Accuracy of the Withings BPM Pro 2 (WPM07) Device for Self-Blood Pressure Measurements in the General Populations According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 2:2024) Protocol.

IF 3.2 Q2 PERIPHERAL VASCULAR DISEASE
Vascular Health and Risk Management Pub Date : 2026-02-25 eCollection Date: 2026-01-01 DOI:10.2147/VHRM.S585947
Parounak Zelveian, Zoya Hakobyan, Jirar Topouchian, Avag Avagyan, Heghine Gharibyan, Roland Asmar
{"title":"Accuracy of the Withings BPM Pro 2 (WPM07) Device for Self-Blood Pressure Measurements in the General Populations According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 2:2024) Protocol.","authors":"Parounak Zelveian, Zoya Hakobyan, Jirar Topouchian, Avag Avagyan, Heghine Gharibyan, Roland Asmar","doi":"10.2147/VHRM.S585947","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The accuracy of blood pressure (BP) measurements performed by electronic devices is crucial in clinical practice. International professional societies emphasize the need for rigorous, standardized validation of electronic BP measuring devices to ensure their accuracy and reliability.</p><p><strong>Objective: </strong>This study evaluated the accuracy of the Withings BPM Pro2 (WPM07) upper‑arm oscillometric BP monitor in the adult general population using the ISO 81060‑2:2018/AMD 2:2024 Universal Standard.</p><p><strong>Methods: </strong>Following the ISO same‑arm sequential measurement protocol, 89 eligible pre-selected participants underwent paired test‑device and mercury reference auscultatory measurements. Nine BP measurements were performed on each participant, four with the tested device and five with the auscultatory method. The accuracy of BP measurements was assessed based on the difference between simultaneous measurements taken by two observers and those displayed by the tested device using ISO Criterion 1 (mean and standard deviation of test-reference differences) and Criterion 2 (within‑subject variability of averaged difference values).</p><p><strong>Results: </strong>Eighty-nine participants underwent the BP measurements protocol, data from 87 were analyzed after excluding 2 participants due to excessive BP variability. Inter‑observer agreement for reference measurements was 0.64 ± 2.14 mmHg for systolic BP (SBP) and 0.39 ± 1.98 mmHg for diastolic BP (DBP). ISO criterion 1 was met with mean test-reference differences of 1.0 ± 4.9 mmHg (SBP) and 0.6 ± 4.1 mmHg (DBP) below ISO requirements (≤ 5±8 mmHg). ISO criterion 2 was also satisfied, with within‑subject standard deviations of 3.50 mmHg (SBP) and 3.01 mmHg (DBP), well below the ISO thresholds (≤6.87/6.91 mmHg).</p><p><strong>Conclusion: </strong>The BPM Pro2 (WPM07) fulfills the ISO 81060‑2:2018/AMD 2:2024 accuracy requirements and is suitable for reliable home BP self‑measurement in adults.</p>","PeriodicalId":23597,"journal":{"name":"Vascular Health and Risk Management","volume":"22 ","pages":"585947"},"PeriodicalIF":3.2000,"publicationDate":"2026-02-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12950639/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Vascular Health and Risk Management","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/VHRM.S585947","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The accuracy of blood pressure (BP) measurements performed by electronic devices is crucial in clinical practice. International professional societies emphasize the need for rigorous, standardized validation of electronic BP measuring devices to ensure their accuracy and reliability.

Objective: This study evaluated the accuracy of the Withings BPM Pro2 (WPM07) upper‑arm oscillometric BP monitor in the adult general population using the ISO 81060‑2:2018/AMD 2:2024 Universal Standard.

Methods: Following the ISO same‑arm sequential measurement protocol, 89 eligible pre-selected participants underwent paired test‑device and mercury reference auscultatory measurements. Nine BP measurements were performed on each participant, four with the tested device and five with the auscultatory method. The accuracy of BP measurements was assessed based on the difference between simultaneous measurements taken by two observers and those displayed by the tested device using ISO Criterion 1 (mean and standard deviation of test-reference differences) and Criterion 2 (within‑subject variability of averaged difference values).

Results: Eighty-nine participants underwent the BP measurements protocol, data from 87 were analyzed after excluding 2 participants due to excessive BP variability. Inter‑observer agreement for reference measurements was 0.64 ± 2.14 mmHg for systolic BP (SBP) and 0.39 ± 1.98 mmHg for diastolic BP (DBP). ISO criterion 1 was met with mean test-reference differences of 1.0 ± 4.9 mmHg (SBP) and 0.6 ± 4.1 mmHg (DBP) below ISO requirements (≤ 5±8 mmHg). ISO criterion 2 was also satisfied, with within‑subject standard deviations of 3.50 mmHg (SBP) and 3.01 mmHg (DBP), well below the ISO thresholds (≤6.87/6.91 mmHg).

Conclusion: The BPM Pro2 (WPM07) fulfills the ISO 81060‑2:2018/AMD 2:2024 accuracy requirements and is suitable for reliable home BP self‑measurement in adults.

根据国际标准化组织通用标准(ISO 81060-2:2018/AMD 2:2024)协议,用于普通人群自我血压测量的Withings BPM Pro 2 (WPM07)设备的准确性。
背景:在临床实践中,电子设备测量血压(BP)的准确性至关重要。国际专业协会强调需要对电子BP测量设备进行严格、标准化的验证,以确保其准确性和可靠性。目的:本研究使用ISO 81060‑2:2018/AMD 2:2024通用标准评估Withings BPM Pro2 (WPM07)上臂振荡式血压监测仪在成年普通人群中的准确性。方法:遵循ISO同臂顺序测量方案,89名符合条件的预选参与者接受了配对测试装置和汞参考听诊测量。对每位参与者进行了9次血压测量,其中4次使用测试设备,5次使用听诊方法。根据ISO标准1(测试参考差异的平均值和标准差)和标准2(平均差异值的受试者内部变异性),根据两名观察者同时测量的结果与被测设备显示的结果之间的差异来评估血压测量的准确性。结果:89名参与者接受了血压测量方案,在排除2名因血压变异性过大的参与者后,对87名参与者的数据进行了分析。参考测量的观察者间一致性收缩压(SBP)为0.64 ± 2.14 mmHg,舒张压(DBP)为0.39 ± 1.98 mmHg。ISO标准1的平均测试参考差值分别为1.0 ± 4.9 mmHg (SBP)和0.6 ± 4.1 mmHg (DBP),低于ISO要求(≤5±8 mmHg)。ISO标准2也得到满足,受试者内标准差为3.50 mmHg (SBP)和3.01 mmHg (DBP),远低于ISO阈值(≤6.87/6.91 mmHg)。结论:BPM Pro2 (WPM07)满足ISO 81060‑2:2018/AMD 2:2024精度要求,适用于成人可靠的家庭血压自我测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Vascular Health and Risk Management
Vascular Health and Risk Management PERIPHERAL VASCULAR DISEASE-
CiteScore
4.20
自引率
3.40%
发文量
109
审稿时长
16 weeks
期刊介绍: An international, peer-reviewed journal of therapeutics and risk management, focusing on concise rapid reporting of clinical studies on the processes involved in the maintenance of vascular health; the monitoring, prevention, and treatment of vascular disease and its sequelae; and the involvement of metabolic disorders, particularly diabetes. In addition, the journal will also seek to define drug usage in terms of ultimate uptake and acceptance by the patient and healthcare professional.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书