First clinical implementation of dynamic tumor-tracking volumetric modulated arc therapy with a gimbal-mounted linac: short delivery time and high precision.
{"title":"First clinical implementation of dynamic tumor-tracking volumetric modulated arc therapy with a gimbal-mounted linac: short delivery time and high precision.","authors":"Noriko Kishi, Mitsuhiro Nakamura, Hideaki Hirashima, Takanori Adachi, Masahiro Yoneyama, Kohei Kawata, Yukako Kishigami, Takashi Ogata, Takashi Murakami, Yusuke Iizuka, Hiroshi Seno, Takashi Mizowaki","doi":"10.1093/jrr/rraf093","DOIUrl":null,"url":null,"abstract":"<p><p>Dynamic tumor tracking (DTT) is an effective respiratory motion management strategy for thoracic and abdominal tumors and enables treatment delivery under free breathing using gimbal-mounted linac, without requiring breath-holding. To date, however, clinical implementations have used three-dimensional conformal radiotherapy (3D-CRT) or fixed-field intensity-modulated radiotherapy (IMRT), and integration with volumetric modulated arc therapy (VMAT) has been anticipated to shorten treatment time while improving dose distribution. This short communication reports the first clinical implementation of stereotactic body radiotherapy with DTT-VMAT using a novel gimbal-mounted linac. Planned treatment using DTT-VMAT for liver tumor over five treatment days was delivered successfully, with a mean treatment time of 23.2 min (range, 12-33 min), shorter than previously reported gimbal-based DTT with 3D-CRT or fixed-field IMRT. The average 3D tracking error was 2.4 ± 1.8 mm overall, comprising 0.3 ± 1.0 mm laterally, -1.5 ± 2.2 mm longitudinally and - 0.1 ± 1.0 mm vertically, maintaining tracking performance comparable to earlier systems. No updates to the four-dimensional model were required during beam delivery. This introduction of novel gimbaled-based DTT-VMAT approach is advantageous among available respiratory motion management techniques, as it shortens treatment time, mitigates baseline drift and improves patient compliance.</p>","PeriodicalId":16922,"journal":{"name":"Journal of Radiation Research","volume":" ","pages":"303-307"},"PeriodicalIF":2.1000,"publicationDate":"2026-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13019128/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiation Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/jrr/rraf093","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"BIOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Dynamic tumor tracking (DTT) is an effective respiratory motion management strategy for thoracic and abdominal tumors and enables treatment delivery under free breathing using gimbal-mounted linac, without requiring breath-holding. To date, however, clinical implementations have used three-dimensional conformal radiotherapy (3D-CRT) or fixed-field intensity-modulated radiotherapy (IMRT), and integration with volumetric modulated arc therapy (VMAT) has been anticipated to shorten treatment time while improving dose distribution. This short communication reports the first clinical implementation of stereotactic body radiotherapy with DTT-VMAT using a novel gimbal-mounted linac. Planned treatment using DTT-VMAT for liver tumor over five treatment days was delivered successfully, with a mean treatment time of 23.2 min (range, 12-33 min), shorter than previously reported gimbal-based DTT with 3D-CRT or fixed-field IMRT. The average 3D tracking error was 2.4 ± 1.8 mm overall, comprising 0.3 ± 1.0 mm laterally, -1.5 ± 2.2 mm longitudinally and - 0.1 ± 1.0 mm vertically, maintaining tracking performance comparable to earlier systems. No updates to the four-dimensional model were required during beam delivery. This introduction of novel gimbaled-based DTT-VMAT approach is advantageous among available respiratory motion management techniques, as it shortens treatment time, mitigates baseline drift and improves patient compliance.
期刊介绍:
The Journal of Radiation Research (JRR) is an official journal of The Japanese Radiation Research Society (JRRS), and the Japanese Society for Radiation Oncology (JASTRO).
Since its launch in 1960 as the official journal of the JRRS, the journal has published scientific articles in radiation science in biology, chemistry, physics, epidemiology, and environmental sciences. JRR broadened its scope to include oncology in 2009, when JASTRO partnered with the JRRS to publish the journal.
Articles considered fall into two broad categories:
Oncology & Medicine - including all aspects of research with patients that impacts on the treatment of cancer using radiation. Papers which cover related radiation therapies, radiation dosimetry, and those describing the basis for treatment methods including techniques, are also welcomed. Clinical case reports are not acceptable.
Radiation Research - basic science studies of radiation effects on livings in the area of physics, chemistry, biology, epidemiology and environmental sciences.
Please be advised that JRR does not accept any papers of pure physics or chemistry.
The journal is bimonthly, and is edited and published by the JRR Editorial Committee.