The effect of implant-abutment connection fit and proximal contact tightness on screw loosening.

IF 2 3区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Journal of Advanced Prosthodontics Pub Date : 2026-02-01 Epub Date: 2026-02-13 DOI:10.4047/jap.2026.18.1.16
Hilal Gülgezen Aydın, Volkan Şahin
{"title":"The effect of implant-abutment connection fit and proximal contact tightness on screw loosening.","authors":"Hilal Gülgezen Aydın, Volkan Şahin","doi":"10.4047/jap.2026.18.1.16","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aims to evaluate how the fit between implant-abutment connection surfaces and the proximal contact tightness of crowns affect abutment screw loosening.</p><p><strong>Materials and methods: </strong>Forty implants with internal hex connections were placed in stainless steel models in the maxillary right first molar region. Monolithic zirconia crowns with standard and 15 µm reduced proximal contacts were used with abutments with standard and 25 µm horizontally reduced hex connections. After cementation, four groups of specimens-standard abutment/standard proximal contact (control), reduced fit abutment/standard proximal contact (RS), standard abutment/reduced proximal contact (SR), and reduced fit abutment/reduced proximal contact (RR)-were subjected to simultaneous thermocycling (± 5 - 55℃) and chewing cycles (100 N, 1.6 Hz, 40 mm/sec lateral and vertical speed, lateral movement: 6 mm, vertical movement: 0.5 mm). Removal torque values (RTVs) were measured using a digital torque meter. Data were analyzed using two-way ANOVA and the Bonferroni test (α = .05).</p><p><strong>Results: </strong>All groups exhibited statistically significant differences compared to each other and the control group (<i>P</i> < .05). The RR group had the lowest removal torque values (RTVs) (21.70 ± 0.99 N·cm), while the control group had the highest (26.82 ± 0.70 N·cm). The RS group (24.78 ± 0.89 N·cm) had lower RTVs than the SR group (25.51 ± 0.52 N·cm).</p><p><strong>Conclusion: </strong>The findings indicate that inadequate proximal contacts and horizontal misfit of the implant-abutment connection may lead to torque loss and screw loosening.</p>","PeriodicalId":51291,"journal":{"name":"Journal of Advanced Prosthodontics","volume":"18 1","pages":"16-24"},"PeriodicalIF":2.0000,"publicationDate":"2026-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12956434/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Advanced Prosthodontics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4047/jap.2026.18.1.16","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/2/13 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study aims to evaluate how the fit between implant-abutment connection surfaces and the proximal contact tightness of crowns affect abutment screw loosening.

Materials and methods: Forty implants with internal hex connections were placed in stainless steel models in the maxillary right first molar region. Monolithic zirconia crowns with standard and 15 µm reduced proximal contacts were used with abutments with standard and 25 µm horizontally reduced hex connections. After cementation, four groups of specimens-standard abutment/standard proximal contact (control), reduced fit abutment/standard proximal contact (RS), standard abutment/reduced proximal contact (SR), and reduced fit abutment/reduced proximal contact (RR)-were subjected to simultaneous thermocycling (± 5 - 55℃) and chewing cycles (100 N, 1.6 Hz, 40 mm/sec lateral and vertical speed, lateral movement: 6 mm, vertical movement: 0.5 mm). Removal torque values (RTVs) were measured using a digital torque meter. Data were analyzed using two-way ANOVA and the Bonferroni test (α = .05).

Results: All groups exhibited statistically significant differences compared to each other and the control group (P < .05). The RR group had the lowest removal torque values (RTVs) (21.70 ± 0.99 N·cm), while the control group had the highest (26.82 ± 0.70 N·cm). The RS group (24.78 ± 0.89 N·cm) had lower RTVs than the SR group (25.51 ± 0.52 N·cm).

Conclusion: The findings indicate that inadequate proximal contacts and horizontal misfit of the implant-abutment connection may lead to torque loss and screw loosening.

Abstract Image

Abstract Image

Abstract Image

种植体-基台连接配合及近端接触紧密度对螺钉松动的影响。
目的:探讨种植体-基牙连接面与冠近端接触紧密度对基牙螺钉松动的影响。材料和方法:在上颌右第一磨牙区采用不锈钢模型种植40颗内六角连接种植体。采用标准和15µm近端缩小接触的单片氧化锆冠与标准和25µm水平缩小内六角连接的基台配合使用。胶结后,标准基台/标准近端接触(对照)、缩小配合基台/标准近端接触(RS)、标准基台/缩小近端接触(SR)和缩小配合基台/缩小近端接触(RR)四组样品同时进行热循环(±5 ~ 55℃)和咀嚼循环(100 N, 1.6 Hz, 40 mm/秒的横向和纵向速度,横向运动6 mm,纵向运动0.5 mm)。使用数字扭矩计测量去除扭矩值(rtv)。数据分析采用双因素方差分析和Bonferroni检验(α = 0.05)。结果:各组与对照组比较,差异均有统计学意义(P < 0.05)。RR组去除扭矩值(RTVs)最低(21.70±0.99 N·cm),对照组最高(26.82±0.70 N·cm)。RS组rtv(24.78±0.89 N·cm)低于SR组(25.51±0.52 N·cm)。结论:种植体-基牙近端接触不足和水平错位可导致扭矩损失和螺钉松动。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Advanced Prosthodontics
Journal of Advanced Prosthodontics DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
4.70
自引率
3.80%
发文量
25
期刊介绍: This journal aims to convey scientific and clinical progress in the field of prosthodontics and its related areas to many dental communities concerned with esthetic and functional restorations, occlusion, implants, prostheses, and biomaterials related to prosthodontics. This journal publishes • Original research data of high scientific merit in the field of diagnosis, function, esthetics and stomatognathic physiology related to prosthodontic rehabilitation, physiology and mechanics of occlusion, mechanical and biologic aspects of prosthodontic materials including dental implants. • Review articles by experts on controversies and new developments in prosthodontics. • Case reports if they provide or document new fundamental knowledge.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书