Advanced spectrofluorimetric determination of ascorbic acid via serotonin hydrochloride quenching using a twin solar cell device with eight-angled blue radiation sources

IF 5.8 Q2 CHEMISTRY, MULTIDISCIPLINARY
Results in Chemistry Pub Date : 2026-05-05 Epub Date: 2026-03-05 DOI:10.1016/j.rechem.2026.103207
Ghufran K. Allawi , N.S. Turkey
{"title":"Advanced spectrofluorimetric determination of ascorbic acid via serotonin hydrochloride quenching using a twin solar cell device with eight-angled blue radiation sources","authors":"Ghufran K. Allawi ,&nbsp;N.S. Turkey","doi":"10.1016/j.rechem.2026.103207","DOIUrl":null,"url":null,"abstract":"<div><div>A novel spectrofluorimetric flow injection analysis (FIA) method was developed for the selective quantification of ascorbic acid via fluorescence quenching of serotonin hydrochloride. The system employs a custom-designed photometric array comprising 16 irradiation sources arranged in a dual-axis matrix—eight aligned horizontally and eight orthogonally, enabling multi-angle excitation and enhanced spectral resolution. Fluorescence signals were captured using a twin-pair solar cell detector, offering high sensitivity and minimal optical interference. The method exhibited a linear calibration range of 0.1–30 μg/mL with a correlation coefficient (r<sup>2</sup>) of 0.9966, a limit of detection (LOD) of 0.025 μmol /L, equivalent to 4.403 * 10<sup>−4</sup> μg/mL per sample (0.1 mL (and limit of quantitative (LOQ) of 99.381 * 10<sup>−3</sup> <span><math><mi>μg</mi><mo>/</mo><mi>ml</mi></math></span>. Precision and accuracy were confirmed through intra- and inter-day validation, yielding relative standard deviations (RSD) below 2% and recovery rates exceeding 99%. Statistical comparison with conventional fluorometric techniques using a paired <em>t</em>-test revealed no significant difference at the 95% confidence level (<em>p</em> &gt; 0.05), affirming the method's reliability. The proposed system supports high-throughput analysis of up to 50 samples per hour, with reduced reagent consumption and minimal environmental impact. Its robustness, cost-efficiency, and analytical performance render it a promising tool for routine pharmaceutical quality control and environmental monitoring.</div></div>","PeriodicalId":420,"journal":{"name":"Results in Chemistry","volume":"24 ","pages":"Article 103207"},"PeriodicalIF":5.8000,"publicationDate":"2026-05-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Results in Chemistry","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2211715626001803","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/5 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0

Abstract

A novel spectrofluorimetric flow injection analysis (FIA) method was developed for the selective quantification of ascorbic acid via fluorescence quenching of serotonin hydrochloride. The system employs a custom-designed photometric array comprising 16 irradiation sources arranged in a dual-axis matrix—eight aligned horizontally and eight orthogonally, enabling multi-angle excitation and enhanced spectral resolution. Fluorescence signals were captured using a twin-pair solar cell detector, offering high sensitivity and minimal optical interference. The method exhibited a linear calibration range of 0.1–30 μg/mL with a correlation coefficient (r2) of 0.9966, a limit of detection (LOD) of 0.025 μmol /L, equivalent to 4.403 * 10−4 μg/mL per sample (0.1 mL (and limit of quantitative (LOQ) of 99.381 * 10−3 μg/ml. Precision and accuracy were confirmed through intra- and inter-day validation, yielding relative standard deviations (RSD) below 2% and recovery rates exceeding 99%. Statistical comparison with conventional fluorometric techniques using a paired t-test revealed no significant difference at the 95% confidence level (p > 0.05), affirming the method's reliability. The proposed system supports high-throughput analysis of up to 50 samples per hour, with reduced reagent consumption and minimal environmental impact. Its robustness, cost-efficiency, and analytical performance render it a promising tool for routine pharmaceutical quality control and environmental monitoring.

Abstract Image

采用八角蓝色辐射源双太阳能电池装置,通过盐酸5 -羟色胺猝灭法测定抗坏血酸
建立了一种利用盐酸5 -羟色胺荧光猝灭法选择性测定抗坏血酸的新方法。该系统采用定制设计的光度阵列,包括16个辐照源,排列在双轴矩阵中,8个水平排列,8个正交排列,可实现多角度激发和提高光谱分辨率。荧光信号捕获使用双对太阳能电池探测器,提供高灵敏度和最小的光学干扰。方法线性校准范围为0.1 ~ 30 μmol /mL,相关系数(r2)为0.9966,检出限(LOD)为0.025 μmol /L,相当于每样样品4.403 * 10−4 μg/mL (0.1 mL),定量限(LOQ)为99.381 * 10−3 μg/mL。精密度和准确度经日内、日间验证,相对标准偏差(RSD)小于2%,回收率大于99%。使用配对t检验与传统荧光技术进行统计比较,在95%置信水平上无显著差异(p > 0.05),证实了该方法的可靠性。该系统支持每小时多达50个样品的高通量分析,减少了试剂消耗,对环境的影响最小。它的稳健性、成本效益和分析性能使其成为常规药品质量控制和环境监测的有前途的工具。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Results in Chemistry
Results in Chemistry Chemistry-Chemistry (all)
CiteScore
2.70
自引率
8.70%
发文量
380
审稿时长
56 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书