Comparative assessment of spherical and cylindrical ionization chambers (PTW 32002 and PTW 23361) in 60Co and 137Cs reference beams: experimental and Monte Carlo evaluation for installation compliance in gamma irradiator.
{"title":"Comparative assessment of spherical and cylindrical ionization chambers (PTW 32002 and PTW 23361) in 60Co and 137Cs reference beams: experimental and Monte Carlo evaluation for installation compliance in gamma irradiator.","authors":"Taibi Zidouz, Asmaa Mouhssine, Hasbiya Tamiss, Abdessamad Didi, Abdellatif Talbi, Khalid Nabaoui, Elmehdi Alibrahmi, El Mahjoub Chakir, Abdelwahab Badague, Mohamed Zaryah","doi":"10.1093/rpd/ncag014","DOIUrl":null,"url":null,"abstract":"<p><p>Accredited calibration laboratories operating under ISO/IEC 17025 must demonstrate compliance of their facilities with technical standards such as ISO 4037 when performing calibrations in gamma and X-ray fields, typically using ionization chambers as reference instruments. In this framework, the accurate determination of the air-kerma rate is a critical step to ensure both calibration reliability and international traceability in radiation protection dosimetry. This study reports a comparative evaluation of two ionization chambers with different geometries: the PTW 32002 (spherical, 1 L) and the PTW 23361 (cylindrical, 30 cm3). Both were employed to characterize 137Cs and 60Co reference radiation fields at the CNESTEN calibration laboratory. The experimental results were compared against a theoretical benchmark provided by Monte Carlo simulations using the MCNP6.1 code to perform a comparative study with the measured values. The aim was to verify the installation's compliance with the inverse square law and to assess the metrological suitability of the two detectors. The results showed overall agreement between experiment and simulation, with deviations ranging from ~0.2% to 10.5%, depending on photon energy and source-to-detector distance. The spherical PTW 32002 chamber demonstrated greater stability in 60Co beams, while the cylindrical PTW 23361 exhibited better accuracy in 137Cs fields. These findings highlight the importance of chamber selection with respect to photon energy and dose rate, and confirm the compliance of the CNESTEN facility with international standards, thereby supporting reduced uncertainties and improved metrological traceability in calibration services.</p>","PeriodicalId":20795,"journal":{"name":"Radiation protection dosimetry","volume":" ","pages":"827-841"},"PeriodicalIF":0.8000,"publicationDate":"2026-06-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation protection dosimetry","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1093/rpd/ncag014","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENVIRONMENTAL SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
Accredited calibration laboratories operating under ISO/IEC 17025 must demonstrate compliance of their facilities with technical standards such as ISO 4037 when performing calibrations in gamma and X-ray fields, typically using ionization chambers as reference instruments. In this framework, the accurate determination of the air-kerma rate is a critical step to ensure both calibration reliability and international traceability in radiation protection dosimetry. This study reports a comparative evaluation of two ionization chambers with different geometries: the PTW 32002 (spherical, 1 L) and the PTW 23361 (cylindrical, 30 cm3). Both were employed to characterize 137Cs and 60Co reference radiation fields at the CNESTEN calibration laboratory. The experimental results were compared against a theoretical benchmark provided by Monte Carlo simulations using the MCNP6.1 code to perform a comparative study with the measured values. The aim was to verify the installation's compliance with the inverse square law and to assess the metrological suitability of the two detectors. The results showed overall agreement between experiment and simulation, with deviations ranging from ~0.2% to 10.5%, depending on photon energy and source-to-detector distance. The spherical PTW 32002 chamber demonstrated greater stability in 60Co beams, while the cylindrical PTW 23361 exhibited better accuracy in 137Cs fields. These findings highlight the importance of chamber selection with respect to photon energy and dose rate, and confirm the compliance of the CNESTEN facility with international standards, thereby supporting reduced uncertainties and improved metrological traceability in calibration services.
期刊介绍:
Radiation Protection Dosimetry covers all aspects of personal and environmental dosimetry and monitoring, for both ionising and non-ionising radiations. This includes biological aspects, physical concepts, biophysical dosimetry, external and internal personal dosimetry and monitoring, environmental and workplace monitoring, accident dosimetry, and dosimetry related to the protection of patients. Particular emphasis is placed on papers covering the fundamentals of dosimetry; units, radiation quantities and conversion factors. Papers covering archaeological dating are included only if the fundamental measurement method or technique, such as thermoluminescence, has direct application to personal dosimetry measurements. Papers covering the dosimetric aspects of radon or other naturally occurring radioactive materials and low level radiation are included. Animal experiments and ecological sample measurements are not included unless there is a significant relevant content reason.