg/mL with a correlation coefficient (r<sup>2</sup>) of 0.9966, a limit of detection (LOD) of 0.025 μmol /L, equivalent to 4.403 * 10<sup>−4</sup> μg/mL per sample (0.1 mL (and limit of quantitative (LOQ) of 99.381 * 10<sup>−3</sup> <span><math><mi>μg</mi><mo>/</mo><mi>ml</mi></math></span>. Precision and accuracy were confirmed through intra- and inter-day validation, yielding relative standard deviations (RSD) below 2% and recovery rates exceeding 99%. Statistical comparison with conventional fluorometric techniques using a paired <em>t</em>-test revealed no significant difference at the 95% confidence level (<em>p</em> > 0.05), affirming the method's reliability. The proposed system supports high-throughput analysis of up to 50 samples per hour, with reduced reagent consumption and minimal environmental impact. Its robustness, cost-efficiency, and analytical performance render it a promising tool for routine pharmaceutical quality control and environmental monitoring.</div></div>","PeriodicalId":420,"journal":{"name":"Results in Chemistry","volume":"24 ","pages":"Article 103207"},"PeriodicalIF":5.8000,"publicationDate":"2026-05-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Results in Chemistry","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2211715626001803","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/5 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
A novel spectrofluorimetric flow injection analysis (FIA) method was developed for the selective quantification of ascorbic acid via fluorescence quenching of serotonin hydrochloride. The system employs a custom-designed photometric array comprising 16 irradiation sources arranged in a dual-axis matrix—eight aligned horizontally and eight orthogonally, enabling multi-angle excitation and enhanced spectral resolution. Fluorescence signals were captured using a twin-pair solar cell detector, offering high sensitivity and minimal optical interference. The method exhibited a linear calibration range of 0.1–30 μg/mL with a correlation coefficient (r2) of 0.9966, a limit of detection (LOD) of 0.025 μmol /L, equivalent to 4.403 * 10−4 μg/mL per sample (0.1 mL (and limit of quantitative (LOQ) of 99.381 * 10−3 . Precision and accuracy were confirmed through intra- and inter-day validation, yielding relative standard deviations (RSD) below 2% and recovery rates exceeding 99%. Statistical comparison with conventional fluorometric techniques using a paired t-test revealed no significant difference at the 95% confidence level (p > 0.05), affirming the method's reliability. The proposed system supports high-throughput analysis of up to 50 samples per hour, with reduced reagent consumption and minimal environmental impact. Its robustness, cost-efficiency, and analytical performance render it a promising tool for routine pharmaceutical quality control and environmental monitoring.