Lei Wang, Da-Wei Wang, Rui-Ming Xu, Jian-Zhong Li, Jing Zhang
{"title":"Comparative Evaluation of Three Transthoracic Echocardiographic Techniques for Cardiac Output and Stroke Volume Assessment Across Hemodynamic States","authors":"Lei Wang, Da-Wei Wang, Rui-Ming Xu, Jian-Zhong Li, Jing Zhang","doi":"10.1111/anec.70162","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objective</h3>\n \n <p>This study compared the accuracy of three transthoracic echocardiographic (TTE) techniques—fractional shortening (FS), left ventricular outflow tract velocity-time integral (LVOT/VTI), and Simpson's method—for measuring cardiac output (CO) and stroke volume (SV) in hemodynamically unstable patients (systolic pressure < 90 mmHg), using pulse index continuous cardiac output (PiCCO) as the reference.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A retrospective analysis was conducted involving 12 patients admitted to an Emergency Intensive Care Unit between October 2023 and October 2024, who underwent a total of 54 echocardiographic examinations. The median length of hospital stay is 14 days. CO and SV values obtained using the three TTE methods as part of routine assessment were compared with simultaneous PiCCO measurements. The median number of examinations for patients is 4. Statistical analysis was performed using correlation methods.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>All three TTE methods demonstrated the capability to estimate CO and SV. The VTI method showed the highest overall accuracy (CO-VTI vs. CO-PiCCO: <i>r</i> = 0.950, <i>p</i>< 0.001). Under reduced CO conditions, correlations between echocardiographic and PiCCO-derived measurements decreased for all methods; however, the VTI method maintained superior reliability (<i>r</i> = 0.606, <i>p</i> = 0.006). In contrast, the Simpson's method did not accurately reflect CO in this setting (SV-Simpson vs. SV-PiCCO: <i>r</i> = 0.408, <i>p</i> = 0.083). Notably, the performance of the VTI method remained consistent regardless of SV or heart rate variations (SV-VTI vs. SV-PiCCO: overall <i>r</i> = 0.970; with heart rate > 100 bpm, <i>r</i> = 0.946; with heart rate ≤ 100 bpm, <i>r</i> = 0.988).</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The LVOT/VTI method exhibits the highest accuracy and consistency for measuring SV and CO, making it the preferred non-invasive technique for hemodynamic evaluation in critically ill patients.</p>\n </section>\n </div>","PeriodicalId":8074,"journal":{"name":"Annals of Noninvasive Electrocardiology","volume":"31 2","pages":""},"PeriodicalIF":0.9000,"publicationDate":"2026-02-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12932115/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Noninvasive Electrocardiology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/anec.70162","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
This study compared the accuracy of three transthoracic echocardiographic (TTE) techniques—fractional shortening (FS), left ventricular outflow tract velocity-time integral (LVOT/VTI), and Simpson's method—for measuring cardiac output (CO) and stroke volume (SV) in hemodynamically unstable patients (systolic pressure < 90 mmHg), using pulse index continuous cardiac output (PiCCO) as the reference.
Methods
A retrospective analysis was conducted involving 12 patients admitted to an Emergency Intensive Care Unit between October 2023 and October 2024, who underwent a total of 54 echocardiographic examinations. The median length of hospital stay is 14 days. CO and SV values obtained using the three TTE methods as part of routine assessment were compared with simultaneous PiCCO measurements. The median number of examinations for patients is 4. Statistical analysis was performed using correlation methods.
Results
All three TTE methods demonstrated the capability to estimate CO and SV. The VTI method showed the highest overall accuracy (CO-VTI vs. CO-PiCCO: r = 0.950, p< 0.001). Under reduced CO conditions, correlations between echocardiographic and PiCCO-derived measurements decreased for all methods; however, the VTI method maintained superior reliability (r = 0.606, p = 0.006). In contrast, the Simpson's method did not accurately reflect CO in this setting (SV-Simpson vs. SV-PiCCO: r = 0.408, p = 0.083). Notably, the performance of the VTI method remained consistent regardless of SV or heart rate variations (SV-VTI vs. SV-PiCCO: overall r = 0.970; with heart rate > 100 bpm, r = 0.946; with heart rate ≤ 100 bpm, r = 0.988).
Conclusion
The LVOT/VTI method exhibits the highest accuracy and consistency for measuring SV and CO, making it the preferred non-invasive technique for hemodynamic evaluation in critically ill patients.
目的:本研究比较了三种经胸超声心动图(TTE)技术——分数缩短(FS)、左心室流出道速度-时间积分(LVOT/VTI)和辛普森法在血流动力学不稳定患者(收缩压)中测量心输出量(CO)和卒中容积(SV)的准确性。对2023年10月至2024年10月期间入住急诊重症监护病房的12例患者进行了回顾性分析,共进行了54次超声心动图检查。住院时间中位数为14天。使用三种TTE方法作为常规评估的一部分获得的CO和SV值与同时测量的PiCCO进行比较。患者检查次数中位数为4次。采用相关法进行统计分析。结果:三种TTE方法均能估计CO和SV。VTI法整体准确率最高(CO-VTI vs. CO-PiCCO: r = 0.950, p 100 bpm, r = 0.946;心率≤100 bpm, r = 0.988)。结论:LVOT/VTI法测量SV和CO具有较高的准确性和一致性,是危重患者血流动力学评估的首选无创技术。
期刊介绍:
The ANNALS OF NONINVASIVE ELECTROCARDIOLOGY (A.N.E) is an online only journal that incorporates ongoing advances in the clinical application and technology of traditional and new ECG-based techniques in the diagnosis and treatment of cardiac patients.
ANE is the first journal in an evolving subspecialty that incorporates ongoing advances in the clinical application and technology of traditional and new ECG-based techniques in the diagnosis and treatment of cardiac patients. The publication includes topics related to 12-lead, exercise and high-resolution electrocardiography, arrhythmias, ischemia, repolarization phenomena, heart rate variability, circadian rhythms, bioengineering technology, signal-averaged ECGs, T-wave alternans and automatic external defibrillation.
ANE publishes peer-reviewed articles of interest to clinicians and researchers in the field of noninvasive electrocardiology. Original research, clinical studies, state-of-the-art reviews, case reports, technical notes, and letters to the editors will be published to meet future demands in this field.