Tali Drori, Amir Dori, Zehavit Goldberg, Valery Golderman, Polina Sonis, Michael Gurevich, Rina Zilkha-Falb, Joab Chapman, Efrat Shavit-Stein
{"title":"Neurofilament Light Chain Measurements by Centaur and Simoa Systems in Human and Murine Samples.","authors":"Tali Drori, Amir Dori, Zehavit Goldberg, Valery Golderman, Polina Sonis, Michael Gurevich, Rina Zilkha-Falb, Joab Chapman, Efrat Shavit-Stein","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Neurofilament light chain (NfL) is an established biomarker for detecting axonal injury in various neurological disorders. The Quanterix Single Molecule Array (Simoa) is the current standard; however, automated immunoassays, such as the Siemens Atellica and Centaur, may serve as alternatives.</p><p><strong>Objectives: </strong>To compare NfL measurements obtained with the Centaur system to those from the Simoa-SR-X. To assess their agreement and applicability in clinical practice, research, and animal studies.</p><p><strong>Methods: </strong>NfL levels were measured in 27 human serum, 8 plasma, and 16 cerebrospinal fluid (CSF) samples, and 9 murine serum samples, by Centaur and Simoa systems. NfL levels in concomitantly drawn serum and plasma were compared in 8 humans. The agreement between platforms was evaluated.</p><p><strong>Results: </strong>NfL levels measured by Centaur and Simoa systems demonstrated a strong correlation in serum (Spearman r=0.97, P < 0.0001) and plasma (Pearson R²=0.95, P < 0.0001). Centaur measurements were higher (P = 0.01) than Simoa. Most importantly, system-specific Z-scores corrected these differences. Serum and plasma levels measured by the Centaur system correlated strongly (R²=0.98, P < 0.0001) and showed similar results. CSF levels measured by the Centaur system were lower (52% bias) than those measured by Simoa, with poor correlation at concentrations within the normal range (R2=0.32, P = 0.11). Mouse serum results showed a strong correlation between systems (R²=0.86, P < 0.001) with similar values.</p><p><strong>Conclusions: </strong>The Centaur system offers an alternative to Simoa for measuring NfL in human serum, plasma, and murine serum. System-specific age-adjusted Z-scores are essential for interpretation. CSF evaluation requires further assessment.</p>","PeriodicalId":50268,"journal":{"name":"Israel Medical Association Journal","volume":"28 2","pages":"77-81"},"PeriodicalIF":0.8000,"publicationDate":"2026-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Israel Medical Association Journal","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Neurofilament light chain (NfL) is an established biomarker for detecting axonal injury in various neurological disorders. The Quanterix Single Molecule Array (Simoa) is the current standard; however, automated immunoassays, such as the Siemens Atellica and Centaur, may serve as alternatives.
Objectives: To compare NfL measurements obtained with the Centaur system to those from the Simoa-SR-X. To assess their agreement and applicability in clinical practice, research, and animal studies.
Methods: NfL levels were measured in 27 human serum, 8 plasma, and 16 cerebrospinal fluid (CSF) samples, and 9 murine serum samples, by Centaur and Simoa systems. NfL levels in concomitantly drawn serum and plasma were compared in 8 humans. The agreement between platforms was evaluated.
Results: NfL levels measured by Centaur and Simoa systems demonstrated a strong correlation in serum (Spearman r=0.97, P < 0.0001) and plasma (Pearson R²=0.95, P < 0.0001). Centaur measurements were higher (P = 0.01) than Simoa. Most importantly, system-specific Z-scores corrected these differences. Serum and plasma levels measured by the Centaur system correlated strongly (R²=0.98, P < 0.0001) and showed similar results. CSF levels measured by the Centaur system were lower (52% bias) than those measured by Simoa, with poor correlation at concentrations within the normal range (R2=0.32, P = 0.11). Mouse serum results showed a strong correlation between systems (R²=0.86, P < 0.001) with similar values.
Conclusions: The Centaur system offers an alternative to Simoa for measuring NfL in human serum, plasma, and murine serum. System-specific age-adjusted Z-scores are essential for interpretation. CSF evaluation requires further assessment.